《EN 61262-3-1994 en Medical Electrical Equipment - Characteristics of Electro-Optical X-Ray Image Intensifiers Part 3 Determination of the Luminance Distribution and Luminance Non-U.pdf》由会员分享,可在线阅读,更多相关《EN 61262-3-1994 en Medical Electrical Equipment - Characteristics of Electro-Optical X-Ray Image Intensifiers Part 3 Determination of the Luminance Distribution and Luminance Non-U.pdf(20页珍藏版)》请在麦多课文档分享上搜索。
1、CENELEC ENxbL2b2-3 94 3404583 OL49LY3 629 BRITISH STANDARD Characteristics of electro-optical X-ray image intensifiers for medical electrical equipment Part 3. *Determination of the luminance distribution and luminance non-uniformity The European Standard EN 61262-3 : 1994 has the status of a Britis
2、h Standard BS EN 61262-3 : 1995 IEC 1262-3 : 1994 CENELEC EN*bl12b2-3 94 m 3404583 0149144 565 m BS EN 61262-3 : 1996 This British Standard, having been prepared under the direction of the Health and Environment Sedor Board (W-), was published under the authority of the Standards Board and comes int
3、o effect on 15 March 1995 O BSI 1995 The following BSI references relate to the work on this standard: Committee reference HCC/72 Draft for comment 94/501343 DC ISBN O 680 23694 7 Committees responsible for this British Standard The preparation of this British Standard was entrusted to %chnical Comm
4、ittee HCC/72, Image intensifiers, upon which the foliowing bodies were represented : Association of X-ray Equipment Manufacturers (BEAMA Ltd.) British Institute of Radiology College of Radiographers Institute of Physical Sciences in Medicine (Ipsm) Amendments issued since publication Amd. No. I Date
5、 I Rxt affected CENELEC EN*bLZb2-3 94 = 3404583 0147145 4TL BS EN 61262-3 : 1995 Contents hge Committees responsible Inside front cover National foreword ii Foreword 2 Rxt of EN 61262-3 4 i CENELEC EN*b12b2-3 94 m 340Y583 0149146 338 m BS EN 61262-3 : 1995 National foreword This British Standard has
6、 been prepared by Bchnical Committee HCC/72 and is the English language version of EN 61262-3 : 1994 Medical electrical equi- - Chamctertics of electro-optical X-ray image in implies that compliance with a requirement is strongly recommended but is not mandatory for compliance with the standard; imp
7、lies that compliance with a requirement is permitted to be accomplished in a particular manner, for compliance with the standard; and the following words have the meaning: - “specific“ - “specified“ when used in combination with parameters or conditions: refers to a particular value or standardized
8、arrangement, usually to those required in an IEC standard or a legal requirement; see IEC 788, rm-74-01. when used in combination with parameters or conditions: refers to a value or arrangement to be chosen for the purpose under consideration and indicated usually in the ACCOMPANYING DOCUMENTS; see
9、JEC 788, rm-74-02. - “designed for“ when used in standards to characterize equipment, devices, components or arrangements:. designates an intended and usually apparent purpose or use for the product. CENELEC ENubLZb2-3 94 3404583 OL49L53 578 Page 7 EN 61262-3 : 1994 4 Requirements 4.1 Test ser-up a)
10、 The SED shall be 100 cm f 1 cm. b) The FOCAL SPOT of the X-RAY TUBE shall be on the CENTRAL AXIS. c) The distance between the FOCAL SPOT of the X-RAY TUB and any ADDED FILTERS shall not exceed 33 cm. d) The X-RAY FIELD shall cover the ENTRANCE FIELD entirely, but its area shall not exceed the ENTRA
11、NCE FIELD area by more than 10 %. This limitation of the X-RAY FIELD shall be made by a DIAPHRAGM placed immediately in front of the ENTRANCE PLANE. The BEAM LIMITING DEVICE Of the X-RAY SOURCE ASSEMBLY shall be Used to limit the X-RAY BEAM on this DIAPHRAGM. The DIAPHRAGM shall provide an ATENUATIO
12、N FIATIO Of at least 100 for !he quantity AIR KERMA RATE. 4.2 X-RAY IMAGE INTENSIFIER - Operating conditions a) The XRII shall be operated under the conditions for NORMAL USE as specified by the manufacturer. b) NO ANTI-SCAITER GRID or protective cover shall be Used. c) In the case of multiple-field
13、 XRIIs, the measurement shall be made for the largest specified ENTRANCE FIELD. Measurements for other ENTRANCE FIELDS are optional. 4.3 input radiation a) The TOTAL FiLTRAnoN of the incident X-RAY BEAM shall be 22,5 mm f 0,5 mm of aluminium equivalent of which at least 20 mm is aluminium (99,9 % pu
14、rity). The HALF-VALUE LAYER at the position corresponding to the CENTRE OF THE ENTRANCE FIELD shall be 7,O mm Al f 0,2 mm Al (99,9 % purity). This condition corresponds to an approximate X-RAY TUBE VOLTAGE of 75 kV. b) The AIR KERMA RATE in the ENTRANCE PLANE shall be such that linear operation of t
15、he XRII and the detectorsemployed in the measurement is ensured. An AIR KERMA RATE of 1 pGy s- to 100 pGy s- is recommended. c) The temporal fluctuations of the AIR KERMA RATE shall not impair the measurement uncertainty by more than 2 %. If the AIR KERMA RATE measurements or the luminance measureme
16、nts are made at different time instances, the X-RAY TUBE output shall be monitored and the results of the measurements shall be corrected to meet this requirement. d) The AIR KERMA RATE at any location in the ENTRANCE FIELD shall not be less than 70 % Of the highest value of the AIR KERMA RATE in th
17、e ENfRANCE FIELD. 4.4 TEST DEVICE Not used. Page 8 EN 61262-3 : 1994 4.5 Measurement equipment 4.5.1 Device for measuring the uniformity of the X-RAY FIELD a) Any device measuring a quantity which has a known relationship to the relative AIR KERMA RATE may be used, for example, a photomultiplier tub
18、e coupled to a scintillator or a NON-SCREEN FILM with a known sensitometric curve. The device shall have such a precision, that the error of measuring relative AIR KERMA RATE is 2 % or less. No particular specifications are required with respect to the absolute accuracy of these measurements. b) The
19、 maximum diagonal of the effective measuring area of the device shall not exceed 10 % Of the ENTRANCE FIELD SIZE. 4.5.2 Device for measuring the luminance of the OUTPUT SCREEN a) Any device measuring a quantity which has a known relationship to the relative luminance may be used. The device shall ha
20、ve such a precision, that the error of measuring relative luminance is 2 % or less. No particular specifications are required with respect to the absolute accuracy of these measurements. b) The device shall measure the relative luminance in the direction perpendicular to the plane defined by the OUT
21、PUT SCREEN. The acceptance cone of the detector shall be small enough to avoid errors when measurement approaches a dark edge. c) The diagonal of the effective area of the measuring field shall be at least 0,l mm and shall not exceed 5 Yo of the OUTPUT IMAGE diameter. d) If the luminance is measured
22、 through a lens with a one- or two-dimensional image detector in fixed position, the lens vignetting, the detector non-uniformity and the non-linearity of its response shall be corrected for. 5 Determination Of the LUMINANCE DISTRIBUTION and LUMINANCE NON-UNIFORMIM 5.1 Preparation a) The requirement
23、s for the measurement Set-up given in 4.1 shall be fulfilled. b) The TOTAL FILTRATION Of the RADIATION BEAM and the X-RAY TUBE VOLTAGE are adjusted according to 4.3 a) to obtain the required HALF-VALUE LAYER at the position corresponding to the CENTREOFTHE ENTRANCE FIELD according to 4.1. 5.2 Measur
24、ement 5.2.1 Measurement of the X-RAY FIELD uniformity a) After the determination Of the ENTRANCE PLANEand the CENTREOFTHE ENTRANCE FIELD, the XRII is removed from the X-RAY BEAM. No objects capable of scattering X-RADIATION shall remain in the X-RAY BEAM at a distance less than 50 cm from the ENTRAN
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- EN6126231994ENMEDICALELECTRICALEQUIPMENTCHARACTERISTICSOFELECTROOPTICALXRAYIMAGEINTENSIFIERSPART3DETERMINATIONOFTHELUMINANCEDISTRIBUTIONANDLUMINANCENONUPDF

链接地址:http://www.mydoc123.com/p-719893.html