ASTM E1647-2016 Standard Practice for Determining Contrast Sensitivity in Radiology《测定放射医学中对比敏感度的标准实施规程》.pdf
《ASTM E1647-2016 Standard Practice for Determining Contrast Sensitivity in Radiology《测定放射医学中对比敏感度的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1647-2016 Standard Practice for Determining Contrast Sensitivity in Radiology《测定放射医学中对比敏感度的标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1647 16Standard Practice forDetermining Contrast Sensitivity in Radiology1This standard is issued under the fixed designation E1647; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number
2、in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the design and material selection ofa contrast sensitivity measuring gauge used to determine theminimum change in materia
3、l thickness or density that may beimaged without regard to unsharpness limitations.1.2 This practice is applicable to transmitted-beam radio-graphic imaging systems (film, radioscopy, computedradiography, and digital detector array image detectors) utiliz-ing X-ray and gamma ray radiation sources.1.
4、3 The values stated in inch-pound units are to be regardedas standard. The SI units given in parentheses are for informa-tion only.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establi
5、sh appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific safetystatements, see NIST/ANSI Handbook 114 Section 8, Code ofFederal Regulations 21 CFR 1020.40 and 29 CFR 1910.96.2. Referenced Documents2.1 ASTM Standards:2B139/B139M
6、 Specification for Phosphor Bronze Rod, Bar,and ShapesB150/B150M Specification for Aluminum Bronze Rod, Bar,and ShapesB161 Specification for Nickel Seamless Pipe and TubeB164 Specification for Nickel-Copper Alloy Rod, Bar, andWireB166 Specification for Nickel-Chromium-Iron Alloys (UNSN06600, N06601,
7、 N06603, N06690, N06693, N06025,N06045, and N06696), Nickel-Chromium-Cobalt-Molybdenum Alloy (UNS N06617), and Nickel-Iron-Chromium-Tungsten Alloy (UNS N06674) Rod, Bar, andWireE747 Practice for Design, Manufacture and Material Group-ing Classification of Wire Image Quality Indicators (IQI)Used for
8、RadiologyE1025 Practice for Design, Manufacture, and MaterialGrouping Classification of Hole-Type Image Quality In-dicators (IQI) Used for RadiologyE1255 Practice for RadioscopyE1316 Terminology for Nondestructive ExaminationsE1411 Practice for Qualification of Radioscopic SystemsE1734 Practice for
9、Radioscopic Examination of CastingsE1742 Practice for Radiographic ExaminationE2002 Practice for Determining Total Image Unsharpnessand Basic Spatial Resolution in Radiography and Radios-copyE2445 Practice for Performance Evaluation and Long-TermStability of Computed Radiography Systems2.2 Federal S
10、tandards:321 CFR 1020.40 Safety Requirements for Cabinet X-raySystems29 CFR 1910.96 Ionizing Radiation2.3 NIST/ANSI Standards:NIST/ANSI Handbook 114 General Safety Standard forInstallations Using Non-Medical X-ray and SealedGamma Ray Sources, Energies to 10 MeV42.4 ISO Standard:5ISO 192325 Duplex Wi
11、re Image Quality Indicator2.5 Other Standards:EN 462 5 Duplex Wire Image Quality Indicator(withdrawn, replaced by ISO 192325)6EN 130681 Radioscopic Testing-Part 1: Qualitative Mea-surement of Imaging Properties61This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing
12、 and is the direct responsibility of Subcommittee E07.01 onRadiology (X and Gamma) Method.Current edition approved June 1, 2016. Published July 2016. Originally approvedin 1994. Last previous edition approved in 2009 as E1647 09. DOI: 10.1520/E1647-16.2For referenced ASTM standards, visit the ASTM w
13、ebsite, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop
14、: SDE, Washington, DC 20401, http:/www.access.gpo.gov.4Available fromAmerican Society for NondestructiveTesting (ASNT), P.O. Box28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http:/www.asnt.org.5Available from International Organization for Standardization (ISO), ISOCentral Secretariat, BIBC II
15、, Chemin de Blandonnet 8, CP 401, 1214 Vernier,Geneva, Switzerland, http:/www.iso.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13. Terminology3.1 DefinitionsDefinitions of terms applicable to this testmethod may be found in Term
16、inology E1316.4. Summary of Practice4.1 It is often useful to evaluate the contrast sensitivity of apenetrating radiation imaging system separate and apart fromunsharpness measurements. Conventional image quality indi-cators (IQIs), such as Test Method E747 wire and PracticesE1025 or E1742 plaque IQ
17、Is, combine the contrast sensitivityand resolution measurements into an overall performancefigure of merit, other methods such as included in PracticeE2002 do not address contrast specifically. Such figures ofmerit are often not adequate to detect subtle changes inimaging system performance. For exa
18、mple, in a high contrastimage, unsharpness can increase with almost no noticeableeffect upon overall image quality. Similarly, in an applicationin which the imaging system provides a very sharp image,contrast can fade with little noticeable effect upon the overallimage quality. These situations ofte
19、n develop and may gounnoticed until the system performance deteriorates belowacceptable image quality limits.5. Significance and Use5.1 The contrast sensitivity gauge measures contrast sensi-tivity independent of the imaging system spatial resolutionlimitations. The thickness recess dimensions of th
20、e contrastsensitivity gauge are large with respect to the unsharpnesslimitations of most imaging systems. Four levels of contrastsensitivity are measured: 4 %, 3 %, 2 %, and 1 %.5.2 The contrast sensitivity gauge is intended for use inconjunction with a high-contrast resolution measuring gauge,such
21、as Practice E2002, ISO 19232 5 Duplex Wire ImageQuality Indicator7, or a line-pair gauge. Such gauges measuresystem unsharpness essentially independent of the imagingsystems contrast sensitivity. Such measurements are appropri-ate for the qualification and performance monitoring of radio-graphic and
22、 radioscopic imaging systems with film, realtimedevices, Computed Radiography (CR) and Digital DetectorArrays (DDA).5.3 Radioscopic/radiographic system performance may bespecified by combining the measured contrast sensitivity ex-pressed as a percentage with the unsharpness expressed inmillimetres o
23、f unsharpness. For the duplex wire image qualityindicator, the unsharpness is equal to twice the wire diameter.For the line pair gauge, the unsharpness is equal to thereciprocal of the line-pair/mm value. As an example, animaging system that exhibits 2 % contrast sensitivity andimages the 0.1 mm pai
24、red wires of the duplex wire IQI(equivalent to imaging 5 line-pairs/millimeter resolution on aline-pair gauge) performs at a 2 %0.2 mm sensitivity level. Astandard method of evaluating overall radioscopic systemperformance is given in Practice E1411 and in EN 130681. Aconversion table from duplex wi
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16472016STANDARDPRACTICEFORDETERMININGCONTRASTSENSITIVITYINRADIOLOGY 测定 放射 医学 对比 敏感度 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-529136.html