ECA EIA-364-80-2002 TP-80 Low Frequency Shielding Effectiveness Test Procedure for Electrical Connectors and Sockets《TP-80 电连接器和插座的低频防护效果试验方法》.pdf
《ECA EIA-364-80-2002 TP-80 Low Frequency Shielding Effectiveness Test Procedure for Electrical Connectors and Sockets《TP-80 电连接器和插座的低频防护效果试验方法》.pdf》由会员分享,可在线阅读,更多相关《ECA EIA-364-80-2002 TP-80 Low Frequency Shielding Effectiveness Test Procedure for Electrical Connectors and Sockets《TP-80 电连接器和插座的低频防护效果试验方法》.pdf(30页珍藏版)》请在麦多课文档分享上搜索。
1、 EIA STANDARD TP-80 Low Frequency Shielding Effectiveness Test Procedure for Electrical Connectors and Sockets EIA-364-80 NOVEMBER 2002 ELECTRONIC COMPONENTS, ASSEMBLIES RF amplifier; 200 ohm loads (non-inductive); 50 ohm loads; analyzer(s), network or spectrum; shielded cable and connectors; curren
2、t probe. 2.1.1.1 Signal source The signal source shall be an RF generator(s) with an output frequency of 10 kHz to 100 MHz, (method A), or 30 MHz to 500 MHz, (method B). The signal source may be part of the network analyzer(s). 2.1.1.2 RF amplifier The RF amplifier shall provide enough gain to raise
3、 the signal source output to 1 watt. 2.1.1.3 Analyzer The analyzer shall be either a network or spectrum analyzer and shall provide amplitude and may provide phase output from 10 kHz to 100 MHz, (method A), or 30 MHz to 500 MHz, (method B). More than one unit may be used to obtain the required plots
4、/values frequency band. 2.1.1.4 Data collection system This system shall be capable of interfacing with the analyzer and recording amplitude outputs. EIA-364-80 Page 4 2.2 Special test equipment, method A 2.2.1 Quadraxial test facility This shall be shielded metal enclosure with provisions for test
5、specimen installation, drive plate and guard coaxial chamber installation. See Appendix A for construction guidelines. 2.2.2 Connector test specimen adapters These shall be solid metal tubular backshells designed to interface with the connector accessory threads on one end and a type N plug on the o
6、ther. The adapters shall be designed to minimize transfer impedance contributions from other than the connector under test. The inner diameter shall be such that a 50 ohm characteristic impedance is maintained in conjunction with the center contact adapters; see annex A for fabrication details. 2.2.
7、3 Center contact adapters These shall be solid brass or bronze rods with the appropriate contacts on one end and type N socket on the other. 3 Test specimen 3.1 Preparation 3.1.1 Method A All wires shall be removed. Install the appropriate center contact adapters in both the plug and receptacle. Ins
8、tall the backshell adapters on the connectors. 3.1.1.1 Installation Install the test specimen in the inner chamber of the Quadraxial fixture, placing the current probe as shown in figure A-1. Tighten jam nuts to hold specimen in place. Connect one end of the adapter to the analyzer receiver channel
9、and terminate the other end to the 50 ohm load 3.1.1.2 Connection Connect instrumentation and measurement equipment as shown in figure 1. EIA-364-80 Page 5 Figure 1 - Test configuration 3.1.2 Method B 3.1.2.1 Panel mounted connector specimen Mount connector in the center of a bulkhead mounting plate
10、 (figure C.2, Annex C). Common all contacts as close to the bulkhead as possible. A 2 cm - 3 cm wide copper strap shall be soldered to the bulkhead mounting plate close to the CUT. The free end should be very close to the bussed panel mount connector conductors. The shield of a short probe cable sha
11、ll be soldered to the strap to form a low inductance path to the bulkhead, not the panel mount connector shield. This connection must be short as possible. Solder the center conductor to the bussed panel mount connector conductors. This connection should be as short as possible. Assemble the fixturi
12、ng of figure C.2. Be sure metal surfaces joined are cleaned for a good connection. ANALYZER EIA-364-80 Page 6 3.1.2.1.1 Measure specimen Zowith TDR A specimen connector shall be mounted on several inches of shielded cable. All conductors in this connector shall terminate a conductor in the cable. Ma
13、te to panel mounted connector specimen fixtured in 3.1.2.1. Use time domain reflectometry (TDR) to determine the voltage reflection coefficient of the mated pair. Estimate an average value if necessary. This will be used to calculate a correction factor (CF). 3.1.2.2 Cable mounted connector specimen
14、 Cut the cable on each specimen about 3 cm from the rear. Remove the outer jacket without nicking the braid, leaving 0.5 cm of outer jacket. Fold back braid over remaining outer jacket. Remove second foil shield, if present. Take care to avoid stressing the connector shield braid capture region. Str
15、ip all internal conductors, as close to the braid as practical. Twist together all internal conductors and solder. Cut soldered wires to 0.3 cm. Solder a non-inductive (metal film) termination resistor to this. Wrap this joint with high temperature insulating tape (Kapton). The termination resistor
16、shall be chosen to give a voltage reflection coefficient close to that measured in 3.1.2.1.1. Wrap the joint and termination resistor with copper tape. Dress the braid up over the copper tape. Wrap the braid and copper tape up to the free resistor lead with small gauge solid hook-up wire. Solder the
17、se together to prevent any RF leakage. 4 Test procedure 4.1 Method A With the connector and chamber configured per clause 3, the following procedure shall be performed. The test configuration shall be per figure A.1. 4.1.1 Calibration and initial set-up 4.1.1.1 Load calibration data (see Appendix B)
18、 into data processor system. 4.1.1.2 Connect instrumentation per Figure 1 (below). Adjust system gains and attenuations to ensure that all system elements are operating within their dynamic ranges. This is done by performing trial sweeps until satisfactory performance is obtained. 4.1.2 Specimen tra
19、nsfer impedance measurements 4.1.2.1 Set low frequency limit to 10 kHz maximum. 4.1.2.2 Set upper frequency limit to 100 MHz minimum. 4.1.2.3 Perform transfer impedance measurement. Measurements shall be taken from 10 kHz to 100 MHz. a minimum of 10 points per decade shall be sampled. When more than
20、 one instrument is required, the measurement bands shall overlap by 25% minimum. EIA-364-80 Page 7 4.1.2.4 Plot Z() for test specimen. Measured results shall be reported in dB from one ohm, )ohm 1( / log 20dBZ= 4.2 Method B 4.2.1 Do a full 2 port calibration if using a vector network analyzer for im
21、proved accuracy. 4.2.2 Collect a noise floor plot; see C.1.4). 4.2.3 Prepare panel mounted connector specimen(s); see 3.1.2.1. 4.2.4 Measure specimen Zo; see 3.1.2.1.1. NOTE Note the voltage reflection coefficient (p) for use in 4.2.8. 4.2.5 Prepare cable mounted specimens; see 3.1.2.2.2. 4.2.6 Coll
22、ect a reference measurement; see C.1.1. Place a cable mounted specimen in the fixturing of figure C.1. Solder the electrical connection at the right so that pressure is applied at the mechanical connection at the left (the fixture connection to the mating interface). Measure S21(reference) expressed
23、 in dB, or measure the power at the spectrum analyzer in dBm. The frequency range shall be 30 MHz to 500 MHz. 4.2.7 Collect the specimen measurements; see C.1.2. With the same cables, substitute the specimen measurement fixturing of figure C.2 for the reference measurement fixturing of 4.2.6 (figure
24、 C.1). Mate the connectors under test and connect to the type “N“ connector at the right. Measure S21(specimen) expressed in dB, or measure the power at the spectrum analyzer in dBm while driving the specimen with the same forward power (into fixture port 1, figure C.2 ) as used in 4.2.6. 4.2.8 Calc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ECAEIA364802002TP80LOWFREQUENCYSHIELDINGEFFECTIVENESSTESTPROCEDUREFORELECTRICALCONNECTORSANDSOCKETSTP80

链接地址:http://www.mydoc123.com/p-704314.html