1、Q ri A-A-54037 65 m 9399774 0053793 B A-A-54 O 3 9 30 March 1990 COMMERCIAL ITEM DESCRIPTION SUCTION APPARATUS, OROPHARYNGEAL Tracheal, Aeromedical, Portable/Dual Voltage The General Services Administration has authorized , the use of this Commercial Item Description. This Commercial Item Descriptio
2、n covers a portable, battery operated, oropharyngeal suction apparatus, 120 volts, 50/60/400 Hz, AC, with separate stepdown transformer for operation on 230 volts, 50/60 Ha, AC. Salient characteristics: Suction apparatus. oropharyngeal/tracheal suction apparatus. It shall be self- contained in a car
3、rying case complete with a rechargeable battery, suitable for use in hospitals, ambulances, and field uses with no external power available, and in miitary aeromedical aircrafts. It shall be suitable for providing aspiration of secretions, blood or vomitus.duning emergency treatment of an unconsciou
4、s or injured patient. The suction apparatus shall be a portable, e Major components. The suction apparatus shall be provisioned with the following integrated-components: vacuum pump and motor, battery charger, internal battery, interlocking switch(es1, indicating lamps, circuit breaker, EM1 and RFI
5、shielding circuitry, collection container, patient suction tube, tracheal catheters, AC power cord, vehicle power cord, vacuum valve, vacuum gauge, rinse bottle, litter mounting attachment, and separate transformer Vacuum pump and motor. The design of the punip-motor shall be such that overflow of p
6、atient debris shall not jam or clog it during usage. It shall be readily cleanable for next use. Vacuum range. The vacuum capacity shall include an adjustable range from O to at least 500 mmHg. The high end of this range shall be adjustable by a control valve which regulates the pump output vacuum.
7、Using the internal battery, the unit shall be capable of maintaining the specified vacuum for at AMSC N/A FSC 6515 distribution is unlimited. DISTRIBUTION STATEMENT A. Approved for public release; Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5
8、4 O 3 9 least 30 minutes. This requirement shall be tested in accordance with ASTM F960 using the internal battery, the 120 amp hour external battery, and using 110 volts AC at 50, 60, and 400 Hz. It shall also be tested using its internal battery in accordance with test C, “Vacuum Capacity Using Ba
9、ttery“ stated herein. Flow rate. The free airflow at the patient end of the suction tube shall be at least 30 LPM using any of the specified power sources (see paragraph titled “Power Sources“) and operating at specified temperature extremes (see paragraph titled “Operating Temperature Range“). This
10、 requirement shall be tested in accordance with ASTM F960 using the internal battery, the 120 amp hour externa battery, and 110 volts at 50, 60 and 400 Hz. Battery charger. This apparatus shall be provisioned with a battery charger, operational from nominal 110-125 volts, 50/60 Hz. It shall be capab
11、le of recharging the totally depleted internal battery within 24 hours. Internal battery. The apparatus shal be equipped with a 12 volt internal battery capable of operating the unit in accordance with the requirements specified herein. (Multiple batteries wired to provide 12 volts DC is acceptable)
12、. Totally depleted internal battery shall be rechargeable within 24 hours. This requirement shall be tested in accordance with test B, “Recharge Time“ stated herein. Internal battery operating time. The battery operating time, with vacuum adiusted at 120 mmHq, shal be at least 30 minutes when cycled
13、 thiough 5 cycles of-6 minutes each. cycle shall have periods of a 2 minute period with patient tube occluded, a 1 minute period with patient tube open (free air flow), a 2 minute period with patient tube occluded, and a 1 minute period with the pump turned off (rest period). This requirement shall
14、be tested in accordance with test C, “Fully Charged Battery Operating Time“ stated herein. Interlocking switch(es1. with interlocking switch(es1 to seect “OFF/CHARGE“, “AC/ON“, “EXT. DC/ON“, and “INT. BAT/ON“. Switch arrangement shall not allow inappropriate selection of circuitry, causing damage to
15、 apparatus. Indicating lamps. The apparatus shall contain separate indicating lamps, indicating “PUMP ON“ and “CHARGER ON“. Each The apparatus shall be provisioned 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-540 3 9 Electromaqnetic interfer
16、ence (EMI) and radio frequency interference (RFI). The apparatus shall be designed to control electromagnetic interference (EMI) and radio frequency - interference (RFI) aircraft (including associated ground - support equipment and portable medical equipment used for aeromedical airlift). Proof of c
17、ompliance shall be accomplished by aeromedical testing of first article sample dedicated solely for these tests. Collection container. The collection container shall have a minimum capacity of 500 ML. The container shall meet the testing reguirements of ASTM F960. Patient suction tube. The apparatus
18、 shall be provisioned with a patient suction tube not less than 45 inches long. Tracheal suction catheters. The apparatus shall be provisioned with 2 catheters, both with thumb vacuum control to adjust vacuum level. Two sizes are required, one 14 FR and the other 18 FR. AC power cord. The apparatus
19、shall be provisioned with a power cord at least 6 feet long, with hospital grade male Vehicle power cord. The apparatus shall include a 6 foot electric cord assembly to obtain power from external vehicle DC source. Plug . Vacuum valve. The apparatus shall be provisioned with a control valve (in addi
20、tion to catheter thumb control), which shall regulate the high end of vacuum level. Vacuum gauqe. The apparatus shall be provisioned with a vacuum gauge calibrated both in millimeters and inches of mercury. It shall include the range of O to 500 mmHg . Rinse bottle. A rinse bottle shall be provided
21、with at least 250 ml capacity. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-_l_l_ A-A-54037 65 m 7777774 0053774 3 W A-A-5 4 039 Litter mountinq attachment. The self-contained apparatus shall include a litter mounting attachment consisting of fo
22、ur D-rings, two attached to the case front and two attached to I the case rear, and two 24 inch long hook and pile/nylon webbing attachment straps. The straps shall be shipped in the carrying case. Separate stepdown transformer. The unit shall include a separate stepdown transformer as a separate co
23、mponent not required to be in the carrying case, but shipped in the same shipping package. The transformer shall accept 220-250 volts, 50/60 Hz and provide one half output voltage for the suction apparatus. The transformer shall include a hospital grade, 110 volt receptacle to accept the suction app
24、aratus specified male line cord. The transformer shall include 3 posts in a compartment with a wire strain-relief attachment for SJT-16/3 wire. The three posts shall be wired to accept any worldwide wiring scheme required to connect a cord for 220-250 volts, single phase AC service. The separate ste
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA540391990SUCTIONAPPARATUSOROPHARYNGEALTRACHEALAEROMEDICALPORTABLEDUALVOLTAGE 便携式 电压 航空 医学 管口 吸力

链接地址:http://www.mydoc123.com/p-701707.html