DOD A-A-54759-1992 CANISTER AND TIMER SYSTEM (Oropharyngeal Suction Apparatus)《滤毒罐和定时系统》.pdf
《DOD A-A-54759-1992 CANISTER AND TIMER SYSTEM (Oropharyngeal Suction Apparatus)《滤毒罐和定时系统》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54759-1992 CANISTER AND TIMER SYSTEM (Oropharyngeal Suction Apparatus)《滤毒罐和定时系统》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、-_I_ - - A-A-54759 m 9999974 0103930 514 I INCH-POUND 1 A-A-5 4 7 5 9 15 October 1992 COMMERCIAL ITEM DESCRIPTION CANISTER AND TIMER SYSTEM (Oropharyngeal Suction Apparatus) The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Descripti
2、on covers a system of 3 canisters connected to a manifold and an electrical timer to convert a 115 volt suction apparatus from constant flow to intermittent flow. Salient characteristics: The unit shall be a three-canister and timer system ,for Wangensteen type intermittent suctioning of 3 patients
3、simultaneously providing an on-off timer for intermittent interruption of power supply to a suction apparatus. shall be suitable for use in hospital and military aeromedical aircraft. it The system shall include the following: least 6 feet long to connect the systems manifold to a suction apparatus.
4、 a. A vacuum supply hose (tubing), 3/8 inch ID by at b. with a range from O to 760 mm Hg vacuum, connected to the manif old. A vacuum gauge, with connector for the supply hose, c. A manifold to receive vacuum from a suction apparatus and feed adjustable vacuum to 3 suction canisters. d. Three vacuum
5、 adjustment valves, mounted on the manifold, each with a fitting for connection of the canisters vacuum lines. AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from
6、 IHS-,-,-_ A-A-54759 9999974 OL0393L 450 A-A-54 7 59 e. Three vacuum lines, 1/4 inch ID by at least 6 feet long, for connection between the manifolds adjustable vacuum valves and 3 canisters. f. Three disposable suction canisters, each with working capacity of at least 1200 cc having graduation mark
7、ings conforming with ANSI 279.9 and having a barb or tapered connector for 1/4 patient tubing. g. An electrical control box to provide ON-OFF cycling of 115 volt, 50 to 400 Hz power source to a suction apparatus. The control box, with its power cord assembly, shall be designed to control Electromagn
8、etic Interference (EMI) and Radio Frequency Interference (RFI). The control box shall include: 1. A flush mounted 115 volt receptacle to receive a power cord with a hospital grade plug to provide cycled on-off power to a surgical suction apparatus (marked to identify as timed/cycled on-off power). 2
9、. A flush mounted 115 volt hospital grade receptacle for general purpose use (not time/cycled controlled). 3. A 115 volt cord with hospital grade plug, hard 4. A timer to intermittently interrupt power from wired, to allow input power from a wall receptacle. 115 VAC wall source providing power cycli
10、ng of 15 seconds off and 45 seconds on. The switch circuit shall be rated at least 5 amps. h. Three mounting bracket sets to mount each of the canisters on a stanchion with cross-section of either 2 inches by 2 inches or 2-3/4 inches by 2 inches. The system shall be compatible with the oropharyngeal
11、 suction apparatus described in Commercial Item Description A-A-54039. Workmanship. defects which detract from its appearance or impair its serviceability. The canister and timer system shall be free from Unit. Each (EA). One canister and timer system, as specified, constitutes one unit. 2 Provided
12、by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-_-I_- - A-A-54759 = 9999974 0303932 397 m A-A-54 7 5 9 Operational and Service Data Manuals shall be furnished as follows : a. information content: The contractor shall furnish manuals, handbooks, and/or brochur
13、es containing complete operation, installation and service/maintenance instructions (including pictures or illustrations, as necessary) with complete schematics and wiring diagrams (if unit has electric circuitry). All manuals, handbooks, and/or brochures shall be written in the English language and
14、 all schematics and wiring diagrams shall use American electrical data and connection diagrams for all I applicable utilities. The instructions shall also contain a complete list of all replaceable parts showing part number, name and quantity required. All service notes, service memo, etc., pertaini
15、ng to the equipment and issued subsequent to the printing of the original instructions shall also be included. These instructions shall be, as a minimum, that furnished to service engineers (or servicemen) that normally install and service the equipment for the company or distributor. procured inclu
16、des, or will operate ancillary equipment, the service data shall include complete instructions and drawings which show interfacing of all system components. When the system being b. Quantities: Two (2) complete sets of the above information shall be packaged with each unit of the required equipment.
17、 Quality Assurance Provisions. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA547591992CANISTERANDTIMERSYSTEMOROPHARYNGEALSUCTIONAPPARATUS 滤毒罐 定时 系统 PDF

链接地址:http://www.mydoc123.com/p-701771.html