《ASTM F2082-2015 Standard Test Method for Determination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery《通过弯曲和自由恢复测定镍钛形状记忆合金转变温度的标准试验方法.pdf》由会员分享,可在线阅读,更多相关《ASTM F2082-2015 Standard Test Method for Determination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery《通过弯曲和自由恢复测定镍钛形状记忆合金转变温度的标准试验方法.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2082 15Standard Test Method forDetermination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery1This standard is issued under the fixed designation F2082; the number immediately following the designation indicates the year oforiginal adoption
2、 or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes a procedure for determiningthe martensite-to-aust
3、enite transformation temperatures of ei-ther fully annealed or heat-treated nickel titanium alloys bymeasuring the deformation recovered during the thermal trans-formation.1.2 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach
4、system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith this standard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is th
5、eresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE220 Test Met
6、hod for Calibration of Thermocouples ByComparison TechniquesE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodF2005 Terminology for Nickel-Titanium Shape MemoryAlloys3. Terminology3.1 DefinitionsSpecific technical terms used in this testmethod are found
7、in Terminology F2005.3.2 free recoveryunconstrained motion of a shapememory alloy upon heating and transformation to austeniteafter deformation in a lower temperature phase.3.3 Abbreviations:3.3.1 LVDTlinear variable differential transducer.3.3.2 RVDTrotary variable differential transducer.4. Summar
8、y of Test Method4.1 This test method involves cooling a test specimen to itsnominally fully martensitic phase, deforming the specimen,and heating the specimen to its fully austenitic phase. Duringheating, the motion of the specimen is measured and plottedversus the specimen temperature. For a two-st
9、agetransformation, the Rs, Rf, As, and Af, as defined in Terminol-ogy F2005, are determined. For a single-stage transformation,the Asand Afare determined.5. Significance and Use5.1 This test method provides a rapid, economical methodfor determination of transformation temperatures.5.2 Measurement of
10、 the specimen motion closely parallelsmany shape memory applications and provides a result that isapplicable to the function of the material.5.3 This test method uses wire, tube, or strip samples; thus,it is able to provide an assessment of the product in itssemifinished form.5.4 This test method ma
11、y be used on annealed samples todetermine the transformation temperatures and assure the alloyformulation, since chemical analysis is not precise enough todetermine adequately the nickel-to-titanium ratio of shapememory alloys.5.5 Transformation temperatures derived from this testmethod may differ f
12、rom those derived from other methods as aresult of the effects of strain and load on the transformationtemperature.5.6 The test method is applicable to shape memory alloyswith Aftemperatures in the range of approximately 25 to+90C.1This test method is under the jurisdiction of ASTM Committee F04 on
13、Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved May 1, 2015. Published July 2015. Originally approvedin 2001. Last previous edition approved in 2006 as F2082 06, which waswithdrawn in January 2015 and
14、reinstated in May 2015. DOI: 10.1520/F2082-15.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM
15、International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States16. Apparatus6.1 LVDT, with range greater than half the mandrel diameter(see 9.2), with power supply, mounted in an appropriate fixturewith counterbalanced probe (see Fig. 1); or RVDT with rangegreater
16、than 45, with power supply, mounted in an appropriatefixture (see Fig. 2); or a vision system; or equivalent means ofmeasuring sample displacement.6.2 Thermocouple and Indicator, with resolution of 0.1C(0.2F) or better.6.3 XY Chart Recorder, or equivalent manual or automateddata acquisition system.6
17、.4 Hot Plate and Stirrer.6.5 Bath of Heat Transfer Fluid, for example, denaturedalcohol, ethylene glycol, water, and so forth.6.6 Mandrel, for deforming the sample in the martensiticstate.6.7 Fixture, for holding the sample during recovery.6.8 Liquid Nitrogen, or dry ice.7. Sampling7.1 Test specimen
18、 can be a wire, tube, or strip with diameteror thickness in the range of 0.3 to 3.0 mm (0.012 to 0.12 in.).For test systems that do not contact the specimen (for example,vision system), the diameter or thickness may be less than 0.3mm.7.2 Specimens may be tested in the semifinished (heat-treated) or
19、 annealed condition. Anneal is defined in Terminol-ogy F2005.8. Calibration8.1 The thermocouple and indicator shall be kept in acalibrated condition, traceable to the National Institute forStandards and Technology or appropriate National MetrologyInstitute that successfully participates in relevant
20、internationalinterlaboratory comparisons.8.2 The thermocouple shall be calibrated using Test MethodE220.9. Procedure9.1 For alloys that are superelastic at room temperature,cool a bath of appropriate heat transfer fluid to 55C (67F)or lower using liquid nitrogen, dry ice, or other suitablemethod. Fo
21、r alloys that are martensitic at room temperature,cool the bath to 10C (50F) or lower.9.2 Select a mandrel according to the sample diameter orthickness to give an outer fiber strain of 2 to 2.5 %. For thesestrains, mandrel diameter shall be between 39 and 49 timesspecimen diameter or thickness.9.3 C
22、ut a test specimen long enough to wrap 90 to 180around the mandrel.9.4 Place the recovery fixture and the mandrel, along withthe test specimen, in the bath and wait a minimum of 3 min forthe fixtures to equilibrate to the bath temperature.9.5 Deform the specimen in the bath by wrapping it 90 to180 a
23、round the mandrel.9.6 Place the specimen on a fixture (recovery fixture) thatholds the sample so as not to interfere with the free recovery ofthe specimen on heating.FIG. 1 Schematic Showing Side View of Test Apparatus Using a Vertically Mounted and Counterbalanced LVDT(LVDT Power Supply, Thermocoup
24、le Indicator, and Data Acquisition System are not shown)F2082 1529.7 Remove the mandrel from the bath. Alternatively, themandrel can be attached to the recovery fixture and left in thebath. In this case, the thermal mass of the mandrel and fixtureshall be such that the temperature of the fixture and
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF20822015STANDARDTESTMETHODFORDETERMINATIONOFTRANSFORMATIONTEMPERATUREOFNICKELTITANIUMSHAPEMEMORYALLOYSBYBENDANDFREERECOVERY

链接地址:http://www.mydoc123.com/p-536509.html