ASTM E275-2001 Standard Practice for Describing and Measuring Performance of Ultraviolet Visible and Near-Infrared Spectrophotometers《说明和测量紫外线 可见和近红外线分光光度计的性能的标准操作规程》.pdf
《ASTM E275-2001 Standard Practice for Describing and Measuring Performance of Ultraviolet Visible and Near-Infrared Spectrophotometers《说明和测量紫外线 可见和近红外线分光光度计的性能的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E275-2001 Standard Practice for Describing and Measuring Performance of Ultraviolet Visible and Near-Infrared Spectrophotometers《说明和测量紫外线 可见和近红外线分光光度计的性能的标准操作规程》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 275 01Standard Practice forDescribing and Measuring Performance of Ultraviolet,Visible, and Near-Infrared Spectrophotometers1This standard is issued under the fixed designation E 275; the number immediately following the designation indicates the year oforiginal adoption or, in the ca
2、se of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.INTRODUCTIONIn devel
3、oping a spectrophotometric method it is the responsibility of the originator to describe theinstrumentation and the performance required to duplicate the precision and accuracy of the method.It is necessary to specify this performance in terms that may be used by others in applications of themethod.
4、The tests and measurements described in this practice are for the purpose of determining theexperimental conditions required for a particular analytical method. In using this practice an analysthas either a particular analysis for which he describes requirements for instrument performance, or heexpe
5、cts to test the capability of an instrument to perform a particular analysis. To accomplish eitherof these objectives it is necessary that instrument performance be obtained in terms of the factors thatcontrol the analysis. Unfortunately, it is true that not all the factors that can affect the resul
6、ts of ananalysis are readily measured and easily specified for the various types of spectrophotometricequipment.Of the many factors that control analytical results, this practice covers selection of the setting ofanalytical wavelength, selection of slit width, photometric measurements, and character
7、istics ofabsorption cells as the parameters of spectrophotometry that are likely to be affected by the analyst inobtaining data. Other important factors, particularly those primarily dependent on instrument design,are not covered in this practice.1. Scope1.1 This practice covers the description of r
8、equirements ofspectrophotometric performance especially for ASTM meth-ods, and the testing of the adequacy of available equipment fora specific method. The tests give a measurement of some of theimportant parameters controlling results obtained in spectro-photometric methods, but it is specifically
9、not to be concludedthat all the factors in instrument performance are measured.1.1.1 This practice is not to be used ( 1) as a rigorous test ofperformance of instrumentation, or (2) to intercompare thequantitative performance of instruments of different design.1.1.2 This practice is primarily direct
10、ed to dispersive spec-trophotometers used for transmittance measurements ratherthan instruments designed for diffuse transmission and diffusereflection.2. Referenced Documents2.1 ASTM Standards:E 131 Terminology Relating to Molecular Spectroscopy2E 168 Practices for General Techniques of Infrared Qu
11、anti-tative Analysis2E 169 Practices for General Techniques of Ultraviolet-Visible Quantitative Analysis2E 387 Test Method for Estimating Stray Radiant PowerRatio of Spectrophotometers by the Opaque Filter Method2E 958 Practice for Measuring Practical Spectral Bandwidthof UltravioletVisable Spectrop
12、hotometers23. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this practice, refer toTerminology E 131.4. Significance and Use4.1 This practice permits an analyst to compare the generalperformance of his instrument, as he is using it in a specificspectrophotometric method, with the
13、 performance of instru-ments used in developing the method.1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and is the direct responsibility of Subcommittee E13.01 on Ultravioletand Visible Spectroscopy.Current edition approved Feb. 10, 2001. Published April 20
14、01. Originallypublished as E 275 65 T. Last previous edition E 275 93.2Annual Book of ASTM Standards, Vol 03.06.1Copyright ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.5. Reference to This Practice in Standards5.1 Reference to this practice in any ASTM spectrophoto-me
15、tric method (preferably in the section on apparatus wherethe spectrophotometer is described) shall constitute due noti-fication that the adequacy of the spectrophotometer perfor-mance is to be evaluated by means of this practice. Perfor-mance is considered to be adequate when the instrument can beop
16、erated in a manner to give test results equivalent to thoseobtained on instruments used in establishing the method or incooperative testing of the method.5.2 It is recommended that the apparatus be described interms of the results obtained on application of this practice toinstruments used in establ
17、ishing the method. This descriptionshould give a numerical value showing the wavelength accu-racy, wavelength repeatability, and photometric repeatabilityfound to give acceptable results. A recommended spectral slitwidth maximum should be given along with typical spectra ofthe components to be deter
18、mined to indicate the resolutionfound to be adequate to perform the analysis. If it is considerednecessary in a particular analysis, the use of only the linearportion of an analytical curve (absorbance per centimetreversus concentration) may be specified, or if nonlinearity isencountered, the use of
19、 special calculation methods may bespecified. However, it is not permissible to specify the amountof curvature if a nonlinear working curve is used.6. Parameters in Spectrophotometry6.1 Any spectrophotometer may be described as a source ofradiant energy, a dispersing optical element, and a detectort
20、ogether with a photometer for measuring relative radiantpower. Accurate spectrophotometry involves a large number ofinterrelated factors that determine the quality of the radiantenergy passing through a sample and the sensitivity andlinearity with which this radiant energy may be measured.Assuming p
21、roper instrumentation and its use, the instrumentalfactors responsible for inaccuracies in spectrophotometry in-clude resolution, linearity, stray radiant energy, and cell con-stants. Rigorous measurement of these factors is beyond thescope of this practice. The measurement of stray radiant energyis
22、 described in Method E 387.6.2 Modern spectrophotometers are capable of more accu-racy than most analysts obtain. The problem lies in theselection and proper use of instrumentation. In order to ensureproper instrumentation and its use in a specific spectrophoto-metric method, it is necessary for an
23、analyst to evaluate certainparameters that can control the results obtained. These param-eters are wavelength accuracy and precision, spectral slitwidth, photometry, and absorption-cell constants. Unsatisfac-tory measurement of any of these parameters may be due toimproper instrumentation or to impr
24、oper use of availableinstrumentation. It is therefore first necessary to determine thatinstrument operation is in accordance with the manufacturersrecommendations. Tests shall then be made to determine theperformance of an instrument in terms of each of the param-eters in 6.1 and 6.2.7. Instrument O
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME2752001STANDARDPRACTICEFORDESCRIBINGANDMEASURINGPERFORMANCEOFULTRAVIOLETVISIBLEANDNEARINFRAREDSPECTROPHOTOMETERS

链接地址:http://www.mydoc123.com/p-527279.html