1、 A-A-59604 12 July 2004 COMMERCIAL ITEM DESCRIPTION TEST SET, RADIO FREQUENCY POWER (50 MHz to 40 GHz) The General Services Administration has authorized the use of this commercial item description for all federal agencies. 1. SCOPE. This commercial item description describes the minimum requirement
2、s for a Radio Frequency Power Test Set, which consists of a dual channel radio frequency (RF) power meter, sensors, and applicable accessories. The equipment, as described in this CID, has a minimum frequency range of 50 MHz to 40 GHz and will be used for measuring RF average power over selected fre
3、quencies. Sensors supplied with the test set have a minimum frequency range of 50 MHz to 18 GHz. This CID is meant as a minimum requirement for the power meter in which only those manufacturers that meet or surpass the following requirements are supplied per this CID. 2. SALIENT CHARACTERISTICS. The
4、 equipment shall be capable of operation within the accuracies, limits, and specifications herein. 2.1 Classification. Equipment covered by this CID shall be commercially available equipment and may be modified to the extent necessary to meet the following description. The equipment shall be Class 3
5、, in accordance with MIL-PRF-28800, except as specified he rein. 2.2 Safety and Environmental. The equipment shall meet all safety and environmental requirements as specified in MIL-PRF-28800 for the classification as stated herein except as specified herein. 2.2.1 Temperature Operating. The equipme
6、nt shall conform to its specified performance and accuracy while being operated in a temperature range of 0C to +50C. AMSC N/A FSC 6625 INCH-POUNDComments, suggestions, or questions on this document should be addressed to: WR-ALC/LEEC, 295 Byron Street, Robins AFB, GA 31098-1611. Since contact infor
7、mation can change, you may want to verify the currency of this address information using the ASSIST Online database at www.dodssp.daps.mil. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59604 2 2.2.2 Temperature Non-operating. The equipment sha
8、ll conform to its specified performance and accuracy after being stored in an environment in the temperature range of -20C to +70C. 2.3. Electrical Power Sources. The equipment shall operate from nominal commercial, military, and shipboard power sources of 120 VAC (10 %) at line frequencies of 48 Hz
9、 to 63 Hz and 380 to 420 Hz and 240 VAC (10 %) at line frequencies of 47 Hz to 63 Hz. Transient state conditions shall meet MIL-PRF-28800 requirements for the classification specified herein. 2.4 Reliability. The design of the equipment shall be such that under normal use and operation the equipment
10、 does not fail within 10000 hours of operation with a statistical certainty of 95%. 2.5 Calibration and Maintenance Adjustments. The design of the system shall provide for readily accessible calibration adjustments and maintenance adjustments. The calibration adjustments, wherever possible, shall be
11、 accessible without removal of the equipment case or modules. The calibration interval shall be a period of one year or greater based on an operating time of 2000 hours. No special tools shall be required for alignment or adjustment except those furnished with the equipment. 2.6 Performance Requirem
12、ents. The equipment shall meet all requirements specified herein, after the warm-up period specified in MIL-PRF-28800, under all combinations of input power conditions, output signal conditions and operating service condition as specified in MIL-PRF-28800. 2.6.1 Meter/Chassis. The averaging power me
13、ter shall have a minimum of two measurement channels. 2.6.1.1 Remote Operation. The averaging power meter shall have an IEEE-488 interface for remote set up, control, and measurement operation. 2.6.1.2 Front Panel Controls. Controls for the power meter shall be included on the front-panel. 2.6.1.3 M
14、odes. The power meter shall have the selection of either linear (mW) or logarithmic (dBm) modes for the display of the power measurements. 2.6.1.4 Frequency Measurement Range. The minimum frequency range of the averaging power meter without regard to sensors shall be 50 MHz to 40 GHz. The minimum fr
15、equency range of the averaging power meter with sensors supplied shall be 50 MHz to 18 GHz. 2.6.1.5 Power Measurement Range. The averaging power meter with sensors shall have, as a minimum, a range of -30 dBm to 20 dBm, regardless of modulation format. 2.6.1.5.1 Power Meter Accuracy. The instrument
16、uncertainty shall be within 0.02 dB at full scale in logarithmic mode and within 0.5% in linear mode. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59604 3 2.6.1.6 Display/Time Base. The averaging power meter shall have a digital display for al
17、l functions. The display shall provide, as a minimum, information on the amplitude of the measured signal. 2.6.1.6.1 Indicators. The display shall have indicators for measurement mode, display units (watts, dBm, dB), calibration factor, error conditions, and offset value. 2.6.1.6.2 Amplitude Informa
18、tion Display. As a minimum, averaging power amplitude information shall be available for display in logarithmic form. The equipment shall be capable of displaying the amplitude measurement range specified herein. 2.6.1.6.2.1 Range. The equipment shall include a display range of 99.99 dB to +99.99 dB
19、. 2.6.1.6.2.2 Offset. The equipment shall have an amplitude offset capability to include a range from 99.99 dB to +99.99 dB. 2.6.1.7 Instrument Zeroing. The power meter shall have automatic zeroing to within 0.5 % of full scale using the highest sensitivity range. 2.6.1.8 Power Reference Calibration
20、 Source. An internal power reference source, and associated operating mode, shall be provided to allow user-selected automatic calibration of the combination of the sensor(s) and the equipment. 2.6.1.8.1 Frequency. The power calibration source frequency shall be 50 MHz 5%. 2.6.1.8.2 Amplitude. The p
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA596042004TESTSETRADIOFREQUENCYPOWER50MHZTO40GHZ 射频 功率 测试 装置 50 MHZ 40 GHZPDF

链接地址:http://www.mydoc123.com/p-427128.html