FORD FLTM EU-BQ 010-1-2000 FILM THICKNESS MEASUREMENT FOR ANODIZED ALUMINUM.pdf
《FORD FLTM EU-BQ 010-1-2000 FILM THICKNESS MEASUREMENT FOR ANODIZED ALUMINUM.pdf》由会员分享,可在线阅读,更多相关《FORD FLTM EU-BQ 010-1-2000 FILM THICKNESS MEASUREMENT FOR ANODIZED ALUMINUM.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、 FORD LABORATORY TEST METHOD EU-BQ 010-01 Date Action Revisions 2000 11 09 Editorial no technical change A. Cockman 1994 02 28 Printed copies are uncontrolled Page 1 of 5 Copyright 2000, Ford Global Technologies, Inc. FILM THICKNESS MEASUREMENT FOR ANODIZED ALUMINUM Application This method describes
2、 the use of the light section microscope for the non - destructive measurement of anodised aluminum film thickness. Apparatus and Materials Required Zeiss Light Section Microscope This instrument has its design based upon the light section principle, and has been found to be adaptable to the measure
3、ment of thin film thicknesses of any transparent coating over both magnetic and non - magnetic surfaces. The basic light sectio n microscope unit (see drawing) consists of the following major components: A microscope with built - in illumination device, built -in measuring eyepiece and two pairs of
4、objectives for 200X and 400X total magnification respectively. A Stand with Stage Plate Mechanical Stage With Vee blocks for positioning relatively small test samples. Flat Glass Reference Surface A flat glass reference surface is required for proper alignment of the slit image. Dimensions approx 50
5、 x 50 mm and a minimum thickness of 0. 8 mm. Sample Holder A clamping device (or equivalent) compatible with the size and shape of the samples to be tested, which will facilitate positioning the test sample so that the test area is very nearly normal to the axis of the microscope. Conditioning and T
6、est Conditions All test values indicated herein are based on material conditioned in a controlled atmosphere of 23 +/ - 2 C and 50 +/ - 5 % relative humidity for not less than 24 h prior to testing and tested under the same conditions unless otherw ise specified. FORD LABORATORY TEST METHOD EU-BQ 01
7、0-01 Page 2 of 5 Copyright 2000, Ford Global Technologies, Inc. Procedure A. Instrument Set-up 1. Place the instrument on a flat and relatively vibration - free surface. 2. Switch the power supply unit on by setting the knob provided to any mark between 1 and 8 on the graduated disc. This setting ma
8、y later be changed to obtain a desired level of illumination. B. Initial Focus and Alignment 1. Focus the cross lines, visible through the eyepiece, as outlined in the instrument manual (supplied by the manufacturer). 2. Place the flat glass reference su rface on the stage plate. The surface of this
9、 glass piece must be thoroughly clean (free of fingerprints, grease film, etc.). 3. Looking into the eyepiece, lower the microscope body by turning the vertical drive knob slowly, to obtain a well focused ligh t band. 4. This light band must be parallel to the horizontal portion of the cross - lines
10、. If it is not parallel, it must be aligned accordingly by means of the knob provided, as outlined in the instrument manual. Once this light band has been properly a ligned, the aligning knob should not be touched in the course of instrument operation. 5. By means of the vertical drive knob elevate
11、the microscope body to a safe clearing distance between objectives and stage plate. 6. Remove the glass reference surfac e from the stage plate. C. Selecting the Objectives 1. Select the 400X total magnification pair of objectives. At this level of magnification each division on the measuring drum i
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- FORDFLTMEUBQ01012000FILMTHICKNESSMEASUREMENTFORANODIZEDALUMINUMPDF

链接地址:http://www.mydoc123.com/p-747851.html