ANSI ISO ASTM 52116-2013 Standard Practice for Dosimetry for a Self-Contained Dry-Storage Gamma Irradiator.pdf
《ANSI ISO ASTM 52116-2013 Standard Practice for Dosimetry for a Self-Contained Dry-Storage Gamma Irradiator.pdf》由会员分享,可在线阅读,更多相关《ANSI ISO ASTM 52116-2013 Standard Practice for Dosimetry for a Self-Contained Dry-Storage Gamma Irradiator.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、ISO/ASTM 52116:2013(E)Standard Practice forDosimetry for a Self-Contained Dry-Storage GammaIrradiator1This standard is issued under the fixed designation ISO/ASTM 52116; the number immediately following the designation indicates theyear of original adoption or, in the case of revision, the year of l
2、ast revision.1. Scope1.1 This practice outlines dosimetric procedures to be fol-lowed with self-contained dry-storage gamma irradiators. Forirradiators used for routine processing, procedures are given toensure that product processed will receive absorbed doseswithin prescribed limits.1.2 This pract
3、ice covers dosimetry in the use of dry-storagegamma irradiators, namely self-contained dry-storage137Cs or60Co irradiators (shielded freestanding irradiators). It does notcover underwater pool sources, panoramic gamma sources, nordoes it cover self-contained bremsstrahlung X-ray units.1.3 The absorb
4、ed-dose range for the use of the dry-storageself-contained gamma irradiators covered by this practice istypically 1 to 105Gy, depending on the application. Theabsorbed-dose rate range typically is from 102to 103Gy/min.1.4 For irradiators supplied for specific applications, spe-cific ISO/ASTM or ASTM
5、 practices and guides providedosimetric procedures for the application. For proceduresspecific to dosimetry in blood irradiation, see ISO/ASTMPractice 51939. For procedures specific to dosimetry in radia-tion research on food and agricultural products, see ISO/ASTMPractice 51900 . For procedures spe
6、cific to radiation hardnesstesting, see ASTM Practice E1249. For procedures specific tothe dosimetry in the irradiation of insects for sterile releaseprograms, see ISO/ASTM Guide 51940. In those cases coveredby ISO/ASTM 51939, 51900 , 51940, or ASTM E1249, thosestandards take precedence.1.5 This doc
7、ument is one of a set of standards that providesrecommendations for properly implementing and utilizingdosimetry in radiation processing. It is intended to be read inconjunction with ASTM E2628, “Practice for Dosimetry inRadiation Processing”.1.6 This standard does not purport to address all of thes
8、afety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced documents2.1 ASTM Standards:2E170 Terminology Relating to
9、Radiation Measurements andDosimetryE1249 Practice for Minimizing Dosimetry Errors in Radia-tion Hardness Testing of Silicon Electronic Devices UsingCo-60 SourcesE2628 Practice for Dosimetry in Radiation ProcessingE2701 Guide for Performance Characterization of Dosim-eters and Dosimetry Systems for U
10、se in Radiation Pro-cessing2.2 ISO/ASTM Standards:251261 Practice for Calibration of Routine Dosimetry Sys-tems for Radiation Processing51539 Guide for Use of Radiation-Sensitive Indicators51707 Guide for Estimating Uncertainties in Dosimetry forRadiation Processing51900 Guide for Dosimetry in Radia
11、tion Research on Foodand Agricultural Products51939 Practice for Blood Irradiation Dosimetry51940 Guide for Dosimetry for Sterile Insects Release Pro-grams2.3 International Commission on Radiation Units and Mea-surements (ICRU) Reports:3ICRU 85a Fundamental Quantities and Units for IonizingRadiation
12、2.4 ANSI Standards:4ANSI/HPS N43.7 Safe Design and Use of Self-Contained,Dry Source Storage Gamma Irradiators (Category I)2.5 Joint Committee for Guides in Metrology (JCGM)Reports:JCGM 100:2008, GUM 1995 with minor corrections, Evalu-ation of measurement data Guide to the Expression of1This practice
13、 is under the jurisdiction of ASTM Committee E61 on RadiationProcessing and is the direct responsibility of Subcommittee E61.04 on SpecialtyApplication, and is also under the jurisdiction of ISO/TC 85/WG 3.Current edition approved Aug. 16, 2012. Published April 2013. Originallypublished as ASTM E 21
14、1600. Last previous edition ASTM E 211600. Thepresent International Standard ISO/ASTM 52116:2013(E) replaces E 2116-00 and isa major revision of the last previous edition ISO/ASTM 52116:2002:(E).2For referenced ASTM and ISO/ASTM standards, visit the ASTM website,www.astm.org, or contact ASTM Custome
15、r Service at serviceastm.org. ForAnnual Book of ASTM Standards volume information, refer to the standardsDocument Summary page on the ASTM website.3International Commission on Radiation Units and Measurements (ICRU), 7910Woodmont Ave., Suite 800, Bethesda, MD 20810, U.S.A.4Available from the Health
16、Physics Society, http:/hps.org. ISO/ASTM International 2017 All rights reservedThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Rec
17、ommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1Uncertainty in Measurement5JCGM 100:2008, VIM International vocabulary of metrol-ogy Basis and general concepts and associated terms63. Terminology3.1 Definitions:3.1.1 absorbed-dose mappingmeasurement o
18、f absorbeddose within an irradiated product to produce a one-, two-, orthree-dimensional distribution of absorbed dose, thus renderinga map of absorbed-dose values.3.1.2 calibrationVIM, 6.11 set of operations underspecified conditions, which establishes the relationship be-tween values indicated by
19、a measuring instrument or measur-ing system, and the corresponding values realised by standardstraceable to a nationally or internationally recognised labora-tory.3.1.2.1 DiscussionCalibration conditions include environ-mental and irradiation conditions present during irradiation,storage and measure
20、ment of the dosimeters that are used for thegeneration of a calibration curve. To achieve stable environ-mental conditions, it may be necessary to condition thedosimeters before performing the calibration procedure.3.1.3 dose uniformity ratioratio of the maximum to theminimum absorbed dose within th
21、e irradiated product.3.1.4 measurement management systemset of interrelatedor interacting elements necessary to achieve metrologicalconfirmation and continual control of measurement processes.3.1.5 transit doseabsorbed dose delivered to a product (ora dosimeter) while it travels between the non-irra
22、diationposition and the irradiation position, or in the case of amovable source while the source moves into and out of itsirradiation position.3.2 Definitions of other terms used in this standard thatpertain to radiation measurement and dosimetry may be foundin ASTM Terminology E170. Definitions in
23、ASTM Terminol-ogy E170 are compatible with ICRU 85a; that document,therefore, may be used as an alternative reference.4. Significance and use4.1 The design and operation of a self-contained irradiatorshould ensure that reproducible absorbed doses are obtainedwhen the same irradiation parameters are
24、used. Dosimetry isperformed to determine the relationship between the irradiationparameters and the absorbed dose.4.1.1 For most applications, the absorbed dose is expressedas absorbed dose to water (see ISO/ASTM Practice 51261).For conversion of absorbed dose to water to that to othermaterials, for
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIISOASTM521162013STANDARDPRACTICEFORDOSIMETRYFORASELFCONTAINEDDRYSTORAGEGAMMAIRRADIATORPDF

链接地址:http://www.mydoc123.com/p-437123.html