1、A-A-535 O 1 11 May 1988 AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited, - COMMERCIAL ITEM DESCRIPTION STAPLE UNIT, SURGICAL, THORACIC-ABDOMINAL The General Services Administration has authorized the use of this commercial item description. This com
2、mercial item description covers seven styles of Thoracic -abdominal surgical staple units. Salient characteristics: Style and design. Style and design of staple units shall conform to Figure 1. Staple units shall consist of the following: Staple cartridge. The staple cartridge shall contain staple c
3、hannels, a drive mechanism designed to engage the compatible stapler and drive the staples toward the anvil, a spring-loaded alignment pin, and a slotted alignment pin spring. The number of staple channels (staples) shall be as specified in Table II. Except as specified in the “Notes“ section, the c
4、hannels shall be arranged in a double-staggered row (See Figure 1). Each channel shall be loaded with a surgical staple. Anvil. The anvil shall contain a mated set of staple- clinching pockets.and an engagement hook and detent stop, as shown in Figure 1. Hinge. the anvil and cartridge. The hinge sha
5、ll be a spring-loaded hinge connecting Dimensions. For each NSN, staple dimensions shall be as indicated in Table I. The length of the staple-channel pattern and the number of channels shall be as indicated in Table II. 1 , THIS DOCUMENT CONTAINS 7 PAGES. p. Provided by IHSNot for ResaleNo reproduct
6、ion or networking permitted without license from IHS-,-,-A-A-53501 NSN -No.taples Stase-Pattern (1) (2) 6 !5-L50-1T0-7=819 1 31 6515-01-209-8220 11 31.5 6515-01-207-8222 33 91.5 Length 6m-0ig2o-822i 11 31.5 Table I. Surgical Staple Dimensions NCN (a) Staple- 515-01-207-8220 4.8 4.0 0.28 515-01-207-8
7、221 3.5 4.0 0.23 515-81-207-8222 4.8 4.0 0.28 515-01-207-8223 3.5 4.0 0.23 515-01-207-8224 3.5 4.0 0.23 515-01-207-8225 4.8 4.0 9.28 (a) All dimensions in mm. (1) Tolerance: .E 0.05 mm. (2) Tolerance: 0.005 mm. Table II, Staple Unit Dimensions 5515-01-207-8225 19 53 E _.I_ (1) Length in mm. (2) Tole
8、rance: 0.05 rnm. Matsialc. Staples shall be fabricated from corrosion-resisting stsol wire that coneorms to ACTM Designation F138, Grads 2 or ASTM Desagnation F55, Grade 2. Staple units ahall be Eabricate from a suitable corrosion-resisting steel and a suitable r igfd plastic. Performance. Staple un
9、it shall be compatible with the stapler indicated in Table III. Staple units shall be suitable for use in abdominalt gynecologicall pediatric, and thoracic surgical pmceduxec. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-53501 b5 m 7777774 0
10、030748 T m A-A-535 O 1 * I c 0 TABLE III. Stapler - Staple Unit Compatibility 1 NSN: Staple Unit NSN: Stapler 6515-01-207-8219 6515-01-204-5365 6515-01-207-8220 6515-01-204-5365 6515-01-207-8221 6515-01-204-5365 6515-01-207-8222 6515-01-204-5362 6515-01-207-8223 6515-01-204-5362 6515-01-207-8224 651
11、5-01-204-5364 b 6515-01-207-8224 6515-01-204-5364 Sterility. Staple units shall be sterile in accordance with the latest edition of the United States Pharmacopiea (USP). Workmanship. from their appearance or impair their serviceability. - Unit. Each (PG) . One package containing tweIve individually
12、packaged sterile stapling units, as specified, constitutes one unit. The individual package shall be a sealed peel-open type container capable of maintaining sterility of contents unless the package is damaged or opened. Staple units shall be free from defects that detract Quality Assurance Provisio
13、ns. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of al1 inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or an
14、y facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and servi
15、ces conform to prescribed requirements. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Governments request, at any time, or from time to time, during the-performance of the contract and for a pe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA535011988STAPLEUNITSURGICALTHORACICABDOMINAL 胸腹 外科 原材料 PDF

链接地址:http://www.mydoc123.com/p-701676.html