ASTM E1390-2012 Standard Specification for Illuminators Used for Viewing Industrial Radiographs《观察工业X光照片用照明装置的标准规范》.pdf
《ASTM E1390-2012 Standard Specification for Illuminators Used for Viewing Industrial Radiographs《观察工业X光照片用照明装置的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1390-2012 Standard Specification for Illuminators Used for Viewing Industrial Radiographs《观察工业X光照片用照明装置的标准规范》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1390 12Standard Specification forIlluminators Used for Viewing Industrial Radiographs1This standard is issued under the fixed designation E1390; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisio
2、n. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 The function of the illuminator is to provide sufficientillumination and viewing capabilities for the purpose of iden-tification
3、and interpretation of radiographic images. This speci-fication provides the recommended minimum requirements forIndustrial Radiographic Illuminators used for viewing indus-trial radiographic films using transmitted light sources.1.2 The illuminator has to ensure the same safety forpersonnel, or user
4、s of any electric apparatus, as specified byelectrical standards applicable in the country in which theilluminator is used.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro
5、-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1.4 Values stated in SI units are to be regarded as thestandard. The values given in parentheses are provided forinformation only.2. Referenced Documents2.1 ASTM Standards:2E1316 Terminology f
6、or Nondestructive Examinations3. Terminology3.1 Definitions:3.1.1 diffusing screenthe screen that scatters the light suchthat the light appears the same brightness from any viewingangle.3.1.2 viewing screenthe port through which the light isprojected.3.2 For additional definitions of terms used in t
7、his specifi-cation, see Terminology E1316.4. Ordering Information4.1 This specification is intended to be used by the manu-facturers and purchasers of radiographic illuminators. Require-ments, if imposed on manufacturers, should be established bycontractual agreement or appropriate purchase document
8、.5. Materials and Manufacture5.1 GeneralThe illuminator shall consist of a housingwith one or more of the sides containing a viewing screenilluminated from the inside of the housing. The viewing screenmay also be the diffusing screen. There shall be thermalprotection to prevent overheating, and subs
9、equent damage tothe radiographs placed on the viewing screen. The housing orsystem may or may not require ventilation. A rheostat orsuitable electrical circuit shall be provided to vary the lightintensity.6. Physical Properties6.1 GeneralThe illuminator shall be manufactured ofmaterials deemed suita
10、ble to withstand the environmentalconditions encountered under normal operating conditions.6.2 Viewing ScreenThe viewing screen shall be easy toclean and made of material which is resistant to scratches. Thesize of the screen shall allow the user to view the radiographwithout excessive glare. If the
11、 illuminator is to be used forviewing radiographs of various sizes, masks of various sizesand configurations should be provided. Alternately, an adjust-able aperture may be used.1This specification is under the jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct responsibil
12、ity of Subcommittee E07.01 onRadiology (X and Gamma) Method.Current edition approved June 15, 2012. Published July 2012. Originallyapproved in 1990. Last previous edition approved in 2006 as E1390 - 06. DOI:10.1520/E1390-12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcont
13、act ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West
14、 Conshohocken, PA 19428-2959, United States.6.3 Color of LightThe color of the light used to illuminatethe radiograph shall be white, that is, color temperaturebetween 5000 and 6250K. However, illuminators using non-white or “colored” light may be used if they have beenrecommended by the film manufa
15、cturers.6.4 Diffusing ScreenIf the illuminator has a diffusingscreen, the light shall be sufficiently divergent in accordancewith 7.2.6.5 HousingThe external housing shall be constructed insuch a manner that no disturbing light hinders the viewing ofthe radiographs.6.6 Anti-Glare DeviceThe illuminat
16、or shall be fitted withan anti-glare switch or device that minimizes the probability ofthe operator being subjected to excessive glare when theradiograph is removed. This switch or device may be manual orautomatic.6.7 Illuminators Used for Viewing “Wet” RadiographsIlluminators manufactured for use i
17、n viewing “wet” radio-graphs shall be manufactured to prevent the penetration ofliquid into internal electrical components in such a manner thatsafe operation of the unit would be compromised.7. Performance Requirements7.1 Maximum Luminance outputThe luminance of thetransmitted light shall not be le
18、ss than 30 candelas per squaremetre for film densities equal to or less than 2.5 optical densityand not less than 10 candelas per square metre for filmdensities greater than 2.5 optical density.7.2 Divergence and Diffusion of LightIf the illuminatorhas a diffusing screen, the light shall be sufficie
19、ntly divergentso that both eyes of the observer receive rays from all parts ofthe screen. The divergence factor shall exceed 0.7.7.3 The viewing screen shall be uniformly illuminated withthe uniformity factor d being higher than 0.5.7.4 Appropriate precautions shall be taken by the manufac-turer to
20、ensure that temperature of the housing does not exceed60C (140F) at the usual contact surfaces after one hour ofoperation at a 50 % duty cycle.7.5 A radiograph having a density of 2.0 optical densitywhen placed onto the illuminator viewing surface shall notwarp or curl after one minute of continuous
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME13902012STANDARDSPECIFICATIONFORILLUMINATORSUSEDFORVIEWINGINDUSTRIALRADIOGRAPHS 观察 工业 照片 照明 装置 标准规范

链接地址:http://www.mydoc123.com/p-528599.html