SAE AS 1197A-1998 Continuous Flow Oxygen Regulator《连续流氧气调节器》.pdf
《SAE AS 1197A-1998 Continuous Flow Oxygen Regulator《连续流氧气调节器》.pdf》由会员分享,可在线阅读,更多相关《SAE AS 1197A-1998 Continuous Flow Oxygen Regulator《连续流氧气调节器》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、_ SAE Technical Standards Board Rules provide that: “This report is published by SAE to advance the state of technical and engineering sciences. The use of this report is entirely voluntary, and its applicability and suitability for any particular use, including any patent infringement arising there
2、from, is the sole responsibility of the user.” SAE reviews each technical report at least every five years at which time it may be revised, reaffirmed, stabilized, or cancelled. SAE invites your written comments and suggestions. Copyright 2013 SAE International All rights reserved. No part of this p
3、ublication may be reproduced, stored in a retrieval system or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of SAE. TO PLACE A DOCUMENT ORDER: Tel: 877-606-7323 (inside USA and Canada) Tel: +1 724-776-497
4、0 (outside USA) Fax: 724-776-0790 Email: CustomerServicesae.org SAE WEB ADDRESS: http:/www.sae.org SAE values your input. To provide feedback on this Technical Report, please visit http:/www.sae.org/technical/standards/AS1197A AEROSPACE STANDARD AS1197 REV. A Issued 1970-11 Revised 1998-12 Reaffirme
5、d 2013-12 Superseding AS1197 Continuous Flow Oxygen Regulator RATIONALE AS1197A has been reaffirmed to comply with the SAE five-year review policy. 1. SCOPE:This standard covers regulators of the following types:Type I - Automatic Continuous FlowType II - Adjustable Continuous FlowType III - Pre-set
6、 Continuous FlowClass A - Cylinder MountedClass B - Line MountedSpecial - Composite Continuous Flow (Appendix)1.1 Purpose:To establish requirements for construction, performance and testing of continuous flow oxygen regulators. This document supersedes in part AS463, 12-15-56, which is cancelled.FOR
7、EWORDChanges in this revision are format/editorial only.2. REFERENCES:AIR825 Oxygen Equipment for AircraftAS861 Minimum General Standards for Oxygen SystemsAS1046 Portable Oxygen EquipmentMIL-E-5272CMIL-O-27210I.C.C. Regulations, Tariff 19, Section 173.302Compressed Gas Association Pamphlet S-1, Par
8、t I, Paragraph 5.3.33. DESCRIPTION:The regulators described herein are intended to perform as oxygen supply regulators in aircraft supplemental oxygen breathing systems. The output of the regulator is delivered to a suitable sized outlet orifice at the pressures prescribed herein.4. DEFINITION:The r
9、egulator types are defined as follows:Type I - This type of regulator provides an output flow or pressure which varies automatically with cabin pressure to insure adequate delivery to the mask.Type II - This type of unit may be manually adjusted to deliver the proper amount of pressure or flow for a
10、 selected cabin pressure.Type III - This regulator has a fixed outlet flow or pressure which has been pre-set to provide adequate flow up to indicated cabin pressure.Class A - The regulators in this class are designed to be cylinder mounted.Class B - The regulators in this class are intended for use
11、 in a line between the oxygen source and the dispensing unit or on an oxygen cylinder valve.SAE INTERNATIONAL AS1197A Page 2 of 8_5. REQUIREMENTS:5.1 General:The requirements set forth in AS861, “Minimum General Standards for Oxygen Systems” and AS1046, “Portable Oxygen Equipment,” shall be consider
12、ed as part of this standard, except that should there be any conflict the requirements of this standard shall take precedence.5.2 Materials:Materials shall conform to applicable specifications, shall be of the best quality and shall be suitable for the purpose intended. Materials that are easily det
13、eriorated or otherwise affected adversely by continued service when used with oxygen shall not be used.5.2.1 Metals: Metals shall be of the corrosion-resistant type or suitably treated to resist corrosion due to salt spray, or atmospheric conditions likely to be met in storage or normal service life
14、 of the regulator.5.2.2 Elastomers: All elastomers used shall be manufactured within 24 months or less of the date of regulator manufacture. The cure date (month and year) shall be stamped on the elastomer component providing there is sufficient area. The figures shall be at least 3/16 inch high. Cu
15、re dates and age limitation shall not apply to silicone. The cure date of the oldest elastomers shall be stamped on the outside of the regulator.5.2.3 Protective Treatment: Materials which are used in the construction of the regulator that are subject to corrosion in oxygen, salt air, or other atmos
16、pheric conditions likely to occur during service usage, shall be protected against such corrosion in a manner that will in no way prevent compliance with the performance requirements of this Specification.5.3 Design:5.3.1 Class A Regulators: Class A regulators are intended to be mounted directly on
17、the oxygen cylinder and in addition to the pressure regulating characteristics shall include the following provisions:5.3.1.1 On-Off Control: A means shall be provided for manually turning the flow on or off. Both positions shall be clearly and permanently marked.5.3.1.2 Cylinder Contents Indicator:
18、 A pressure gage or other suitable means shall be provided to give continuous indication of cylinder contents.5.3.1.3 Charging: A self-closing charging valve or other suitable means shall be provided for charging the oxygen cylinder.SAE INTERNATIONAL AS1197A Page 3 of 8_5.3.1.4 Wrench Flats: The val
19、ve body shall have adequate wrench flats and be sturdy enough to withstand removal from the cylinder and attachment of outlet fittings.5.3.1.5 Standpipe: A standpipe shall be provided in the valve inlet to minimize the possibility of moisture and loose foreign matter from entering the regulator when
20、 the cylinder is inverted.5.3.1.6 Safety Outlet: A safety outlet shall be provided in accordance with I.C.C. Regulations, Tariff 19, Section 173.302. The safety outlet shall be the frangible disc type with no fusible metal and shall be designed to provide force equilibrium while relieving.5.3.2 Clas
21、s B Regulators: Class B regulators are intended to be mounted in line downstream of a cylinder and valve or other suitable oxygen source. Its function is simply one of pressure regulation.5.3.3 Oxygen Inlet Filter: A filter capable of filtering materials approximately 25 microns in size shall be pro
22、vided in the oxygen inlet port and filler valve port where applicable on all regulators.5.3.4 Relief Valve: A relief valve shall be provided to protect the regulator outlet from over-pressurization on all regulators.5.3.5 Type II Regulator: A means shall be provided to indicate setting of the manual
23、 control. Such means shall be clearly marked for proper cabin pressure setting.5.3.6 Type III Regulator: This regulator shall be clearly and permanently marked to show maximum altitude at which the regulator will provide sufficient oxygen to meet physiological requirements.5.4 Performance:The regula
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- SAEAS1197A1998CONTINUOUSFLOWOXYGENREGULATOR 连续流 氧气 调节器 PDF

链接地址:http://www.mydoc123.com/p-1023425.html