1、A-A-53807 66 m 7979774 001L127 O m A-A-5 3 8 O 7 18 January 1989 COMMERCIAL ITEM DESCRIPTION REGULATOR, PRESSURE, MEDICAL GAS ADMINISTRATION APPARATUS The General Services Administration has authorized the use of this commercial item description as a replacement for MIL-R-36557C, which is cancelled.
2、 This commercial item description covers the requirements for a single stage pressure regulator, with a Thorpe tube flowmeter suitable for use with oxygen cylinders and therapy equipnients. Salient characteristics: General requirements. The regulator shall conform to the applicable requirements of C
3、GA E-7, Standard for Flowmeters, Pressure Reducing Regulators, Regulator/Flowmeter and Regulator/Flowgauge combination for the Administration of Medical Gases, and UL 404, Standard for Safety, High Pressure Gas Gauges. 0 Order of precedence. In the event of a conflict between the requirements of CGA
4、 E-7 and UL 404, and the requirements described herein, the requirements specified in this document shall take precedence. Material. The body of the regulator shall be fabricated from forged or machined copper alloy, nickel plated or chromium plated over nickel plated. There shall be no internal pla
5、ting of the regulator body. Alloys shall be tempered and shall have a minimum tensile strength of 50,000 pounds per square inch. All external brass or copper alloy surfaces, such as gland nut, standard threaded fitting for low pressure oxygen connection and captive tubing adapter shall be nickel pla
6、ted or chromium plated over nickel plated. AMSC N/A FSC 6680 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. a THIS DOCUMENT CONTAINS 4 PAGES. - Licensed by Information Handling ServicesA-A-53807 bb m 9797974 0011i30 7 m A- A-53807 Construction, style and design. Th
7、e regulator shall consist essentially of a body, a pressure reducing valve, an inlet pressure gage, a Thorpe tube flow indicating device with a flow control valve for manually controlling the rate of Pow pressure flow, a standard externally-threaded fitting for connection to low pressure oxygen devi
8、ces and a captive tubing adapter. The general style and design of the complete unit shall be in accordance with Figure 1. Regulator. Regulator shall be of the single stage type which shall maintain uniform discharge rate by means of a sensitive diaphragm. The outlet pressure of the requlator shall b
9、e preset at 49 + 1 pounds per square inch gage (psig). A self- reseating pressure relief valve shall be incorporated to prevent accumulation o hazardous pressure within the regulator which shall open at 130 psig maximum and reseat at 85 psig minimum. The regulator shall be equipped with a commercial
10、 union nut and gland inlet connection which shal be capable of making connection with suitable oxygen cylinders having a 0.903-14 NGO right-hand thread on the outlet valve in accordance with connection No. 540 of ANSI B57.1. The regulator inlet shall have a suitable assembled filter, fabricated from
11、 bronze, to prevent foreign matter from entering the device. The regulator shall exhibit no leakage as evidenced by formation of bubbles, when a film of soapy water (grease free) is applied to all junctions of the reguator. The internal pressure applied to the regulator shall be at least 2000 psig a
12、nd the flowmeter valve shall be completely closed (no output) . Pressure gage. The regulator shall be equipped with a pressure gage suitable for pressure measurement in the range f 0-4000 psig. pressure and content. The pressure shall be marked at 100 psig and numbered at every 500 psig. The content
13、 shall be indicated as 1/4, 1/2, 3/4 and full The face cover of the gage shall be fabricated from a clear, durable plastic. The gage shall be calibrated to indicate both Flow indicating device. The flow indicating device shall be a Thorpe tube flowmeter suitable for accurate measurement of oxygen fl
14、ow in the range of 1 to 15 liters per minute (LPM). The flowmeter shall be calibrated in the range of 1 to 5 LPM on the lower half of the tube and 5 to 15 LPM in the upper half of the tube. At pressure of 50 psig and temperature of 700 Fahrenheit, the measurement shall be accurate to + 1/2 LPM in th
15、e 1-5 LPM range and - + 1 LPM in the 5-15 LPM range. 2 Licensed by Information Handling ServicesA-A-53807 66 m 7777774 0012231 7 m A-A-53807 A clear flowmeter shield shall be provided over the flowtube. Both the flowtube and the shield shall be fabricated from impact-resistant plastic. The flowmeter
16、 shal possess a pressure relief device or feature that will prevent the flowmeter (flowtube) from bursting in the event of overpressurization. The flowmeter valve, when tested at O degree and 110 degrees Fahrenheit, shall meet the following performance criteria: a. Joints and shaft seals shall not l
17、eak when subjected to a pressure equal to 1.5 times the rated pressure of the flowmeter. b. A shut-off torque of no more than 8 inch-pound shall be required.to achieve a bubble tight seal at the outlet of the valve when the inlet of the valve is pressurized to the rated pressure of the flowmeter. c.
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA538071989REGULATORPRESSUREMEDICALGASADMINISTRATIONAPPARATUS 医学 气体 管理 仪器 压力 调整器 PDF

链接地址:http://www.mydoc123.com/p-701685.html