EN 12884-1999 en Biotechnology - Performance Criteria for Centrifuges《生物技术 离心机性能标准》.pdf
《EN 12884-1999 en Biotechnology - Performance Criteria for Centrifuges《生物技术 离心机性能标准》.pdf》由会员分享,可在线阅读,更多相关《EN 12884-1999 en Biotechnology - Performance Criteria for Centrifuges《生物技术 离心机性能标准》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD Biotechnology - Performance criteria for centrifuges The European Standard EN 12884: 1999 has the status of a British Standard ICs 07.080 NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAW BS EN 12884:1999 - - STD=BSI BS EN 12884-ENGL 1999 m IbZ4bb9 0778945 3Tb m
2、 been prepared under the direction of the Sector Amd.No. Date Committee for Materials and Chemicals, was published under the authority of the Standards Committee and comes into effect on 10 June 1999 O BSI O-1999 ISBN O 580 32271 8 BS EN 128W1999 Comments National foreword This British Standard is t
3、he English language version of EN 12W1999. The UK participation in its preparation was entrusted to Technical Cornmittee CW58, Biotechnology, which has the responsibility to: - aid enquirers to understand the text; - present to the responsible European committee any enquiries on the - monitor relate
4、d international and European developments and promulgate interpretation, or proposals for change, and keep the UK interests informed; them in the UK A lict of organizations represented on this committee can be obtained on request to its secretary. Cross-references The British Standards which impleme
5、nt international or European publications referred to in this document may be found in the BSI Standards Catalogue under the section entited ?International Standards Correspondence Index?, or by using the ?Find facility of the BSI Standards Electronic Catalogue. A British Standard does not purport t
6、o include all the necessasy provisions of a contract. Users of British Standards are responsible for their correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover,
7、 the EN title page, pages 2 to 8, an inside back cover and a back cover. STD.BSI BS EN 22884-ENGL 2779 H LbZLibb9 077874b 232 m EUROPEIN STANDARD NORM3 EUROPENNE EUROP - release of microorganisms after operation due to insuffcient inactivation andlor removal of microorganisms when the centrifuge is
8、opened or dismantled. 5 Performance classes 5.1 General The centrifuge shali be classified for the following performance criteria with regard to the contained use of microorganisms: - leaktightmess; - cleanability; - sterilizability. The performance of the centrifuge shali be determined separately f
9、or each of these criteria in accordance wih Tables 1,2 and 3. NOTE A centrifuge can for example be in class SL4 for sterilizability, but in class LI-C for leaktightness. O BSI O81999 STD.BSI BS EN 12884-ENGL 1977 LbZIbb7 0778950 7b3 Performance class for leaktightness Leaktightness Index (LI) LI-A L
10、I-B LI-C Description of performance class leakage of target microorganism not defined leakage11 of target microorganism detected and quantified under defined conditions leakage) of target microorganism tested under defined conditions and leakage below detection limit or threshold vaJu$) 5.3 Cleanabi
11、lity The performance classes for cleanability of the centrifuge are given in Table 2. Table 2 - Cleanability performance for cleanability Cleanability Index CI-A visible soil or cleanliness not defined Performance class for sterilizability Sterilizability Index (SI) SI-A CI-B Description of performa
12、nce class centrifuge not suitable or not tested for reduction of viable target microorganisms cleanability) tested and quantified under defined conditions or designed with regard to specified technical criteria SI-C CI-c centrifuge can be sterilized cleanability) tested under defined conditions and
13、soil below detection limit or threshold value 2) i) Based on assessment by BATNEEC (Best Available Techniques Not Entailing Excessive Costs)*. 2, Prescribed threshold value should be based on the required safety level and can for example be the detection limit of an Page 5 EN 12884:1999 SI-B centrif
14、uge can be treated for a specified reduction of viable target microorganisms 5.5 Use of performance classes in equipment selection The selection of the appropriate class for performance of a centrifuge in a biotechnological process shall be made in accordance with the risk analysis as described in E
15、N 12460. NOTE In practice the following has been found to be suitable for guidance for the usage of classes given in Tables 1,2 and 3. If the release of microorganisms is to be minimized, a centrifuge can be used which is classified SI-B, CI-A and LI-B or better. If the release of microorganisms is
16、to be prevented, a centrifuge can be used which is classified SI-C, CI-B and LI-C or better. If these performance classes are not used as described or are considered to be inappropriate, alternative and/or additional measures are necessary e.g. procedural measures, secondary containment or disposabl
17、e equipment. * Use of BATNEEC doed not mean that financial issues moderate the degree of safety. Where several methods are available, the user can choose the most convenient, provided that it gives results of the necessary qualis. O BSI 06-1999 STD-BSI BS EN 12884-ENGL L777 Lb24bb7 0778753 bTT Page
18、6 EN 128849999 6 Classification and verification of performance 6.1 General Centrifuges covered by this European Standard shall be classified in accordance with Tables 1,2 and 3. The performance of centrifuges is determined and verified by the manufacturer or by the user. NOTE 1 Under certain circum
19、stances the user can reclassify the centrifuge. if for example the centrifuge could comply with steriiizability class SI-C but is classified SI-B because the manufacturer has done only a simple test, then the user can reclassify the centrifuge to SI-C if he executes and documents the appropriate tes
20、ts. Tests shall be performed under representative process conditions. To be classied according to this European Standard the centrifuge shall be designed according to EN 2921 If centrifuges are manufactured in accordance with an already classified prototype they are classified in the same class as t
21、he prototype, if: ad -2 and EN 12547. - there is no change in design or materials used or - there are only minor changes that do not affect biosafew. EXAMPLE It is not to be expected that changes in the drum design will have an effect on leakage from the casing. If there are only minor changes that
22、do not affect biosafew, the manufacturer shall document the changes and shall provide evidence to the user that these changes will not result in the need for reclassification. If, for a classified existing centrifuge, parts or components are replaced by other parts or components of similar performan
23、ce class, the overall classification of the centrifuge remains the same. NOTE 2 Verification of classification can for example be done by operating an appropriate qualis assurance system (see annex A FI, 71). The test report shall clearly state the boundaries of the tested object. NOTE 3 Tests may c
24、oncern a centrifuge without any auxiliary equipment or a centrifuge including all the auxiliary equipment needed to run the centrifuge. In practice it can be difficult or impossible to test for example sterilization for a centrifuge with no auxiliary equipment as the effectiveness of the sterilizati
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- EN128841999ENBIOTECHNOLOGYPERFORMANCECRITERIAFORCENTRIFUGES 生物技术 离心机 性能 标准 PDF

链接地址:http://www.mydoc123.com/p-706908.html