ASTM D4962-2018 Standard Practice for NaI(Tl) Gamma-Ray Spectrometry of Water.pdf
《ASTM D4962-2018 Standard Practice for NaI(Tl) Gamma-Ray Spectrometry of Water.pdf》由会员分享,可在线阅读,更多相关《ASTM D4962-2018 Standard Practice for NaI(Tl) Gamma-Ray Spectrometry of Water.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D4962 17D4962 18Standard Practice forNaI(Tl) Gamma-Ray Spectrometry of Water1This standard is issued under the fixed designation D4962; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numbe
2、r in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the measurement of radionuclides in water by means of gamma-ray spectrometry. It is applicable tonuclides emitting gamm
3、a-rays with energies greater than 50 keV. For typical counting systems and sample types, activity levelsof about 40 Bq (1080 pCi) are easily measured and sensitivities of about 0.4 Bq (11 pCi) are found for many nuclides (1-10).2Count rates in excess of 2000 counts per second should be avoided becau
4、se of electronic limitations. High count rate samples canbe accommodated by dilution or by increasing the sample to detector distance.1.2 This practice can be used for either quantitative or relative determinations. In tracer work, the results may be expressed bycomparison with an initial concentrat
5、ion of a given nuclide which is taken as 100 %. For radioassay, the results may be expressedin terms of known nuclidic standards for the radionuclides known to be present. In addition to the quantitative measurement ofgamma-ray activity, gamma-ray spectrometry can be used for the identification of s
6、pecific gamma-ray emitters in a mixture ofradionuclides. radionuclides but that ability is limited when using low energy resolution Na(Tl) detectors as compared to HighPurity Germanium (HPGe) detectors. General information on radioactivity and the measurement of radiation has been published(11 and 1
7、2). Information on specific application of gamma-ray spectrometry is also available in the literature (13-16).1.3 The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are included forinformation only and are not considered standard.1.4 This sta
8、ndard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability ofregulatory limitations prior to use.1.5 This inte
9、rnational standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Com
10、mittee.2. Referenced Documents2.1 ASTM Standards:3D1129 Terminology Relating to WaterD3648 Practices for the Measurement of RadioactivityD7902 Terminology for Radiochemical AnalysesE181 Test Methods for Detector Calibration and Analysis of Radionuclides3. Terminology3.1 Definitions:3.1.1 For definit
11、ions of terms used in this standard, refer to Terminologies D1129 and D7902.4. Summary of Practice4.1 Gamma-ray spectra are commonly measured with modular equipment consisting of a detector, amplifier, analog-to-digitalconverter, multi-channel analyzer device, and a computer (17 and 18).1 This pract
12、ice is under the jurisdiction ofASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.04 on Methods of RadiochemicalAnalysis.Current edition approved Nov. 1, 2017Oct. 1, 2018. Published November 2017November 2018. Originally approved in 1989. Last previous edition approved
13、 in 20092017as D4962 02 (2009).D4962 17. DOI: 10.1520/D4962-17.10.1520/D4962-18.2 The boldface numbers in parentheses refer to the references at the end of this practice.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annua
14、l Book of ASTM Standardsvolume information, refer to the standardsstandards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may
15、 not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbo
16、r Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.2 Thallium-activated sodium-iodide crystals, NaI(Tl), which can be operated at ambient temperatures, are often used asgamma-ray detectors in spectrometer systems. However, their energy resolution limits their use to the analysis
17、 of single nuclidesor simple mixtures of a few nuclides. A resolution of about 7 % (45 keV full width at one half the 137Cs peak height) at 662 keVcan be expected for a NaI(Tl) detector in a 76 mm by 76 mm-configuration. There are solid scintillators such as cerium dopedLaBr3 that may provide a perf
18、ormance advantage over NaI(Tl) in terms of energy resolution but whose suitability should beevaluated and documented before being considered as a substitute for NaI(Tl).4.3 Interaction of a gamma-ray with the atoms in a NaI(Tl) detector results in light photons that can be detected by aphotomultipli
19、er tube (PMT). The output from the PMT and its preamplifier is directly proportional to the energy deposited by theincident gamma-ray. These current pulses are fed into an amplifier of sufficient gain to produce voltage output pulses in theamplitude range from 0 to 10 V.4.4 A The combination of an a
20、nalog-to-digital converter and multichannel pulse-height analyzer is used to determine theamplitude of each pulse originating in the detector, and accumulates in a memory the number of pulses in each amplitude band(or channel) in a given counting time (17 and 18). For a 0 to 2 MeV spectrum two hundr
21、ed channels may be adequate but mostcurrent systems provide a thousand or more channels.4.5 The distribution of the amplitudes (pulse heights) of the pulse energies, represented by the pulse height, can be separatedinto two principal components. One of these components has a nearly Gaussian distribu
22、tion and is the result of total absorptionof the gamma-ray energy in the detector; this peak is normally referred to as the full-energy peak or photopeak. The othercomponent is a continuous one, lower in energy than the photopeak.This continuous curve is referred to as the Compton continuumand resul
23、ts from interactions wherein the gamma photons lose only part of their energy to the detector.4.6 Other peaks components, such as escape peaks, backscattered gamma-rays, or X-rays from shields, are often superimposedon the Compton continuum. These portions of the curve are shown in Fig. 1 and Fig. 2
24、.4.7 Escape peaks will be present when gamma-rays with energies greater than 1.02 MeV are emitted from the sample (19-24).The positron formed in pair production is usually annihilated in the detector and one or both of the 511 keV annihilation quantaFIG. 1 Compton ContinuumD4962 182may escape from t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD49622018STANDARDPRACTICEFORNAITLGAMMARAYSPECTROMETRYOFWATERPDF

链接地址:http://www.mydoc123.com/p-1243634.html