UOP 927-1992 TRACE METALLIC IMPURITIES IN FRESH CATALYSTS BY ICP-AES.pdf
《UOP 927-1992 TRACE METALLIC IMPURITIES IN FRESH CATALYSTS BY ICP-AES.pdf》由会员分享,可在线阅读,更多相关《UOP 927-1992 TRACE METALLIC IMPURITIES IN FRESH CATALYSTS BY ICP-AES.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、TRACE METALLIC IMPURITIES IN FRESH CATALYSTSBY ICP-AESUOP Method 927-92SCOPEThis method is for determining trace levels of iron, sodium, calcium, magnesium, titanium, molybdenumand silicon in fresh alumina supported catalysts of known composition by Inductively Coupled Plasma-Atomic Emission Spectro
2、metry (ICP-AES). All these elements except silicon can also be determined infresh silica-alumina supported catalysts. The method is sufficiently flexible to permit the determination ofother metals at trace levels if they can be brought into solution using the sample preparation proceduredescribed. C
3、oncentration ranges covered are listed in Table 1. Catalysts containing alpha- or theta-aluminaphases cannot be analyzed by this method.OUTLINE OF METHODAlumina supported catalysts are dissolved using sulfuric acid, while silica-alumina supported catalystsare dissolved in sulfuric and hydrofluoric a
4、cids. The resulting sample solution is aspirated into an ICP-AESinstrument and the concentration of each element is determined by comparison to matrix-matched standardsolutions. Sample results are reported on an as-received basis.APPARATUSReferences to catalog numbers and suppliers are included as a
5、 convenience to the method user. Othersuppliers may be used.Balance, readability 0.1-mgBeaker, Teflon, 250-mLBeaker, borosilicate glass, one-liter, used for warm water bathBottles, polypropylene, 125-mL and one-liter, Fisher Scientific, Cat. Nos. 02-925C and 925F,respectivelyCylinder, graduated, pol
6、ypropylene, 10-, 25- and 50-mLIS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TODETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH ANDSAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THISPROCEDURE
7、 IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS(MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED INTHIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTIONEQUIPMENT (PPE). COPYRIGHT 1992 UOP LLCALL RI
8、GHTS RESERVEDUOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service atserviceastm.org, 610.832.9555 FAX, or 610.832.
9、9585 PHONE.2 of 7927-92Flasks, volumetric, polypropylene, 100-, 200- and 1000-mLFunnel, disposable, 55-mm diameter, Fisher Scientific, Cat. No. 10-320AGloves, polyethyleneHot plate, variable heat to at least 400CInductively Coupled Plasma-Atomic Emission Spectrometer, computer controlled, having a b
10、andpass of0.010 nm (FWHM) or better, and covering the wavelength range specified in Table 1. The instrumentmust have the ability to use internal standards and perform background corrections, and must measurethe internal standard response concurrently with each element response. A Perkin-Elmer Plasma
11、 2000spectrometer, or its equivalent, is required. A corrosion resistant sample introduction system isrecommended.Pipets, total volume, Class A, 5-, 10-, 15-, 20-, 25-, 30- and 50-mL, several of each requiredRegulator, argon, two-stage, high purity, Matheson Gas Products, Model 3104-580Regulator, ni
12、trogen, two-stage, high purity, Matheson Gas Products, Model 3104-580Thermometer, 0 to 100C, for use with warm water bath, Fisher Scientific, Cat. No. 15-077CThermometer, electronic, with surface temperature probe, Omega Engineering, Cat. Nos. HH81 and88108K, respectivelyWatch glass, Teflon, 75-mmWa
13、ter purification system, to provide 18.3 megohm/cm, Type 1 Reagent Grade water, Fisher Scientific,Cat. No. 09-050-256REAGENTS AND MATERIALSAll reagents shall conform to the specifications established by the Committee on Analytical Reagents ofthe American Chemical Society, when such specifications ar
14、e available, unless otherwise specified.References to water mean 18.3 megohm/cm, Type 1 Reagent Grade water.References to catalog numbers and suppliers are included as a convenience to the method user. Othersuppliers may be used.Aluminum metal, shot, 99.9999% minimum purity, Johnson Matthey, Alfa Pr
15、oducts, Cat. No. 10574Argon, 99.995% minimum purity, ICP feed gasHydrochloric acid, concentrated, high purity, Baker Ultrex gradeHydrofluoric acid, 48%Hydrogen peroxide, 30%Mercury metal, 99.99999% purity, SPEX Industries, Cat. No. HG05-50Nitrogen, 99.99% minimum purity, water/oil free, ICP optics p
16、urge gasScandium solution, 1000-g/mL, Inorganic Ventures, Cat. No. CGSC1-53 of 7927-92Standard solutions of metals, Fe, Na, Ca, Mg, Ti, Mo and Si solutions of 1000-g/mL concentrations,Inorganic Ventures, Cat. Nos. CGFE1-5, CGNA1-5, CGCA1-5, CGMG1-5, CGTI1-5, CGMO1-5 andCGSI1-5, respectivelySulfuric
17、acid, concentrated, high purity, Baker Ultrex GradeSulfuric acid, 80%. Mix four parts concentrated sulfuric acid with one part water, adding the acid to thewater slowly to avoid spattering.Yttrium solution, 1000-g/mL, Inorganic Ventures, Cat, No. CGY1-5. Not required if the spectrometerspecified in
18、APPARATUS is used (see Note 1).PROCEDUREAll subsequent steps involving handling and heating of concentrated acids or acid solutions must beperformed in a fume hood.Preparation of 10,000-g/mL Aluminum Solution1. Weigh 10.0 0.1 g of pure aluminum metal to the nearest 0.1 mg, and transfer to a one-lite
19、rpolypropylene bottle.2. Add one drop of mercury metal and 50 mL of concentrated high purity hydrochloric acid to the bottle.3. Cap the bottle, but leave the cap loose to allow pressure release.4. Place the bottle in a warm water bath (40 to 50C) to start the reaction, and adjust the reaction rate t
20、omaintain an approximate 25-mm layer of bubbles (foam) above the liquid. If the layer exceeds thislevel, control by adding a little water to the bottle. As it decreases, add hydrochloric acid dropwiseuntil a total of 100 mL (50 mL from Step 2 plus 50 mL additional) has been added. The dissolutiontak
21、es several hours.5. Transfer the solution to a one-liter polypropylene volumetric flask, quantitatively with water, using afunnel, after dissolution is complete. The remaining mercury is also transferred to the bottle. After the aluminum solution has been consumed,dispose of the mercury by following
22、 appropriate statutory guidelines.6. Cool to ambient temperature and dilute to volume. Cap and invert several times to mix thoroughly.Store in a one-liter polypropylene bottle.Preparation of Calibration Standards1. Prepare standards at the concentrations shown in Table 1 by appropriate dilution of t
23、he 1000-g/mLstandard metal solutions, using the high purity hydrochloric acid. Silicon is not added to Standard 2when determining impurities in silica-alumina supported catalysts. Two standards are used for Ca to cover two different concentration ranges using two differentwavelengths. The standards
24、are prepared in the same matrix as the samples, matching the acid and aluminum levels inthe final sample dilution, see Sample Preparation. If a spectrometer other than that specified in APPARATUS is used, yttrium can be substituted as theinternal standard (see Note 1).4 of 7927-922. Prepare a calibr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9271992TRACEMETALLICIMPURITIESINFRESHCATALYSTSBYICPAESPDF

链接地址:http://www.mydoc123.com/p-1074571.html