1、A-A-50504 bb m 9979974 0064447 4 m I INCH-POUND I A-A- 50504 28 February 1992 COMMERCIAL ITEM DESCRIPTION ANALYZERS, FLUE-GAS, ORSAT-TYPE, PORTABLE The General Services Administration has authorized the use of this Commercial Item Description (CID) as a replacement for Military Specification MIL-A-1
2、8218B which is cancelled, Abstract. the volumetric percentage of C02, This CID covers portable Orsat-type analyzers for the measurement of 02, and CO occurring in flue-gases. 1. Classification. Flue-gas analyzers shall be of the following sizes, as specified in the contract or order: Size 1 - 100 mi
3、lliliters (mL). Size 2 - 50 mL. Salient characteristics 1. Description. The analyzer shall be a metal-frame-mounted, portable assembly, consisting of a jacketed burette, manifold with control valves, absorption pipettes with integral or close-coupled reservoirs for absorption chemicals, an expansion
4、 bag(s) or chamber(s), a leveling bottle, a dirt-excluding protective enclosure, necessary connectors, flexible tubing and stoppers, an aspirating bulb, sampling line, filter, and absorption chemicals. 2. Design. easy and trouble-free manipulation, rapid and thorough gas absorption, easy readability
5、, and portability as a unit. Jars and bumps, such as those encountered during transportation and under the conditions of intended use, shall not cause breakage, misalignment, loosening of components, or leaks. The burette and pipettes shall be arranged in a row, from right to left, as follows: pipet
6、te, and CO pipette. The bottom of the frame, or the support points, The analyzer shall be designed to withstand rough usage and for Measuring burette, CO2 pipette, O2 IBeneficial comments (recommendations, additions, deletions) and any pertinent1 Idata which may be of use in improving this document
7、should be addressed to: I ICommanding Officer (Code 156), Naval Construction Battalion Center, Port I Hueneme. CA 93043-5000. I FSC 6630 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without lice
8、nse from IHS-,-,-A-A-50504 66 I 9999974 0064448 6 I A-A- 50504 shall touch a flat supporting surface simultaneously. With the analyzer standing on a flat surface, the long axis of each burette and pipette shall be essentially vertical. 3. Leakave. Chemicals shall not leak or creep past or through se
9、als, connectors, valves, or cocks. 4. Burette. iacket. and scale. The burette and jacket shall be of clear, high grade materials free of striae, whorls, or other imperfections which would result in false indications of liquid levels or scale reading. The materials shall not become fogged or discolor
10、ed with age. The burette shall be of glass thoroughly annealed after being formed. shall be of clear glass or of clear plastic. Changes in cross-sectional area in the burette shall be made with generously rounded curves. Wall thickness shall be substantially uniform in each portion, and changes in w
11、all thickness between different-diameter portions shall be uniform. On or adjacent to the lower portion of the burette, there shall be a scale marked with parallel division lines. The division lines shall be sharply defined and shall be of a deep dark color. The lines at integral points shall be som
12、ewhat longer that the intermediary lines. The scale intervals shall be 0.2 percent for both size 1 and size 2 burettes. The scale range shall be zero to 50 percent for size 1 burettes and zero to 20 percent for size 2 burettes. Scales shall be etched and filled in with a fusible material on the bure
13、tte or shall be marked in enamel on a corrosion-resistant metal strip. The latter, when used, shall be assembled with the burette in such a way that the whole length of scale will remain in contact with the burette, and so that the scale will not move with respect to the burette because of vibration
14、 or other movement of the analyzer, or because of changes in dimension due to temperature. The jacket 5, Manifold and valves. The manifold and valves shall be of nonbrittle, nonfragile materials which shall not be affected chemically and shall not deteriorate physically as a result of contact with a
15、ir or flue-gas, or with the liquids used in the analyzer. Glass manifolds and glass stopcocks will not be acceptable. Means shall be provided for easy cleaning of gas passages without extensive disassembly. Valve control knobs and handles shall be designed for easy and interference-free manipulation
16、. Valves shall close gas-tight with light seating-pressure and shall not stick or bind in normal use. 6. Absorption pipettes and reservoirs. Absorption pipettes may be integral with or separable from reservoirs. When they are separable, assembly shall be effected by means of a compression-type fitti
17、ng, which shall provide a leak-tight joint without breaking the pipette. The absorption space shall be filled to the neck of the pipette capillary with contact-extending material, or shall be provided with a bubbler, having at least four bubbling openings, approximately 1 mL in diameter, near the bo
18、ttom of the pipette and equipped with an automatic valve. Contact-extending material shall be so arranged that jostling and jarring during transport and handling will not cause the material to pack or fall through the bottom of the pipette or otherwise to become dislodged. Bubbler pipettes shall be
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- NAVYAA505041992ANALYZERSFLUEGASORSATTYPEPORTABLE 便携式 奥尔氏 气体 排气 分析 PDF

链接地址:http://www.mydoc123.com/p-976280.html