VA 11 53 13-2011 LABORATORY FUME HOODS.pdf
《VA 11 53 13-2011 LABORATORY FUME HOODS.pdf》由会员分享,可在线阅读,更多相关《VA 11 53 13-2011 LABORATORY FUME HOODS.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、01-11 11 53 13 - 1 SECTION 11 53 13 LABORATORY FUME HOODS SPEC WRITER NOTES: 1. Delete between / _ / if not applicable to project. 2. Delete other items or paragraphs in the section that are not applicable and renumber the paragraphs. PART 1 - GENERAL 1.1 DESCRIPTION This section includes chemical (
2、general purpose) hoods, radioisotope hoods, and Perchloric acid hoods in designs and configurations specified hereunder. 1.2 DEFINITIONS A. CHEMICAL OR GENERAL-PURPOSE HOODS 1. AUXILIARY-AIR HOOD: A hood with a dedicated duct that supplies up to 50% outside air to the face of the hood. 2. BYPASS HOO
3、D: A hood that contains openings above the sash and below the airfoil, which redistribute the air to reduce fluctuations in face velocity and turbulence within the hood, when the sash is re-positioned. 3. 100% RE-CIRCULATORY (DUCTLESS) FUME HOODS: A ductless, self-contained, carbon and/or HEPA-filte
4、red enclosure, that draws air into the front of the hood, passes it through multi-stage filters, and then re-circulates it into the lab. These hoods can be placed onto fixed work surfaces or on mobile stands. 4. LOW-VELOCITY HOOD: A high-performance hood that yields energy savings by reducing the sa
5、sh opening and the corresponding exhaust volume, while maintaining safe containment levels with the sash raised for set-up and face velocity as low as 60fpm. B. Radioisotope Hood: A hood that is used for radioactive applications, with a stainless steel liner and an integral work surface. C. PERCHLOR
6、IC ACID HOOD: A hood that is dedicated for use with Perchloric Acid only, due to the potential reaction of the acid with other chemicals. The hood, duct and fan must have a water wash-down system for removal of perchlorate crystals (salts) after use. 1.3 RELATED WORK A. Section 22 15 00, GENERAL SER
7、VICE COMPRESSED-AIR SYSTEMS: Connections to Compressed Air System. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-01-11 11 53 13 - 2 B. Section 22 11 00, FACILITY WATER DISTRIBUTION: Plumbing Connections. C. Section 22 13 00, FACILITY SANITARY SEWER
8、AGE: Plumbing Connections. D. Section 22 66 00, CHEMICAL-WASTE SYSTEMS FOR LABORATORY and HEALTHCARE FACILITIES: Plumbing Connections. E. Section 22 62 00, VACUUM SYSTEMS FOR LABORATORY and HEALTHCARE FACILITIES: Connections to Gas and Vacuum Systems. F. Section 22 63 00, GAS SYSTEMS FOR LABORATORY
9、and HEALTHCARE FACILITIES: Connections to Gas and Vacuum Systems. G. Section 23 05 11, COMMON WORK RESULTS FOR HVAC. H. Section 23 05 12, GENERAL MOTOR REQUIREMENTS FOR HVAC and STEAM GENERATION: Integral Blowers on Hoods. I. Section 23 09 23, DIRECT-DIGITAL CONTROL SYSTEM FOR HVAC: Face Velocity Se
10、nsor Controller. J. Section 23 34 00, HVAC FANS. K. Section 23 36 00, AIR TERMINAL UNITS: Airflow Control Valves. L. Section 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS: Electrical Connections. M. Section 26 27 26, WIRING DEVICES: Electrical Devices. N. Section 26 29 11, LOW-VOLTAGE MOTOR ST
11、ARTERS: Motor Starters. 1.4 PERFORMANCE REQUIREMENTS A. Average Face Velocity for general chemical and ductless laboratory fume hoods: 0.51 m/s (100 fpm) with sash located at the average sash stop of 18 inches. B. Average Face Velocity for Low Velocity laboratory fume hoods can be as low as 60 fpm w
12、ith sash located at the average Low Velocity hood sash stop height of 12 to 18 inches. C. Containment: Furnish and install laboratory fume hoods that are tested according to ASHRAE 110 at a release rate of 4.0 L/min. (0.1 CFM) (VA Comment: Provide equivalent inch/pound unit.) 1. Face Velocity Variat
13、ion: Allowable VA variation threshold is not more than +-10% of average face velocity. SPEC WRITER NOTE: Size hood exhaust system for sash in the fully open position for the specified static pressure losses. 2. Sash Position: a. For horizontal-sash units, test with maximum opening on one side, with
14、maximum opening in the center, and with opening at each side equal to half of maximum opening. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-01-11 11 53 13 - 3 b. For combination-sash units, test with sash fully raised, with maximum opening on one
15、side, with maximum opening in the center, and with opening at each side equal to half of maximum opening. c. For Vertical Sash units, test with sash fully open, and at typical sash stop height of 18”. SPEC WRITER NOTE: Options for AM and AI rating requirements included below are examples only; verif
16、y requirements for project. d. VA Comment: Why vertical sash hoods not mentioned? 3. As-Manufactured (AM) Rating: AM / 0.05 (0.05 ppm) / _ /. 4. As-Installed (AI) Rating: AI / 0.05 (0.05 ppm) / _ /. D. Average Static-Pressure Loss: Not more than 93 Pa (3/8-inch wg) at 0.51-m/s (100-fpm) face velocit
17、y when tested according to SEFA 1.2. 1.5 QUALITY CONTROL A. Mechanical, electrical and associated systems shall be safe, reliable, efficient, durable, easily and safely operable, maintainable, and accessible. B. Standard Products: Material and equipment shall be the standard products of the selected
18、 manufacturer, and they should be regularly engaged in the manufacture of such products for at least 3 years. C. All items furnished shall be free from defects that would adversely affect the performance, maintainability and appearance of individual components and overall assembly. D. Standard Produ
19、cts: Material and equipment shall be the standard products of a manufacturer regularly engaged in the manufacture of the products for at least 3 years. The design, model and size of each item shall have been in satisfactory and efficient operation on at least three installations for approximately th
20、ree years. However, digital electronics devices, software and systems such as controls, instruments, computer work station, shall be the current generation of technology and basic design that has a proven satisfactory service record of at least three years. See other specification sections for any e
21、xceptions. E. Electrical Components and Devices: UL listed and labeled for intended use. 1.6 SUBMITTALS A. Submit in accordance with Section 01 33 23, SHOP DRAWINGS, PRODUCT DATA, AND SAMPLES. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-01-11 11
22、53 13 - 4 B. Manufacturers Literature and Data: Include the following: 1. Illustrations and descriptions of laboratory fume hoods and factory-installed devices for fume hoods. 2. Catalog or model numbers for each item incorporated into the work. 3. Static-pressure losses and exhaust volumes for fume
23、 hoods. 4. Results of testing according to ASHRAE 110. C. Shop Drawings: Show details of fabrication, installation, adjoining construction, coordination with mechanical and electrical work, anchorage, and other work required for complete installation. D. Field Test Reports: Indicate dates and times
24、of tests and certify test results. E. Factory Test Reports: Provide manufacturers QC checklist or other reports that indicate comprehensive factory testing has been performed, and the results of these tests have been certified. F. Operating Instructions: Comply with requirements in Section 01 00 00,
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- VA1153132011LABORATORYFUMEHOODSPDF