1、BRITISH STANDARD BS 4376:1982 Incorporating Amendment No. 1 Specification for Electrically operated blood storage refrigerators UDC 621.565.92-83:615.382BS4376:1982 This British Standard, having been prepared under the directionof the Refrigeration, Heating and Air Conditioning Standards Committee,
2、was published under the authority ofthe Board of BSI and comes into effect on 30 September 1982 BSI 08-1999 First published December 1968 First revision September 1982 The following BSI references relate to the work on this standard: Committee reference RHE/19 Draft for comment 81/71454 DC ISBN 0 58
3、0 12671 4 Cooperating organizations The Refrigeration, Heating and Air Conditioning Standards Committee, under whose direction this British Standard was prepared, consists of representatives from the following: Association of Consulting Engineers Association of Manufacturers of Domestic Electrical A
4、ppliances Boiler and Radiator Manufacturers Association Limited British Combustion Equipment Manufacturers Association British Gas Corporation British Refrigeration and Air Conditioning Association* Building Services Research and Information Association Chartered Institution of Building Services Dep
5、artment of Energy (Energy Technology) Department of Health and Social Security* Department of the Environment (PSA) Electricity Supply Industry in England and Wales Heating and Ventilating Contractors Association* Hevac Association Institute of Energy Institute of Refrigeration* Institution of Gas E
6、ngineers Lloyds Register of Shipping Manufacturers Association of Radiators and Convectors Ltd. Ministry of Defence* National Coal Board Society of British Gas Industries Water-tube Boilermakers Association The organizations marked with an asterisk in the above list, together with the following, wer
7、e directly represented on the Technical Committee entrusted with the preparation of this British Standard: Agricultural Research Council Co-operative Union Environmental Health Officers Association Ice Cream Federation Limited Institution of Grocery Distribution London Refrigeration Society National
8、 Blood Transfusion Service UK Association of Frozen Food Producers Amendments issued since publication Amd. No. Date of issue Comments 6326 April 1991 Indicated by a sideline in the marginBS4376:1982 BSI 08-1999 i Contents Page Cooperating organizations Inside front cover Foreword ii Section 1. Gene
9、ral 1 Scope 1 2 Definitions 1 Section 2. Particular requirements 3 Construction 1 4 Refrigerating system 1 5 Electrical equipment 1 6 Instruments and control equipment 1 7 Alarm systems 2 8 Safety of personnel in cold rooms 2 Section 3. Performance requirements 9 Temperature limits 2 Section 4. Test
10、s 10 Text deleted 2 11 Tests for walk-in cold rooms 2 Section 5. Marking and information 12 Marking 3 13 Information 3 Publications referred to Inside back coverBS4376:1982 ii BSI 08-1999 Foreword This revision of BS 4376:1968 has been prepared under the direction of the Refrigeration, Heating and A
11、ir Conditioning Standards Committee to cover all sizes of refrigerator, both “reach-in” and “walk-in” types, designed and equipped for the preservation of whole blood and fluid blood plasma and its derivatives that need to be kept within specified limits of temperature. The refrigerators are intende
12、d for use in hospitals and blood collecting centres. Refrigerators for the storage of frozen blood products are excluded. A revision of the requirements for closed reach-in type cabinets has been prepared and issued as BS4376-1:1991 and the requirements for closed reach-in type cabinets have been de
13、leted by amendment1. It is envisaged that further Parts will be issued to cover other types of blood storage refrigerators. The standard specifies particular requirements for the proper storage of unfrozen blood when in refrigerators that have been constructed according to BS2501 and BS2502. BS 4376
14、:1968 is withdrawn. A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. Summary of pages This
15、 document comprises a front cover, an inside front cover, pagesi andii, pages1 to4, an inside back cover and a back cover. This standard has been updated (see copyright date) and may have had amendments incorporated. This will be indicated in the amendment table on the inside front cover.BS4376:1982
16、 BSI 08-1999 1 Section 1. General 1 Scope This British Standard specifies particular requirements for sectional cold rooms of the walk-in type, constructed according to BS2502, which are intended for the preservation of whole blood and fluid blood plasma and its derivatives at temperatures between4
17、C and6 C. The refrigerators are for use in climates having dry bulb temperatures and relative humidities, respectively, from16 C and80% to40 C and40%. Closed reach-in type cabinets are covered by BS4376-1. It does not apply to refrigerators used for the storage of frozen blood plasma. It does not ap
18、ply to low temperature refrigerators or compartments used for the preservation of anti-sera, reagents, etc. NOTE 1It is recommended that, except for very small storage capacity requirements, separate refrigerators should be provided for stock blood and for cross-matched blood reserved for particular
19、 individuals. NOTE 2The titles of the publications referred to in this standard are listed on the inside back cover. 2 Definitions For the purposes of this British Standard the definitions given in BS2501 and BS2502 apply. Section 2. Particular requirements 3 Construction 3.1 Shelves. Where revolvin
20、g shelves are fitted, the distance from the periphery to the adjacent surface of the supporting centre column shall be not more than330mm. All shelves shall be capable of sustaining a loading of175kg/m 2without undue deflection. 3.2 Lighting 3.2.1 Text deleted 3.2.2 Walk-in cold rooms. Refrigerators
21、 of the walk-in type shall be provided with internal lighting. Fittings shall be of the moisture-proof type. The lighting circuit shall be controlled by a switch to be supplied and fitted to the refrigerator immediately adjacent to the main door, and an indicator light shall be supplied and fitted e
22、xternally. 3.3 Doors 3.3.1 General. Door sealing gaskets shall have an effective seal over the entire length. 3.3.2 Text deleted 3.3.3 Walk-in cold rooms. It shall be possible to open the door(s) of walk-in refrigerators from the inside, even when locked from outside. Means shall be provided for loc
23、king each door, and two keys shall be supplied for each lock. 4 Refrigerating system 4.1 General. The refrigerating system shall comply with the requirements of clause10 of BS2501:1976, together with the requirements, as follows, of4.2 to4.5. 4.2 Noise. The refrigerating system shall not produce und
24、ue noise or vibration. 4.3 Defrosting. The refrigerating system shall be designed to ensure that under normal conditions the evaporator, drip tray and adjacent cold surfaces are automatically defrosted with sufficient regularity to maintain the performance requirements of this standard. In the case
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS43761982SPECIFICATIONFORELECTRICALLYOPERATEDBLOODSTORAGEREFRIGERATORS 电动 血液 冷藏箱 规范 PDF

链接地址:http://www.mydoc123.com/p-544372.html