UOP 411-2013 Normal Paraffins by Subtractive GC.pdf
《UOP 411-2013 Normal Paraffins by Subtractive GC.pdf》由会员分享,可在线阅读,更多相关《UOP 411-2013 Normal Paraffins by Subtractive GC.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1966, 1975, 1992, 2013 UOP LLC. All rights reserved. Nonco
3、nfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 P
4、HONE. Normal Paraffins by Subtractive GC UOP Method 411-13 Scope This method is for determining total normal paraffins (TNP), individual normal paraffins, and total non-normal paraffins in desorbent-free UOP Molex process feed, raffinate, and extract streams. The lower limit of quantitation for an i
5、ndividual normal paraffin is 0.1 mass-%. References ASTM Method D4052, “Density, Relative Density, and API Gravity of Liquids by Digital Density Meter,” www.astm.org ASTM Practice D4307, “Preparation of Liquid Blends for Use as Analytical Standards,” www.astm.org Scanlon, J. T. and Willis, D. E., Jo
6、urnal of Chromatographic Science, 23, 333-340 (1985) UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method Feed and raffinate samples to be analyzed are injected twice into a gas chromatograph that is equipped with a fused silica capillary column internally coated wit
7、h cross-linked methyl silicone, an autosampler, and a flame ionization detector (FID). The first injection is performed with a standard capillary injection port liner; the second injection is performed with a capillary injection port liner packed with molecular sieves that quantitatively remove the
8、normal paraffins from the sample. The total sample chromatogram from injection one is superimposed and aligned with the de-normalized chromatogram from injection two so that the net individual normal paraffin peaks can be measured. The mass-% of each normal paraffin is obtained by the external stand
9、ard calibration technique. An extract sample to be analyzed is injected with the standard capillary injection port liner and the mass-% of each normal paraffin is obtained using the internal normalization calibration technique. Apparatus References to catalog numbers and suppliers are included as a
10、convenience to the method user. Other suppliers may be used. Balance, readable to 0.0001 g Chromatographic column, 30 m of 0.25-mm ID fused silica capillary, internally coated to a film thickness of 0.25 micron cross-linked methyl silicone, RTX-1, Restek, Cat. No. 10123 2 of 11 411-13 Data system, C
11、hemStation, Rev. B.01 or later, Agilent Technologies, for obtaining peak areas. The ChemStation software must have programmable parameters for controlling baseline events, and have signal alignment and subtraction capabilities, no substitution except for the pre-configured systems listed under Gas c
12、hromatograph. Desiccator, VWR, Cat. No. 25035-005, with plate, Cat. No. 89038-068 Gas chromatograph, temperature programmable, built for capillary column chromatography, utilizing a split injection system, equipped with a turn top inlet system, glass injection port inserts, and a flame ionization de
13、tector that will give a minimum peak height response of five times the background noise for 0.01 mass-% n-heptane when operated at the recommended conditions, Agilent Technologies, Model 7890. An autosampler is required. A pre-configured system, complete with gas chromatograph, data system, software
14、, columns, and prepacked injection port liners is available from Separation Systems, with the following catalog numbers: UOP-411-01, UOP 411 GC configuration, single detector/single sided UOP-411-02, UOP 411 GC configuration, double detector/double sided UOP-411-S, SOFTWARE, UOP 411 MOLEX Process so
15、ftware, for distributive and non-distributive mode analysis ss-035-500, LINERS, UOP 411 MOLEX Process liners w/molecular sieves Gas purifier, to remove oxygen from the hydrogen carrier gas, VICI Mat/Sen, Cat. No. P-200-1 Injection port liner, glass, split, focus, SGE Inc., Cat. No. 092002, several r
16、equired, no substitution Leak detector, gas, Grace Davison, Cat. No. 21-250 Puller/inserter tool, Sigma-Aldrich Co., Cat. No. 22406 Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, hydrogen, two-stage, high purity, de
17、livery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen or helium, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Sample injector, autosampler, capable of introducing a repeatable 0.5-L volume of sam
18、ple, Agilent Technologies, Model 7693A Tweezers, curved tip, VWR, Cat. No. 63042-964 Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane, local
19、 supply Autosampler vials, glass, 12 x 32 mm, for Agilent autosampler, Fisher Scientific, Cat. No. 03-391-8 Cyclohexane, 99.9% purity, Sigma-Aldrich, Cat. No. 227048-1L Caps, screw, for autosampler vials, open hole with septum, 9 mm, with Tef/Sil/Tef liner, Fisher Scientific, Cat. No. 03-377-2A Desi
20、ccant, indicating, for desiccator, VWR, Cat. No. 22891-028 3 of 11 411-13 Glass wool, Fisher Scientific, Cat. No. 11-390 Hydrogen, zero gas, 99.95% minimum purity, total hydrocarbons less than 0.5 ppm as methane, local supply Molecular sieve, UNISIV-2, 60 to 80 mesh, available prepacked in injection
21、 port liners from Separation Systems, Cat. No. ss-035-500 Pipet bulbs, Fisher Scientific, Cat. No. 13-678-9B Pipets, disposable, Pasteur, Fisher Scientific, Cat. No. 13-678-20A Nitrogen or helium, zero gas, total hydrocarbons less than 0.5 ppm as methane, local supply Normal paraffins, individual ca
22、rbon numbers, typically C8 to C17, 99 mass-% minimum purity, ChemSampCo. Obtain all carbon numbers represented in the feed. Table 1 Normal Paraffins n-Paraffin CAS Number Cat. No. C8 Octane 111-65-9 7056.00 C9 Nonane 111-84-2 6780.00 C10 Decane 124-18-5 0812.55 C11 Undecane 1120-21-4 9479.00 C12 Dod
23、ecane 112-40-3 2269.50 C13 Tridecane 629-50-5 8745.00 C14 Tetradecane 629-59-4 8438.00 C15 Pentadecane 629-62-9 7290.00 C16 Hexadecane 544-76-3 3278.00 C17 Heptadecane 629-78-7 3030.00 Syringe, for sample injector, 10-L, Agilent Technologies, Cat. No. 5181-1273 Vials, 15-mL, with screw cap, Fisher S
24、cientific, Cat. No. 03-339-21J Procedure The analyst is expected to be familiar with general laboratory practices, the technique of gas chromatography, and the equipment being used. Dispose of used reagents, materials, and samples in an environmentally safe manner according to local regulations. Chr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP4112013NORMALPARAFFINSBYSUBTRACTIVEGCPDF

链接地址:http://www.mydoc123.com/p-1074497.html