《ASTM C1344-1997(2013) Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Mass Spectrometer Method《使用普通气源质谱仪对六氟化铀进行同位素分析的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1344-1997(2013) Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Mass Spectrometer Method《使用普通气源质谱仪对六氟化铀进行同位素分析的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1344 97 (Reapproved 2013)Standard Test Method forIsotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Mass Spectrometer Method1This standard is issued under the fixed designation C1344; the number immediately following the designation indicates the year oforiginal ad
2、option or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the isotopic analysis of uraniumhexafluoride (
3、UF6) and may be used for the entire range of235U isotopic compositions for which standards are available.1.2 This test method is applicable to the determination ofthe isotopic relationship between two UF6samples. If theabundance of a specific isotope of one sample (the standard) isknown, its abundan
4、ce in the other can be determined. This testmethod is flexible in that the number of times a given materialis admitted to the ion source may be adjusted to the minimumrequired for a specified precision level.1.3 The sensitivity with which differences between twomaterials can be detected depends on t
5、he measuring systemused, but ratio-measuring devices can generally read ratio-of-mol ratio differences as small as 0.0001.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of t
6、hesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific hazardsstatements are given in Section 7.2. Referenced Docu
7、ments2.1 ASTM Standards:2C787 Specification for Uranium Hexafluoride for Enrich-mentC996 Specification for Uranium Hexafluoride Enriched toLess Than 5 %235U2.2 Other Document:USEC-651, Uranium Hexafluoride: A Manual of GoodHandling Practices33. Terminology3.1 Definitions of Terms Specific to This St
8、andard:3.1.1 drop through, na measurement of the amount of the238UF5+ion beam that can be passed through the235UF5+collector slit and measured on the235UF5+collector, stated asa percentage of the total238UF5+signal.3.1.2 memory corrections, ncorrections applied to thesample analysis results for memo
9、ry effects.3.1.3 memory effect, nthe inability of the mass spectrom-eter to omit completely the isotopic composition of the sampleanalyzed previously from attributing to the results of furthersamples analyzed.3.1.4 normal isotopic abundance material, nUF6having avalue of 0.711 weight percent (wt %)2
10、35U.3.1.5 ratio-of-mol-ratios, nthe mol ratio (235U/238U) ofthe sample divided by the mol ratio of the standard, or theinverse condition of the mol ratio of the standard divided by themol ratio of the sample.4. Summary of Test Method4.1 Test MethodThe unknown sample and a standard withan isotopic co
11、mposition close to that of the sample areintroduced in sequence into the Neir mass spectrometer. UF5+ions of the isotopes are focused through a mass-resolvingcollector slit and onto a faraday cup collector. Measurementsare made of235UF5+to the total of the other UF5+isotopes.With the known compositi
12、on of the standard, calculation of the235U composition of the sample can be determined.5. Significance and Use5.1 Uranium hexafluoride is a basic material used to preparenuclear reactor fuel. To be suitable for this purpose, thematerial must meet the criteria for isotopic composition. This1This test
13、 method is under the jurisdiction ofASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved July 1, 2013. Published July 2013. Originally approvedin 1997. Last previous edition approved in 2008 as C1344 97 (2008)1.DOI
14、:10.1520/C1344-97R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from U.S. Enrichment Corporat
15、ion, 6903 Rockledge Dr., Bethesda,MD 20817, http:/.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1test method is designed to determine whether the materialmeets the requirements described in Specifications C787 andC996.5.2 ASTM Commi
16、ttee C-26 Safeguards Statement:5.2.1 The material (uranium hexafluoride) to which this testmethod applies is subject to the nuclear safeguards regulationsgoverning its possession and use. The analytical procedure inthis test method has been designated as technically acceptablefor generating safeguar
17、ds accountability data.5.2.2 When used in conjunction with appropriate certifiedreference materials (CRMs), this procedure can demonstratetraceability to the national measurement base. However, adher-ence to this procedure does not automatically guarantee regu-latory acceptance of the regulatory saf
18、eguards measurements.It remains the sole responsibility of the user of this test methodto ensure that its application to safeguards has the approval ofthe proper regulatory authorities.6. Apparatus6.1 Neir Mass Spectrometer, with the following features andcapabilities:6.1.1 A single-focusing spectro
19、meter, with a 127-mm mini-mum deflection radius, is satisfactory when equipped andfocused as follows:6.1.1.1 The sample inlet system must have two sampleholders, to which UF6containers can be attached, and thenecessary valves to evacuate the sample lines through whichthe sample and standard are intr
20、oduced. The sample inletsystem should be nickel or Monel for use with corrosive gases,and should have minimum volume.6.1.1.2 A single adjustable leak, operated by an automaticleak control mechanism for admitting the sample into thespectrometer ion source, is preferred.6.1.1.3 The pumping system of t
21、he spectrometer analyzertube must maintain a pressure below 5 108torr with sampleflowing into the ion source.6.1.1.4 Focus the instrument for resolution consistent withprecision requirements. A high-current ion beam of 5 1010to 1109amps is necessary, with a signal-to-noise ratiogreater than 3000 in
22、the low-current amplifier system.6.1.1.5 A dual collector must be used, so that ions from oneisotope are passed through a resolving slit and focused on alow-current collector, and ions from all other isotopes arefocused on a high-current collector. The preferred method ofmaintaining the low-current
23、ion beam within the collector slitis by an automatic beam positioner circuit.Aresolving slit withadjustable width features enhances the measurement of allisotopes but is not mandatory for isotopic measurements.6.1.1.6 The amplified high- and low-current signals are fedinto a multimeter or other devi
24、ce capable of ratioing high- andlow-current signals. If a multimeter is used, the multimetermust have a minimum of 5.5 digits of resolution, a means ofratioing the high- and low-current signals, and interactivecommunication capability with the controller.6.1.1.7 The memory effect of the spectrometer
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC134419972013STANDARDTESTMETHODFORISOTOPICANALYSISOFURANIUMHEXAFLUORIDEBYSINGLESTANDARDGASSOURCEMASSSPECTROMETERMETHOD

链接地址:http://www.mydoc123.com/p-464301.html