DIN 6868-157-2014 Image quality assurance in diagnostic X-ray departments - Part 157 X-ray ordinance acceptance and constancy test of image display systems in their environment《X射线.pdf
《DIN 6868-157-2014 Image quality assurance in diagnostic X-ray departments - Part 157 X-ray ordinance acceptance and constancy test of image display systems in their environment《X射线.pdf》由会员分享,可在线阅读,更多相关《DIN 6868-157-2014 Image quality assurance in diagnostic X-ray departments - Part 157 X-ray ordinance acceptance and constancy test of image display systems in their environment《X射线.pdf(56页珍藏版)》请在麦多课文档分享上搜索。
1、November 2014DEUTSCHE NORM English price group 20No part of this translation may be reproduced without prior permission ofDIN Deutsches Institut fr Normung e. V., Berlin. Beuth Verlag GmbH, 10772 Berlin, Germany,has the exclusive right of sale for German Standards (DIN-Normen).ICS 11.040.50!%Q“24695
2、92www.din.deDIN 6868-157Image quality assurance in diagnostic Xray departments Part 157: Xray ordinance acceptance and constancy test of image display systems in their environment,English translation of DIN 6868-157:2014-11Sicherung der Bildqualitt in rntgendiagnostischen Betrieben Teil 157: Abnahme
3、 und Konstanzprfung nach RV an Bildwiedergabesystemen in ihrer Umgebung,Englische bersetzung von DIN 6868-157:2014-11Assurance de la qualit des images en diagnostic radiographique Partie 157: Test dacceptance et contrle de constance selon RV des systmes de visualisation des images dans leur environn
4、ement,Traduction anglaise de DIN 6868-157:2014-11SupersedesDIN V 686857:200102www.beuth.deDocument comprises 56 pagesDDIN-Normenausschuss Radiologie (NAR)DIN-Sprachendienst08.16 DIN 6868-157:2014-11 2 A comma is used as the decimal marker. Content Page Foreword . 4 1 Scope 6 2 Normative references 6
5、 3 Terms and definitions . 7 4 Symbols and abbreviations . 13 5 Test conditions . 14 6 Test equipment 15 6.1 General 15 6.2 Luminance meter . 15 6.2.1 Luminance meter for the acceptance test . 15 6.2.2 Luminance meter for the constancy test . 15 6.3 Illuminance meter / luxmeter 15 6.4 Test images 16
6、 7 Room classes . 17 8 Acceptance test for image display systems with diagnostic quality . 19 8.1 Scope of testing . 19 8.2 Visual tests . 20 8.2.1 General 20 8.2.2 Overall image quality . 20 8.2.3 Greyscale resolution of the IMAGE DISPLAY DEVICE . 22 8.2.4 LUMINANCE uniformity (visual) . 23 8.2.5 C
7、olour impression and uniformity 23 8.2.6 Pixel errors 24 8.2.7 Image geometry 24 8.2.8 Test based on clinical reference images . 25 8.2.9 Additional tests for mammography image display systems 25 8.3 Metrological tests 25 8.3.1 Minimum luminance 25 8.3.2 Maximum luminance 25 8.3.3 Ambient luminance
8、. 25 8.3.4 Illuminance 25 8.3.5 Maximum luminance ratio . 25 8.3.6 Uniformity of the image display device . 26 8.3.7 Uniformity of multiple image display devices 27 8.3.8 Luminance response curve 27 8.4 Minimum requirements (without mammography) . 28 8.5 Additional requirements for image display sys
9、tems used for mammography . 30 8.6 Mobile systems . 31 9 Reference values for the constancy test 31 10 Constancy test for image display systems with diagnostic quality . 31 DIN 6868-157:2014-11 3 11 Significant changes to an image display system . 33 12 Image display systems with review quality . 33
10、 Annex A (normative) Methods of measuring luminance 34 A.1 General 34 A.2 Measurement methods 34 A.2.1 Method A: Telescopic method . 34 A.2.2 Method B: Close-range luminance meter in combination with a luxmeter . 35 A.2.3 Method C: Front-integrated luminance meter in combination with a luxmeter . 36
11、 A.2.4 Method D: Back-integrated luminance meter in combination with an illuminance meter/luxmeter 37 A.3 Notes on the measuring methods 37 Annex B (informative) Typical errors 38 Annex C (informative) Test report templates . 39 C.1 Example of a test report template for the acceptance test . 39 C.2
12、Example of a test report template for the constancy test . 50 C.3 Example of a test report template for the daily constancy test 52 Bibliography . 53 Keyword index. 55 Figures Figure 1 Test images for use on image display systems (images a) through f) according to DIN EN 62563-1:2014-01, Table 2, an
13、d images g) through l) according to AAPM). 16 Figure 2 Overall image quality test using test image TG18-OIQ 22 Figure 3 Enlarged view of test image TG18-MP, depicting the 8-bit and 10-bit markers 23 Figure 4 Positions of the measuring points for the luminance uniformity test . 26 Figure A.1 Method A
14、 Telescopic method . 34 Figure A.2 Method B Close-range luminance meter in combination with a luxmeter 35 Figure A.3 Method C Front-integrated luminance meter in combination with luxmeter . 36 Figure A.4 Method D Back-integrated luminance meter in combination with a luxmeter . 37 Tables Table 1 Defi
15、nition of room classes with their respective requirements 18 Table 2 Visual tests for acceptance testing 19 Table 3 Metrological tests for acceptance testing . 20 Table 4 Minimum requirements for image display devices 28 Table 5 Additional requirements for image display systems used in mammography a
16、nd stereotactic mammography 30 Table 6 Visual tests for the constancy test . 32 Table 7 Metrological test items in the constancy test 32 DIN 6868-157:2014-11 4 Foreword This standard has been prepared by Working Committee NA 080-00-06 AA Bildgebende Systeme of DIN-Normenausschuss Radiologie (NAR) (D
17、IN Standards Committee Radiology) together with the Deutsche Rntgengesellschaft (German Radiological Society) and in collaboration with the Deutsche Gesellschaft fr Medizinische Physik (DGMP) (German Society for Medical Physics), the Deutsche Gesellschaft fr Nuklearmedizin (DGN) (German Society for
18、Nuclear Medicine), and the Deutsche Gesellschaft fr Radioonkologie (DEGRO) (German Society for Radiation Oncology). Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. DIN shall not be held responsible for identifying any or all such
19、patent rights. This standard contains specifications within the meaning of the Rntgenverordnung (German X-ray Ordinance) 21 and the Richtlinie zur Durchfhrung der Qualittssicherung bei Rntgeneinrichtungen zur Untersuchung oder Behandlung von Menschen nach den 16 und 17 der Rntgenverordnung (QS-RL) (
20、German Guidelines on quality assurance procedures for X-ray equipment used to examine or treat humans in accordance with sections 16 and 17 of the Rntgenverordnung (X-ray Ordinance) 22. In accordance with IEC practice, defined concepts are written in small capitals in this standard. A keyword index
21、at the end of the standard contains all defined concepts, including source(s) and a German translation. DIN 6868 consists of the following parts, under the general title Image quality assurance in X-ray departments Part 1: General Part 2: Constancy testing of film processing Part 3: Constancy testin
22、g in direct radiography Part 4: Constancy testing of medical X-ray equipment for fluoroscopy Part 5: RV constancy testing of dental radiographic equipment Part 7: Constancy testing of mammographic X-ray equipment Part 12: Constancy tests for hard copy cameras Part 13: RV constancy test of projection
23、 radiography systems with digital image receptors Part 14: RV constancy testing of X-ray installations for digital mammography Part 15: RV constancy testing of X-ray installations for dental radiographic equipment for digital cone-beam computed tomography Part 55: Acceptance testing in direct radiog
24、raphic and fluoroscopic X-ray-systems including film processing Part 56: Acceptance testing for hard copy cameras Part 150: RV acceptance test of medical radiographic and fluoroscopic X-ray equipment DIN 6868-157:2014-11 5 Part 151: Acceptance testing of dental radiographic equipment according to RV
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DIN68681572014IMAGEQUALITYASSURANCEINDIAGNOSTICXRAYDEPARTMENTSPART157XRAYORDINANCEACCEPTANCEANDCONSTANCYTESTOFIMAGEDISPLAYSYSTEMSINTHEIRENVIRONMENTX

链接地址:http://www.mydoc123.com/p-661222.html