1、Designation: D7513 09D7513 17Standard Test Method forCapacity of Mixed Bed Ion Exchange Cartridges1This standard is issued under the fixed designation D7513; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision
2、. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the performance of mixed bed ion exchange resin cartridges in the active formwhen used
3、 for deionization. The test can be used to determine the initial capacity of unused cartridges or the remaining capacityof used cartridges. In this case performance is defined as ion exchange capacity (or throughput) to two defined endpoints. Themethod does not measure organics and does not attempt
4、to determine the ultimate water quality attainable by the cartridge.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use.
5、 It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use1.4 This international standard was developed in accordance with internationally recognized principles on standardizationest
6、ablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D1125 Test Methods for Electrical Conductivity and Resistivit
7、y of WaterD1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on WaterD3375 Test Method for Column Capacity of Particulate Mixed Bed IonExchange MaterialsD5391 Test Method for Elect
8、rical Conductivity and Resistivity of a Flowing High Purity Water Sample3. Terminology3.1 DefinitionsDefinitions: For definitions of terms related to water, refer to Terminology D1129.3.1.1 For definitions of terms used in this standard, refer to Terminology D1129.4. Summary of Test Method4.1 This t
9、est method consists of exhausting a cartridge of mixed bed ion exchange material to a specific end point with aninfluent solution of known composition and volume.5. Significance and Use5.1 This test method can be used to evaluate unused mixed bed ion exchange cartridges for conformance to specificat
10、ions.5.2 This test method provides for the calculation of capacity in terms of the volume of water treated to a conductivity end point.5.3 The test method as written assumes that the ion exchange resins in the cartridge are either partially or fully converted tothe H+ or OH- form. Regeneration of th
11、e resins is not part of this method.5.4 This test method provides for the calculation of capacity on a cartridge basis.1 This test method is under the jurisdiction of ASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.08 on Membranes and Ion ExchangeMaterials.Current ed
12、ition approved March 15, 2009June 1, 2017. Published April 2009June 2017. Originally approved in 2009. Last previous edition approved in 2009 as D7513 09. DOI: 10.1520/D7513-09.10.1520/D7513-17.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at se
13、rviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version
14、. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International,
15、 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.5 This test method may be used to test different size mixed bed resin cartridges. The flow rate of test water and the frequencyof sampling are varied to compensate for the approximate volume of resin in the test c
16、artridge.6. Apparatus6.1 Test Assembly (Fig. 1) , ), consisting of the following:6.1.1 Cartridge, and pressure vessel, if required. Connections shall be provided at the top and the bottom for the admission andremoval of the ion exchange test water as described in 7.37.4.2.6.1.2 Test solution tank. A
17、dequate means of regulating and measuring flow through the cartridge shall be provided. If the testflow rate (8.1) can not be obtained by gravity feed, then a pump should be used.6.1.3 Measuring circuit and in-line conductivity cells shall be as described in Test Methods D1125 or D5391. A continuous
18、recorder is recommended.7. Reagents and Materials7.1 Purity of ReagentsReagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that allreagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, wheresu
19、ch specifications are available.3 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently highpurity to permit its use without lessening the accuracy of the determination.7.2 Purity of WaterAll reference to water in this test method shall be understood to mean
20、Reagent Water Type I or IIconforming to Specification D1193.7.3 Ammonium Hydroxide Solution (1+19)Carefully pour 50 mL of concentrated ammonium hydroxide (NH4OH, sp gr 0.90)into 500 mL of water, stirring constantly. Cool to 25 6 5C and dilute to 1 L with water. Mix well.7.4 Ion Exchange Test Water (
21、10 meq/L, or 500 ppm as CaCOD (10 meq/L)3)Prepare a test water containing, in eachlitre:litre, 0.585 g of oven-dried (105C) sodium chloride (NaCl).Approximately 25 Lof this solution are required for a single test.7.4.1 0.294 g of calcium chloride dihydrate (CaClTo standardize for chloride content, p
22、ipet three 100-mL portions. Add onedrop of methyl orange indicator solution and one drop of phenolphthalein indicator solution to each and neutralize, if required, bydropwise addition of HNO3 (1+9) until the color changes from yellow to orange followed by dropwise addition of NH4OH (1+19)to restore
23、the yellow color. Pipet in 1 mLof K2-2HCrO24O), solution (50 g/L) and titrate each with standard 0.1 NAgNO3 solutionuntil the color of the supernatant solution changes from yellow to red-orange and persists for 30 s with vigorous swirling. Recordthe average number of millilitres of AgNO3 solution us
24、ed to 60.02 mL.3 Reagent Chemicals, American Chemical Society Specifications, American Chemical Society, Washington, DC. For Suggestions on the testing of reagents not listed bythe American Chemical Society, see Annual Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD75132017RED5000STANDARDTESTMETHODFORCAPACITYOFMIXEDBEDIONEXCHANGECARTRIDGES 混合 离子交换 能力 标准 试验 方法

链接地址:http://www.mydoc123.com/p-525709.html