UOP 539-2012 Refinery Gas Analysis by GC.pdf
《UOP 539-2012 Refinery Gas Analysis by GC.pdf》由会员分享,可在线阅读,更多相关《UOP 539-2012 Refinery Gas Analysis by GC.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1963, 1973, 1987, 1997, 2012 UOP LLC.
3、All rights reserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9
4、555 FAX, or 610.832.9585 PHONE. Refinery Gas Analysis by GC UOP Method 539-12 Scope This method is for determining the composition of refinery gas streams or vaporized liquefied petroleum gas (LPG) using a preconfigured, commercially available gas chromatograph. At a minimum, the following component
5、s are reported: non-condensable gases, individual C1 through C4 hydrocarbons and C5 paraffins. C5 olefins and C6+ hydrocarbons are reported as a composite. Some configurations of the analyzer will separate the C5 olefins and certain C6+ hydrocarbons. Oxygen is not separated from argon. Results for h
6、ydrogen sulfide, if present, may not be quantitative on some analyzers. The lower limit of quantitation for each component or composite is 0.05 mol-%. If the measurement of lower concentrations of hydrocarbons are needed, use UOP Method 899, “Trace Hydrocarbons in Hydrogen, Hydrocarbon Gases, or LPG
7、 by GC.” LPG samples may be analyzed by UOP Method 373, “Composition of C2 through C5 Hydrocarbon Mixtures by GC,” or UOP Method 980, “C5 and Lower Boiling Dienes, Olefins, and Paraffins in Naphthas by GC.” Alternatively, LPG samples may be expanded to the gas phase and run by this method. Reference
8、s Analyzer manufacturers operating manual UOP Method 373, “Composition of C2 through C5 Hydrocarbon Mixtures by GC,” www.astm.org UOP Method 899, “Trace Hydrocarbons in Hydrogen, Hydrocarbon Gases, or LPG by GC,” www.astm.org UOP Method 980, “C5 and Lower Boiling Dienes, Olefins, and Paraffins in Na
9、phthas by GC,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is injected into a GC configured for refinery gas analysis with multiple valves and columns, and, usually, more than one detector. The included data system controls the operat
10、ing sequence, measures the resultant peaks, and performs component quantitation based upon calibration with known gas blends. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. 2 of 7 539-12 Analyzer, refinery gas. Ref
11、inery gas analyzers are available pre-configured, including a data system, from several suppliers. A partial list of suppliers is included in Appendix A, and includes instrument features as described in the suppliers website, as of publication of this method. Other vendors also supply similar system
12、s. Confirm with the supplier that the analyzer is appropriate for the streams to be analyzed and the analysis required. Instrument gas supply regulators; not all instruments will require all the regulators listed below. Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-10
13、0 psi), Matheson Tri-Gas, Model 3122-590 Regulator, argon, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Regulator, helium, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Regulator, h
14、ydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Leak detector, gas, Grace Davison, Cat. No. 60229 Reagents a
15、nd Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Calibration blends, as specified by the analyzer operating manual, Matheson Tri-Gas, or local supply. For increased accuracy, one calibration blend should have a co
16、mposition close to that of the samples to be analyzed, but not less than 1 mol-% of each component of interest. Table 1 lists a set of calibration blends that would be generally applicable to a variety of sample types. Gas purifier, for carrier gases, to remove oxygen and moisture, VICI Mat/Sen, Cat
17、. Nos. P100-1 (helium), P200-1 (hydrogen), P300-1 (nitrogen), and (optional) indicating oxygen traps, Restek, Cat. No. 22010 Instrument gases, as needed; not all instruments will require all the gases listed below. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane, local supply Argon, 9
18、9.99% minimum purity, local supply Helium, 99.99% minimum purity, local supply Hydrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane, local supply Nitrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane, local supply Procedure T
19、he analyst is expected to be familiar with general laboratory practices, the technique of gas chromatography, and the equipment being used. Dispose of used reagents, materials, and samples in an environmentally safe manner according to local regulations. 3 of 7 539-12 Table 1 Calibration Blend Examp
20、le Concentrations, mol-% Component Blend 1 Blend 2 Blend 3 Hydrogen 50.0 - 20.0 Nitrogen 15.0 15.0 45.0 Argon 5.0 - 5.0 Methane 5.0 10.0 - Ethane 5.0 10.0 - Ethylene - 5.0 - Propane 12.0 12.0 - Propylene - 6.0 - Propadiene - 1.0 - n-Butane 3.0 10.0 - Isobutane 3.0 10.0 - 1-Butene - 5.0 - Isobutylene
21、 - 5.0 - trans-2-Butene - 5.0 - cis-2-Butene - 5.0 - 1,3-Butadiene - 1.0 - n-Pentane 1.0 - - Isopentane 1.0 - - CarbonDioxide - - 5.0 CarbonMonoxide - - 10.0 Helium - 10.0 HydrogenSulfide - - 5.0 If specific components, e.g., hydrogen sulfide, helium, or 1,3-butadiene are known to be not present in
22、the sample types being analyzed, they may be deleted from the blends, and the composition of the other components adjusted to total 100%. Chromatographic Technique 1. Install the gas purifier in the supply line between the carrier gas source and the carrier gas inlets on the gas chromatograph. Colum
23、n life is significantly reduced if gas purifiers are not used. Replace the gas purifiers at intervals determined by good laboratory practice. Indicating oxygen traps may be placed downstream of the gas purifiers. When the indicator shows one-half used, replace both the gas purifier and the indicatin
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP5392012REFINERYGASANALYSISBYGCPDF
