BS 3970-2-1991 Sterilizing and disinfecting equipment for medical products - Specification for steam sterilizers for aqueous fluids in sealed rigid containers《医疗制品灭菌消毒设备 第2部分 含水液体密.pdf
《BS 3970-2-1991 Sterilizing and disinfecting equipment for medical products - Specification for steam sterilizers for aqueous fluids in sealed rigid containers《医疗制品灭菌消毒设备 第2部分 含水液体密.pdf》由会员分享,可在线阅读,更多相关《BS 3970-2-1991 Sterilizing and disinfecting equipment for medical products - Specification for steam sterilizers for aqueous fluids in sealed rigid containers《医疗制品灭菌消毒设备 第2部分 含水液体密.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD BS 3970-2: 1991 Sterilizing and disinfecting equipment for medical products Part 2: Specification for steam sterilizers for aqueous fluids in sealed rigid containersBS3970-2:1991 This British Standard, having been prepared under the directionof the Laboratory Apparatus Standards Pol
2、icy Committee, was published underthe authority of the Standards Board and comes intoeffect on 30September 1991 BSI 02-1999 The following BSI references relate to the work on this standard: Committee reference LBC/35 Draft for comment 90/54474 DC ISBN 0 580 19797 2 Committees responsible for this Br
3、itish Standard The preparation of this British Standard was entrusted by the Laboratory Apparatus Standards Policy Committee (LBC/-) to Technical Committee LBC/35, upon which the following bodies were represented: Associated Offices Technical Committee Association of British Health-care Industries A
4、ssociation of British Sterilizer Manufacturers Association of National Health Service Supplies Officers British Dental Trade Association British Surgical Trades Association Central Sterilising Club Department of Health Health and Safety Executive Infection Control Nurses Association Institute of Hos
5、pital Engineering Institute of Purchasing and Supply Institute of Sterile Services Management Joint Committee of Professional Nursing, Midwifery and Health Visiting Associations (England) Medical Sterile Products Association National Blood Transfusion Service Public Health Laboratory Service Regiona
6、l Hospital Boards Engineers Association Royal College of Pathologists Royal Pharmaceutical Society of Great Britain Society for General Microbiology Stainless Steel Fabricators Association of Great Britain The following body was also represented in the drafting of the standard, through subcommittees
7、 and panels: Guild of Hospital Pharmacists Amendments issued since publication Amd. No. Date CommentsBS3970-2:1991 BSI 02-1999 i Contents Page Committees responsible Inside front cover Foreword ii 1 Scope 1 2 Definitions 1 3 General requirements 2 4 Design pressure 2 5 Design of spray-jets 2 6 Opera
8、ting cycle 2 7 Instrumentation and controls 3 8 Performance requirements 4 9 Load-handling equipment 5 10 Information to be supplied by the manufacturer 5 Appendix A Recommendations for information to be supplied by the purchaser 6 Appendix B Test for sterilizer function without a load 6 Appendix C
9、Test for sterilizer function with a small load of low thermalcapacity 6 Appendix D Test for sterilizer function with a full load 7 Appendix E Test for coolant water quality 7 Appendix F Test for F subzero (F 0 ) control compliance 8 Appendix G Test for performance of reference instrument 8 Appendix
10、H Potential hazards associated with the sterilization ofaqueousfluidsin sealed rigid containers 9 Figure 1 Pressure generated in a nominal1L bottle, containing differentvolumes of water, when heated to121C 10 Table 1 Operating cycle conditions for acceptable temperature band 2 Table 2 Permitted rang
11、es for F 0values 5 Publication(s) referred to Inside back coverBS3970-2:1991 ii BSI 02-1999 Foreword It is recommended that this foreword be read in conjunction with the foreword to BS 3970-1. This Part of BS 3970 has been prepared under the direction of the Laboratory Apparatus Standards Policy Com
12、mittee. This Part and BS 3970-1:1990 together supersede BS 3970-2:1966, which is withdrawn. Parts 1 and 2 of the first edition of BS 3970 were both published in 1966 and covered two specific types of sterilizer. However, in this revision the structure of BS 3970 has been changed so that general requ
13、irements are now given in the revised (1990) edition of Part 1. Those requirements that are not common to most types of sterilizer are specified in subsequent Parts of this standard. These Parts giving particular requirements should be read in conjunction with BS 3970-1:1990, which they supplement o
14、r modify. The titles of these Parts are given in Part 1. When aqueous fluids in sealed rigid containers are heated, the increased pressure within them becomes a source of danger, especially when temperatures above80C are reached. Hence it is a requirement of this standard that the load in the chambe
15、r of the sterilizer shall remain inaccessible to the operator unless the temperature of all the contents of the load has become lower than80C. Appropriate interlocks on the door-controlling mechanisms are specified, together with appropriate temperature detectors. There are two accepted principles o
16、f exposing the load to a sterilization cycle that gives an adequate level of lethality, each involving the attainment of an adequate temperature for an adequate time. In Table 1 a preferred combination of time and temperature is given, this being entirely acceptable for thermo-stable products. Addit
17、ionally with thermo-labile products the user may choose to use the F 0principle in which other times and temperatures are employed. In this case the onus is on the User to determine the nature of the bioburden in the product(s) and also to determine that the proposed cycle will ensure that the proba
18、bility of survival of microorganisms does not exceed 10 6 . The F 0control function may be incorporated into the automatic controller. With thermo-labile products it will normally be appropriate to employ assisted cooling in order to limit thermal damage. It is also the responsibility of the user to
19、 provide a suitable container which will withstand the physical conditions and remain intact and sealed both during and after the process in order to prevent adulteration of the contents of the container during the process itself or any possible subsequent microbial contamination. The purchaser may
20、have to adjust the timing of the stages of the process when sterilizing a fluid with viscosity markedly higher than that of water. A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application. Comp
21、liance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover, pages i and ii, pages1to10, an inside back cover and a back cover. This standard has been updated (see copyright date) and may have
22、 had amendments incorporated. This will be indicated in the amendment table on theinside front cover.BS3970-2:1991 BSI 02-1999 1 1 Scope This Part of BS 3970 specifies particular requirements for sterilizers intended to be used for the sterilization of aqueous fluids in sealed rigid containers in wh
23、ich the heat transfer is effected by saturated steam under phase-boundary conditions, having a steady-state equilibrium during the holding phase. These requirements apply in addition to the general requirements specified in BS3970-1. NOTE 1Recommendations for the information to be supplied by the pu
24、rchaser to the manufacturer are given in Appendix A. NOTE 2The attention of both the sterilizer manufacturer and the user is drawn to the hazards associated with processing fluids in sealed rigid containers described in Appendix H. NOTE 3The titles of the publications referred to in this standard ar
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS397021991STERILIZINGANDDISINFECTINGEQUIPMENTFORMEDICALPRODUCTSSPECIFICATIONFORSTEAMSTERILIZERSFORAQUEOUSFLUIDSINSEALEDRIGIDCONTAINERS

链接地址:http://www.mydoc123.com/p-544019.html