ASTM D7094-17a Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester.pdf
《ASTM D7094-17a Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester.pdf》由会员分享,可在线阅读,更多相关《ASTM D7094-17a Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7094 17aStandard Test Method forFlash Point by Modified Continuously Closed Cup(MCCCFP) Tester1This standard is issued under the fixed designation D7094; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of the flashpoint of fuels including diesel/biodiesel blends, lube oils,solven
3、ts, and other liquids by a continuously closed cup testerutilizing a specimen size of 2 mL, cup size of 7 mL, with aheating rate of 2.5 C per minute.1.1.1 Apparatus requiring a specimen size of 1 mL, cup sizeof 4 mL, and a heating rate of 5.5 C per minute must be runaccording to Test Method D6450.1.
4、2 This flash point test method is a dynamic method anddepends on definite rates of temperature increase. It is one ofthe many flash point test methods available and every flashpoint test method, including this one, is an empirical method.NOTE 1Flash point values are not a constant physical chemicalp
5、roperty of materials tested. They are a function of the apparatus design,the condition of the apparatus used, and the operational procedure carriedout. Flash point can, therefore, only be defined in terms of a standard testmethod and no general valid correlation can be guaranteed between resultsobta
6、ined by different test methods or where different test apparatus isspecified.1.3 This test method utilizes a closed but unsealed cup withair injected into the test chamber.1.4 This test method is suitable for testing samples with aflash point from 35 C to 225 C.NOTE 2Flash point determinations below
7、 35 C and above 225 Cmay be performed; however, the precision has not been determined belowand above these temperatures.1.5 If the users specification requires a defined flash pointmethod other than this method, neither this method nor anyother test method should be substituted for the prescribed te
8、stmethod without obtaining comparative data and an agreementfrom the specifier.1.6 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard. Temperatures are in degrees Celsius, pressure inkilo-Pascals.1.7 This standard does not purport
9、to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.For specific warning statements, s
10、ee 7.2, 8.5, and 10.1.2.1.8 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organizat
11、ion TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D93 Test Methods for Flash Point by Pensky-MartensClosed Cup TesterD4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4177 Practice for Automatic Sampling of Petroleum andPetroleum ProductsD6
12、450 Test Method for Flash Point by Continuously ClosedCup (CCCFP) TesterD6708 Practice for Statistical Assessment and Improvementof Expected Agreement Between Two Test Methods thatPurport to Measure the Same Property of a MaterialE300 Practice for Sampling Industrial Chemicals2.2 ISO Standards:3ISO
13、Guide 34 Quality Systems Guidelines for the Produc-tion of Reference MaterialsISO Guide 35 Certification of Reference MaterialsGeneraland Statistical Principles3. Terminology3.1 Definitions:1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubr
14、icants and is the direct responsibility ofSubcommittee D02.08 on Volatility.Current edition approved Dec. 1, 2017. Published February 2018. Originallyapproved in 2004. Last previous edition approved in 2017 as D7094 17. DOI:10.1520/D7094-17A.2For referenced ASTM standards, visit the ASTM website, ww
15、w.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.a
16、nsi.org.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardizat
17、ion established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.13.1.1 dynamic, adjthe condition where the vapor abovethe test specimen and the test specimen are
18、 not in temperatureequilibrium at the time at which the ignition source is applied.3.1.2 flash point, nin flash point test methods, the lowesttemperature of the test specimen, adjusted to account forvariations in atmospheric pressure from 101.3 kPa, at whichapplication of an ignition source causes t
19、he vapors of the testspecimen to ignite under specified conditions of test.3.1.2.1 DiscussionFor the purpose of this test method, thetest specimen is deemed to have flashed when the hot flame ofthe ignited vapor causes an instantaneous pressure increase ofat least 20 kPa inside the closed measuring
20、chamber.4. Summary of Test Method4.1 The lid of the test chamber is regulated to a temperatureat least 18 C below the expected flash point.A2 mL 6 0.2 mLtest specimen of a sample is introduced into the sample cup.Both specimen and cup are at a temperature at least 18 Cbelow the expected flash point;
21、 cooled if necessary. The cup isthen raised and pressed onto the lid of specified dimensions toform the continuously closed but unsealed test chamber with anoverall volume of 7.0 mL 6 0.3 mL.4.2 After closing the test chamber, the temperatures of thetest specimen and the regulated lid are allowed to
22、 equilibrate towithin 1 C. Then the lid is heated at a prescribed, constantrate. For the flash tests, a high voltage arc of defined energy isdischarged inside the test chamber at regular intervals. Aftereach ignition, a variable amount of air (see Table 1)isintroduced into the test chamber to provid
23、e the necessaryoxygen for the next flash test. The pressure inside the continu-ously closed but unsealed test chamber remains at ambientbarometric pressure except for the short time during the airintroduction and at a flash point.4.3 After each arc application, the instantaneous pressureincrease abo
24、ve the ambient barometric pressure inside the testchamber is monitored. When the pressure increase exceeds20 kPa, the temperature at that point is recorded as theuncorrected flash point, which is then corrected to barometricpressure.5. Significance and Use5.1 The flash point temperature is one measu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM D7094 17 Standard Test Method for Flash Point by Modified Continuously Closed Cup MCCCFP Tester

链接地址:http://www.mydoc123.com/p-286974.html