ASTM F2129-17b Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices.pdf
《ASTM F2129-17b Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices.pdf》由会员分享,可在线阅读,更多相关《ASTM F2129-17b Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2129 17bStandard Test Method forConducting Cyclic Potentiodynamic PolarizationMeasurements to Determine the Corrosion Susceptibility ofSmall Implant Devices1This standard is issued under the fixed designation F2129; the number immediately following the designation indicates the year of
2、original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method assesses the corrosion susceptibility ofsm
3、all, metallic, implant medical devices, or componentsthereof, using cyclic (forward and reverse) potentiodynamicpolarization. Examples of device types that may be evaluatedby this test method include, but are not limited to, vascularstents, ureteral stents (Specification F1828), filters, supportsegm
4、ents of endovascular grafts, cardiac occluders, aneurysmor ligation clips, staples, and so forth.1.2 This test method is used to assess a device in its finalform and finish, as it would be implanted. These small devicesshould be tested in their entirety. The upper limit on device sizeis dictated by
5、the electrical current delivery capability of thetest apparatus (see Section 6). It is assumed that test methods,such as Reference Test Method G5 and Test Method G61 havebeen used for material screening.1.3 Because of the variety of configurations and sizes ofimplants, this test method provides a va
6、riety of specimenholder configurations.1.4 This test method is intended for use on implantabledevices made from metals with a relatively high resistance tocorrosion.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This st
7、andard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1.7 This international sta
8、ndard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2.
9、Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterE177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodF1828 Specification for Ureteral StentsG3 Practice for
10、 Conventions Applicable to ElectrochemicalMeasurements in Corrosion TestingG5 Reference Test Method for Making PotentiodynamicAnodic Polarization MeasurementsG15 Terminology Relating to Corrosion and Corrosion Test-ing (Withdrawn 2010)3G61 Test Method for Conducting Cyclic PotentiodynamicPolarizatio
11、n Measurements for Localized Corrosion Sus-ceptibility of Iron-, Nickel-, or Cobalt-Based Alloys3. Terminology3.1 Definitions:3.1.1 potentiostat, nan instrument for automatically main-taining an electrode in an electrolyte at a constant potential orcontrolled potentials with respect to a suitable re
12、ferenceelectrode (see Terminology G15).3.1.2 potentiodynamic cyclic polarization (forward and re-verse polarization), na technique in which the potential ofthe test specimen is controlled and the corrosion currentmeasured by a potentiostat. The potential is scanned in thepositive or noble (forward)
13、direction as defined in Practice G3.The potential scan is continued until a predetermined potential1This test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current ed
14、ition approved Dec. 1, 2017. Published January 2018. Originallyapproved in 2001. Last previous edition approved in 2017 as F2129 17a. DOI:10.1520/F2129-17B.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of AST
15、MStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis int
16、ernational standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) C
17、ommittee.1or current density is reached. Typically, the scan is run until thetranspassive region is reached, and the specimen no longerdemonstrates passivity, as defined in Practice G3. The potentialscan direction is then reversed until the specimen repassivatesor the potential reaches a preset valu
18、e.3.1.3 scan rate, nthe rate at which the controlling voltageis changed.3.2 Symbols:3.2.1 Eb= Breakdown or Critical Pitting Potentialtheleast noble potential at which pitting or crevice corrosion orboth will initiate and propagate as defined in Terminology G15.An increase in the resistance to pittin
19、g corrosion is associatedwith an increase in Eb.3.2.2 Er= Rest Potentialthe potential of the workingelectrode relative to the reference electrode measured undervirtual open-circuit conditions (working electrode is not polar-ized).3.2.3 Ezc= Zero Current Potentialthe potential at whichthe current rea
20、ches a minimum during the forward scan.3.2.4 Ef= Final Potentiala preset potential at which thescan is stopped.3.2.5 Ei= Initial Potentialthe potential at which thepotentiostat begins the controlled potentiodynamic scan.3.2.6 Ep= Protection Potentialthe potential at which thereverse scan intersects
21、the forward scan at a value that is lessnoble than Eb. Epcannot be determined if there is nobreakdown. Whereas, pitting will occur on a pit-free surfaceabove Eb, it will occur only in the range of potentials betweenEpand Ebif the surface is already pitted. The severity ofcrevice corrosion susceptibi
22、lity increases with increasing hys-teresis of the polarization curve, the difference between EbandEp.3.2.7 Ev= Vertex Potentiala preset potential, at which thescan direction is reversed.3.2.8 it= Threshold Current Density (mA/cm2)a presetcurrent density, at which the scan direction is reversed.Typic
23、ally, the scan is reversed when a current density twodecades higher than the current density at the breakdownpotential (Eb) is reached.4. Summary of Test Method4.1 The device is placed in an appropriate deaerated simu-lated physiological solution, and the rest potential (Er)isrecorded for1hor,altern
24、atively, until the rest potentialstabilizes to a rate of change less than 3 mV/min. Thepotentiodynamic scan is then started at Erand scanned in thepositive or noble (forward) direction. The scan is reversed aftereither the vertex potential (Ev) is reached or the current densityhas reached a value ap
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM F2129 17 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices

链接地址:http://www.mydoc123.com/p-288040.html