GPA STD 2177-2013 Analysis of Natural Gas Liquid Mixtures Containing Nitrogen and Carbon Dioxide by Gas Chromatography.pdf
《GPA STD 2177-2013 Analysis of Natural Gas Liquid Mixtures Containing Nitrogen and Carbon Dioxide by Gas Chromatography.pdf》由会员分享,可在线阅读,更多相关《GPA STD 2177-2013 Analysis of Natural Gas Liquid Mixtures Containing Nitrogen and Carbon Dioxide by Gas Chromatography.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、 Analysis of Natural Gas Liquid Mixtures Containing Nitrogen and Carbon Dioxide by Gas Chromatography Adopted as Tentative Standard. 1977 Revised and Adopted as a Standard. 1984 Revised 1989, 1995, 2003, 2012, 2013 Gas Processors Association 6526 East 60th Street Tulsa, Oklahoma 74145 GPA Standard 2
2、177-13 DISCLAIMER GPA publications necessarily address problems of a general nature and may be used by anyone desiring to do so. Every effort has been made by GPA to assure accuracy and reliability of the information contained in its publications. With respect to particular circumstances, local, sta
3、te, and federal laws and regulations should be reviewed. It is not the intent of GPA to assume the duties of employers, manufacturers, or suppliers to warn and properly train employees, or others exposed, concerning health and safety risks or precautions. GPA makes no representation, warranty, or gu
4、arantee in connection with this publication and hereby expressly disclaims any liability or responsibility for loss or damage resulting from its use or for the violation of any federal, state, or municipal regulation with which this publication may conflict, or for any infringement of letters of pat
5、ent regarding apparatus, equipment, or method so covered. FOREWARD GPA 2177 provides the gas processing industry a method for determining the compositional analysis of demethanized liquid hydrocarbon streams containing nitrogen, air and carbon dioxide. Purity products such as an ethane/propane mix a
6、nd other products that meet the compositional ranges listed in the standard may also utilize this method. The precision statements contained in this standard are based on the statistical analysis of round-robin laboratory data obtained by GPA Section B. The detail of how this data yielded the equati
7、ons covered in the precisions statements can be found in GPA RR188. This standard was developed by the cooperative efforts of many individuals from industry under the sponsorship of GPA Section B, Analysis and Test Methods. Throughout this publication, the latest appropriate GPA Standards are refere
8、nced. “Copyright2013 by Gas Processors Association. All rights reserved. No part of this Report may be reproduced without the written consent of the Gas Processors Association.” 1 1. SCOPE 1.1 This method is intended for the analysis of demethanized liquid hydrocarbon streams containing nitrogen/air
9、 and carbon dioxide, and purity products such as an ethane/propane mix that fall within the compositional ranges listed in Table I. This method is limited to mixtures containing less than 5 mole percent of heptanes and heavier fractions (20 mole percent of hexanes and heavier fractions). Table I Com
10、ponents and Recommended Compositional Ranges Components Concentration Range (mole %) Nitrogen 0.01 5.0 Carbon Dioxide 0.01 5.0 Methane 0.01 5.0 Ethane 0.01 95.0 Propane 0.01 100.0 Iso-Butane 0.01 100.0 N-Butane and 2,2-Dimethylpropane 0.01 100.0 Iso-Pentane 0.01 15.0 N-Pentane 0.01 15.0 2,2-Dimethyl
11、butane, 2,2-Dimethylbutane, and 2-Methylpentane, 3-Methylpentane, Cyclopentane, N-Hexane 0.01 15.0 Heptanes & Heavier 0.01 5.0 1.2 The heptanes and heavier fraction if present in the sample is analyzed by either (1) use of a pre-cut column to elute heptanes and heavier first as a single peak, or (2)
12、 reverse flow of carrier gas after n-hexane, and peak grouping of the heptanes plus fraction. For purity mixes without heptanes and heavier, no reverse of carrier flow is required. NOTE 1 This method can be used for a hexanes plus analysis by adjusting the timing of the backflush event to occur foll
13、owing the elution of n-pentane. NOTE 2 CAUTION: In the case of unknown samples with a relatively large C6+or C7+fraction and where precise results are important, it is desirable to determine the molecular weight (or other pertinent physical properties) of these fractions. Since this method makes no
14、provision for determining physical Properties, the physical properties needed must be determined by extended analysis or agreed to by the contracting parties. The extended analysis will be made according to GPA Publication 2186. NOTE 3 For samples containing more than 5 mole percent of heptanes and
15、heavier fractions (20 mole percent of hexanes and heavier fractions), analysis will be made according to GPA Publication 2103. 2. SUMMARY OF METHOD 2.1 Components to be determined in a demethanized hydrocarbon liquid mixture are physically separated by gas chromatography and compared to calibration
16、data. A fixed volume of sample in the liquid phase is isolated in a suitable sample inlet system and flashed onto the chromatographic column. NOTE 4 For lab chromatographs, the comparison to calibration data is performed under identical operating conditions. For on-line chromatographs, the sample ma
17、y not necessarily be at the same conditions as the calibrations data so care should be taken to ensure that the pressure at the sample injection point is sufficiently above the vapor pressure of the product. 2.2 A precut column is backflushed when n-hexane has eluted into the main analytical column,
18、 so that the heptanes+ are eluted as a single peak at the beginning of the chromatogram. Two advantages of this method are: (1) Better precision in measuring the C7+portion of the sample, and (2) Reduction in analysis time over the single column approach by approximately 40%. See Figure 1. 2.2.1 As
19、an alternate method, the components nitrogen/air through n-hexane are individually separated with the carrier flow in the forward direction. The numerous heavy end components are grouped into an irregular shaped peak by reversing direction of carrier gas through the column via a switching valve imme
20、diately following the elution of normal hexane. (See Figure 2.) Samples which contain no heptanes+ fraction are analyzed until the final component has eluted with no reverse of carrier flow. 2.3 The chromatogram is interpreted by comparing the areas of component peaks obtained from the unknown sampl
21、e with corresponding areas obtained from a run of a selected reference standard. Any component in the unknown suspected to be outside the linearity range of the detector, with reference to the known amount of that component in the reference standard, must be determined by a response curve. (See Sect
22、ions 5.1 & 5.1.1 for further explanation of instrument linearity check procedures.) 3. APPARATUS 3.1 Any gas chromatograph may be used which meet the criteria listed in Section 8, Precision. The following equipment specifications have been found useful in designing an instrument to meet those criter
23、ia. 3.1.1 Environment - The gas chromatograph shall be housed in a climate controlled environment to ensure stability. Analysis of Natural Gas Liquid Mixtures Containing Nitrogen and Carbon Dioxide by Gas Chromatography 2 3.1.2 Detector - The detector shall be a thermal conductivity type. It must be
24、 sufficiently sensitive to produce a detectable peak for each component of interest at 100 ppm molar volume. 3.1.3 Sample Inlet System. Liquid - A liquid sampling valve must be provided, capable of entrapping a fixed volume of sample at a pressure at least 200 psi (1379 kPa) above the vapor pressure
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- GPASTD21772013ANALYSISOFNATURALGASLIQUIDMIXTURESCONTAININGNITROGENANDCARBONDIOXIDEBYGASCHROMATOGRAPHYPDF

链接地址:http://www.mydoc123.com/p-782332.html