IEC TR 61831-2011 On-line analyser systems - Guide to design and installation《在线分析仪系统设计和安装指导》.pdf
《IEC TR 61831-2011 On-line analyser systems - Guide to design and installation《在线分析仪系统设计和安装指导》.pdf》由会员分享,可在线阅读,更多相关《IEC TR 61831-2011 On-line analyser systems - Guide to design and installation《在线分析仪系统设计和安装指导》.pdf(76页珍藏版)》请在麦多课文档分享上搜索。
1、 IEC/TR 61831 Edition 2.0 2011-08 TECHNICAL REPORT On-line analyser systems Guide to design and installation IEC/TR 61831:2011(E) THIS PUBLICATION IS COPYRIGHT PROTECTED Copyright 2011 IEC, Geneva, Switzerland All rights reserved. Unless otherwise specified, no part of this publication may be reprod
2、uced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or IECs member National Committee in the country of the requester. If you have any questions about IEC copyright or have an enquiry about obtain
3、ing additional rights to this publication, please contact the address below or your local IEC member National Committee for further information. Droits de reproduction rservs. Sauf indication contraire, aucune partie de cette publication ne peut tre reproduite ni utilise sous quelque forme que ce so
4、it et par aucun procd, lectronique ou mcanique, y compris la photocopie et les microfilms, sans laccord crit de la CEI ou du Comit national de la CEI du pays du demandeur. Si vous avez des questions sur le copyright de la CEI ou si vous dsirez obtenir des droits supplmentaires sur cette publication,
5、 utilisez les coordonnes ci-aprs ou contactez le Comit national de la CEI de votre pays de rsidence. IEC Central Office 3, rue de Varemb CH-1211 Geneva 20 Switzerland Email: inmailiec.ch Web: www.iec.ch About IEC publications The technical content of IEC publications is kept under constant review by
6、 the IEC. Please make sure that you have the latest edition, a corrigenda or an amendment might have been published. Catalogue of IEC publications: www.iec.ch/searchpub The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,). It also gi
7、ves information on projects, withdrawn and replaced publications. IEC Just Published: www.iec.ch/online_news/justpub Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available on-line and also by email. Electropedia: www.electropedia.or
8、g The worlds leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical Vocabulary online. Customer Service Centre: www.iec.ch/we
9、bstore/custserv If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service Centre FAQ or contact us: Email: csciec.ch Tel.: +41 22 919 02 11 Fax: +41 22 919 03 00 IEC/TR 61831 Edition 2.0 2011-08 TECHNICAL REPORT On-line analyser systems Gu
10、ide to design and installation INTERNATIONAL ELECTROTECHNICAL COMMISSION XB ICS 13.040; 13.060; 17.020 PRICE CODE ISBN 978-2-88912-633-0 Registered trademark of the International Electrotechnical Commission 2 TR 61831 IEC:2011(E) CONTENTS FOREWORD . 5 INTRODUCTION . 7 1 Scope . 8 2 Normative referen
11、ces . 8 3 Terms and definitions . 8 4 Remarks and considerations. 9 4.1 General remarks 9 4.2 Further considerations . 9 4.3 Reliability 10 4.4 Design . 10 4.5 Centralisation 10 4.6 Local mounting 10 4.7 Pre-assembled systems 10 5 Health, safety and environmental considerations 11 5.1 Overview . 11
12、5.2 Prevention of explosions and fires . 11 5.3 Prevention of toxic and asphyxiant hazards . 11 5.4 Radiation hazards . 12 5.5 Safety facilities 12 5.6 Manual shut-down facilities . 12 5.7 Noise 12 6 Housings 12 6.1 Overview . 12 6.2 Selection of housing 13 6.2.1 Analyser case 13 6.2.2 Analyser cabi
13、net 13 6.2.3 Analyser shelter 13 6.2.4 Analyser house 14 6.3 Area classification and toxic danger 14 6.3.1 Electrical area classification 14 6.3.2 Toxic and asphyxiate danger . 14 6.4 Construction and mounting 14 6.4.1 General . 14 6.4.2 Analyser housings . 15 6.4.3 Analyser houses 15 6.4.4 Analyser
14、 shelters . 15 6.5 Analyser housings with natural ventilation . 16 6.5.1 General . 16 6.5.2 Ventilation requirements 16 6.5.3 Heating requirements 16 6.5.4 Analyser shelters . 17 6.5.5 Analyser cases 17 6.6 Analyser housings with forced ventilation 17 6.6.1 General . 17 6.6.2 Ventilation requirement
15、s 17 6.6.3 Air intake system . 17 TR 61831 IEC:2011(E) 3 6.6.4 Ventilation fan requirements 17 6.6.5 Airflow requirements 18 6.6.6 Heating requirements 18 6.6.7 Safety monitors and alarms . 18 7 Sampling systems 21 7.1 Overview . 21 7.2 Sample system terminology . 22 7.3 General requirements 22 7.4
16、Sample point location 23 7.5 Fast circulating systems (fast loops) 24 7.6 By-pass systems . 24 7.7 Sample recovery systems 24 7.8 Special considerations 25 7.9 Multi-stream systems 25 7.10 Construction 26 7.10.1 General . 26 7.10.2 Material selection 26 7.10.3 Flushing facilities . 27 7.10.4 Blockag
17、e removal 27 7.10.5 Heat tracing and insulation 27 7.10.6 Minimising risks from leakage 28 7.10.7 Location of equipment . 28 7.10.8 Instrumentation 28 7.10.9 Identification 28 7.11 Effluent disposal 29 7.11.1 General . 29 7.11.2 Vapour 29 7.11.3 Liquid 29 7.12 Calibration facilities . 30 7.13 Automa
18、tic calibration . 30 8 Analyser communications . 31 8.1 Overview . 31 8.2 Signal transmission . 31 8.3 Safety . 32 8.4 Cables . 33 8.5 Use of signal . 33 8.6 Alarms . 33 Annex A (informative) Typical analyser process line sampling probe for line sizes NPS 2“ and above 34 Annex B (informative) Determ
19、ination of sample probe lengths . 35 Annex C (informative) Sample system calculations 43 Annex D (informative) Natural ventilation calculations 56 Annex E (informative) Forced ventilation calculations 62 Annex F (informative) Example of verification/calibration sequence of data to computer 63 Annex
20、G (informative) Analyser house with forced ventilation Summary of recommended control shut-down actions for flammable hazard operation . 64 4 TR 61831 IEC:2011(E) Annex H (informative) Analyser houses with forced ventilation Ventilation failure and flammable gas detection trip logic . 65 Annex I (in
21、formative) Typical analyser system schematics . 67 Annex J (informative) Example schematic showing a typical wall penetration using a transit . 69 Bibliography 70 Figure A.1 Typical analyser sample probe design 34 Figure C.1 Sample system configuration 49 Figure D.1 Schematic showing wind induced fr
22、ee ventilation principles with worked example ventilation louvre layout and suggested warm air distribution 56 Figure F.1 Typical signal sequence during verification . 63 Figure H.1 Typical analyser shelter logic diagram 66 Figure I.1 Multi-stream sampling for processes of similar pressure and compo
23、nents (e.g. atmospheric sampling) 67 Figure I.2 Multi-stream sampling for processes of differing pressure and/or components showing typical double block and bleed arrangement 67 Figure I.3 Minimum pollution sampling configuration 68 Figure I.4 Typical double block and bleed arrangement 68 Figure J.1
24、 Typical wall penetration using a transit 69 Table B.1 Vibrational Mode Constants . 36 Table B.2 Example calculations for maximum fluid velocity 40 Table C.1 Nomenclature Used In Calculations . 43 Table C.2 Equivalent lengths of valves and fittings 46 Table C.3 Sample system component pressure drop
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- IECTR618312011ONLINEANALYSERSYSTEMSGUIDETODESIGNANDINSTALLATION 在线 分析 系统 设计 安装 指导 PDF

链接地址:http://www.mydoc123.com/p-1238002.html