BS 7920-2005 Luminance meters - Requirements and test methods《亮度计 要求和试验方法》.pdf
《BS 7920-2005 Luminance meters - Requirements and test methods《亮度计 要求和试验方法》.pdf》由会员分享,可在线阅读,更多相关《BS 7920-2005 Luminance meters - Requirements and test methods《亮度计 要求和试验方法》.pdf(20页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD BS 7920:2005 Luminance meters Requirements and test methods ICS 17.180.20; 29.140.40 BS 7920:2005 This British Standard was published under the authority of the Standards Policy and Strategy Committee on 28 January 2005 BSI 28 January 2005 First published August 1998 Second edition
2、January 2005 The following BSI references relate to the work on this British Standard: Committee reference CPL/34/5 Special announcement in Update Standards September 2004 ISBN 0 580 44392 2 Committees responsible for this British Standard The preparation of this British Standard was entrusted by Te
3、chnical Committee CPL/34, Lamps and related equipment, to Subcommittee CPL/34/5, Photometry and light classifications, upon which the following bodies were represented: Chartered Institution of Building Services Engineers GAMBICA (BEAMA Ltd.) Institution of Civil Engineers Institution of Electrical
4、Engineers (IEE) Institution of Lighting Engineers Lighting Industry Federation Ltd. National Physical Laboratory Co-opted members Amendments issued since publication Amd. No. Date CommentsBS 7920:2005 BSI 28 January 2005 i Contents Page Committees responsible Inside front cover Foreword ii Introduct
5、ion 1 1S c o p e 1 2 Normative references 1 3T e r m s a n d d e f i n i t i o n s 1 4 Classification 3 5 Digital ranges and readouts 3 6 Analogue scales 4 7 Calibration and testing temperature 4 8 Power supply 4 9 Performance requirements for laboratory and field luminance meters 4 10 Marking 5 Ann
6、ex A (informative) Guidance on choice of meter range 6 Annex B (normative) Calibration 6 Annex C (normative) Test methods 7 Bibliography 15 Figure 1 Diagram showing characteristic ray paths in a luminance meter 3 Figure C.1 Co-ordinates for calculation of the directional response function f 2 (,)1 0
7、 Figure C.2 Diagram showing gloss traps for determining surrounding field error 11 Table 1 Test and calibration methods 4 Table 2 Error tolerances for luminance meters 5 Table C.1 Example of calculation of correction factor for high pressure sodium lamp 14BS 7920:2005 ii BSI 28 January 2005 Foreword
8、 This British Standard has been prepared by Subcommittee CPL/34/5. Together with BS EN 13032-1:2004, it supersedes BS 7920:1998, which is withdrawn. Luminance meters are used for measurements relating to many varied applications, such as: road lighting; road tunnel entrances; cathode ray tubes and o
9、ther display screens; light boxes for examining X-ray and other transparencies; road and advertising signs; signal lamps such as traffic lights. In many of these applications luminance levels are specified. The approach of this standard is similar to that used for BS 667:2005 on illuminance meters.
10、This British Standard specifies the requirements for two types of luminance meter: Type L: laboratory instruments and Type F: field instruments. The error tolerances have been considered in a similar manner to that used in CIE Publication 69 1, and the definitions are based on CIE Publication 17.4 2
11、. Error tolerances for Type L meters have been aligned with those given in BS EN 13032-1:2004, Table 4. BS EN 13032-1 specifies laboratory luminance meters for testing luminaires only. It does not cover luminance meters for other laboratory purposes or for field use. This new edition of BS 7920 inco
12、rporates changes necessitated by the publication of BS EN 13032-1:2004. It does not reflect a full review or revision of the standard which will be undertaken in due course. This publication does not purport to include all the necessary provisions of a contract. Users are responsible for its correct
13、 application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover, pages 1 to 15 and a back cover. The BSI copyright notice displayed in this document indicates when the document w
14、as last issued.BS 7920:2005 BSI 28 January 2005 1 Introduction As lighting levels reduce, the response characteristics of the human eye change from photopic during the day through an intermediate mesopic region at dusk to a scotopic response at night time. At present, eye response functions have not
15、 been defined in the mesopic region; research work is still in progress on this subject. Although photopic measurements are generally only applicable at lighting levels above about 2 cd m p2 , in practice lighting levels for installations are specified at values significantly below this. It is there
16、fore necessary to make luminance measurements with known accuracy at these lower levels to check performance against such specifications whilst accepting that the measurements do not accurately represent the visual response. Photometric measurements, even under laboratory conditions, are more diffic
17、ult to determine precisely than many other kinds of physical measurement, and in the case of luminance meters, liability to error is increased by the fact that many of the essential components of such instruments are susceptible to variation with time and use. These errors can be minimized by care i
18、n the design of the instrument as a whole, and also by the user in not exposing it to deleterious conditions of temperature, illumination or atmosphere. 1 Scope This British Standard specifies performance requirements for luminance meters for the measurement of photopic luminance for applications ot
19、her than the measurement of luminaires. It specifies the performance requirements for two types of luminance meter, Type L (laboratory instruments) and Type F (field instruments). This standard is intended for use by meter manufacturers and users. NOTE The requirements specified for a Type L meter i
20、n this standard are identical to the requirements specified for a luminance meter in BS EN 13032-1. However, BS EN 13032-1 specifies additional requirements which are not specified in this standard. 2 Normative references The following referenced documents are indispensable for the application of th
21、is document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. BS ISO 10526:1999, CIE standard illuminants for colorimetry. PD 6461-3:1995, General metrology Part 3: Guide to the expression
22、of uncertainty in measurement (GUM). CIE Publication 18.2:1983, The basis of physical photometry. Second edition. 3 Terms and definitions For the purposes of this British Standard the following terms and definitions apply. 3.1 luminance (in a given direction, at a given point of a real or imaginary
23、surface) (L v ; L) quantity defined by the formula: where CIE 17.4:1987, definition 845-01-35 d v is the luminous flux transmitted by an elementary beam passing through the given point and propagating in the solid angle d containing the given direction; dA is the area of a section of that beam conta
24、ining the given point; is the angle between the normal to that section and the direction of the beam; unit: cd m p2= lm m p2 sr p1 v v d dc o sd L A = BS 7920:2005 2 BSI 28 January 2005 3.2 luminance meter instrument for measuring luminance 3.3 photoelectric detector detector of optical radiation in
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS79202005LUMINANCEMETERSREQUIREMENTSANDTESTMETHODS 亮度计 要求 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-547168.html