1、Ip A-A-53777 65 U 9999774 0033873 T = A-A- 5 3 7 7 7 10 December 1988 COMMERCIAL ITEM DESCRIPTION AUDIOMETER, Manually Operated, Portable, 110/220 Volt, 50/60 Hz, AC The General Services Administration has authorized the use of this commercial item description in preference to Military Specification
2、 MIL-A-36494, which is cancelled. This commercial item description covers a single channel, fixed frequency oscillator, portable, manually operated, audiometer capable of performing basic threshold evaluations with calibrated masking noise and bone conduction. Salient characteristics. General. The a
3、udiometer shall be capable of performing the following pure tone hearing tests: Pure tone air conduction. Pure tone air conduction with masking. Pure tone bone conduction. Pure tone bone conduction with masking. The audiometer shall comply with American National Standards Institute (ANSI) Standard S
4、3.6. In the event of confliction, the requirements specified herein shall take precedence. Exceptions. not apply: The following portions of ANSI Standard S3.6 do 4.2 Test Tones for Limited Range amd Narrow Range Audiometers 4.3 Variation in Method of Presentation of Test Tones 5. Requirements for Sp
5、eech Audiometers AMSC/ NA FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. THIS DOCUMENT CONTAINS 8 PAGES.- _- - Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-Test Tones. least the following fixed frequenci
6、es: The audiometer shall generate test tones of at Air Conduction (Hz) 250 500 1000 1500 2000 3000 4000 5000 6000 8000 Bone Conduction (Hz) 250 500 1000 1500 2000 3000 4000 Accessories. accessories: The unit shall be provided with the following 1 Pr 1 Set 1 Set 1 Ea 6 Ea 1 Pr Description Air Receive
7、rs Head Band and Cord Patch Cord for (See Note 2) Audiometer Booth Bone Oscillator with Headband and Cord Spare Fuses (if applicable) Additional Air Receiver Cord Instructions. Instruction and Service manuals. repair parts and wiring diagrams shall be included. Electrical. from the following line so
8、urces: 110 volts (plus or minus 10 percent) and 220 volt (plus or minus 10 percent), multiple frequency, 50 and 60 Hz (plus or minus 2 Hz). The unit shall be supplied with a UL-approved three-wire ground line cord and a 110 volt green dot hospital grade male plug. Carrying Case. rugged, shock resist
9、ing carrying case with handle. The case shall offer a protective cover for the control panel and be capable of housing all accessory components including Instruction data. The unit shall be provided with 2 sets of A list of high mortality The audiometer shall be capable of operating The audiometer s
10、hall be contained in a 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-. O a a A-A-53777 65 977777q 0031875 3-! A-A-5 3 7 7 7 Calibration. Using the methods described in ANSI S3.6, the audiometer shall be calibrated to conform with specified requir
11、ements. A certificate of calibration shall be provided with the unit. Workmanship. which detract from its appearance or impair its serviceability. The audiometer shall be free from defects Unit. Each (EA). One audiometer, as specified, constitutes one unit. Quality Assurance Provisions. the contract
12、 or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements spec
13、ified herein, unless disapproved by the Government. Government reserves the right to perform any of the inspection set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Records. for the contractor shall be maint
14、ained by the contractor and made available to the Government upon the Governments request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. Inspection. Inspection, as used herein, is
15、 defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item. performed on the audiometer when a first article sample is required.
16、tests. power voltage and frequency. Responsibility for inspection. Unless otherwise specified in The Records of examinations and tests performed by or First article inspection. First article inspection shall be This inspection shall include the examination and all All tests are to be performed using
17、 each specified input 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-I_._- _I_ A-A-53777 65 7777774 0033876 5 Quality Conformance Sampling. A-A- 5 3 7 7 7 I Sampling for inspection. testing shall be conducted in accordance with MIL-STD-105, with a
18、n inspection level of S-2 and an AQL of 1.5 (percent defective). The unit of product for sampling purposes shall be one audiometer. Sampling for examination and Quality conformance inspection. Quality conformance inspection shall include the examination and tests specified herein. Examination and Te
19、sting. Examination. The audiometer shall be examined to determine compliance with the requirements of this specification in accordance with the defects listed below. Defects Any component part including accessories missing, broken or otherwise unusable. Exposed edge not free of burrs or sharpness. I
20、dentification marking not complete, permanent, and legible. Finish not free of chips, cracks, scratches and other appearance defects. Any component not secured in place. Testing. Testing shall be conducted to determine compliance First article testing requires all with specified requirement. tests l
21、isted herein. Quality Conformance testing requires the first three tests at all the specified voitages and frequencies. Testing shall be performed after a 30 minute warm-up period. Testing shall be conducted in accordance with the methods outlined in ANSI S3.6. 4 Provided by IHSNot for ResaleNo repr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA537771988AUDIOMETERMANUALLYOPERATEDPORTABLE110220VOLT5060HZAC 交流电 5060 赫兹 110220 伏特 便携式 操作 听度计 PDF

链接地址:http://www.mydoc123.com/p-701683.html