DLA MIL-H-42067-1989 HUMIDIFIER OPERATING ROOM《手术室加湿器》.pdf
《DLA MIL-H-42067-1989 HUMIDIFIER OPERATING ROOM《手术室加湿器》.pdf》由会员分享,可在线阅读,更多相关《DLA MIL-H-42067-1989 HUMIDIFIER OPERATING ROOM《手术室加湿器》.pdf(13页珍藏版)》请在麦多课文档分享上搜索。
1、-.- . - * NIL-H-42067 45 W 7777706 0236557 7 OB-27-2-8 MIL-H-42067 15 June 1989 MILITARY SPECIFICATION HUMIDIFIER, OPERATING ROOM This specification is approved for use by all departments and agencies of the Department of Defense. 1. SCOPE 1.1 Scope. This specification covers one type of humidifier
2、for use in the operating rooms of the Medical Unit - Self Contained Transportable (MUST). . 2. APPLICABLE DOCUMENTS 2.1 Government documents. 2.1.1 Specifications, standards. The following specifications and standards form a part of this document to the extent specified herein. Unless otherwise spec
3、ified, the issues of these documents shall be those listed in the issue of the Department of Defense Index of Specifications and Standards (DODISS) and supplement thereto, cited in the solicitation (see 6.2). 0 IBeneficial comments (recommendations, additions, deletions) and lany pertinent data whic
4、h may be of use in improving this document should be addressed to: Headquarters, Defense (Personnel Support Center, ATTN: Directorate of Medical Materiel IDPSC-RST, 2800 South 20th Street, Philadelphia, PA 19101, by lusing the self-addressed Standardization Document Improvement IProposal (DD Form 14
5、26) appearing at the end of this document . lor by letter. AMSC/NA FSC 4520 DISTRIBUTION STATEMENT A. Approved for public release: 0 distribution is unlimited. r- . 1 THIS DOCUMENT CONTAI.NS 13 PAGES. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-*
6、 MIL-H-42067 SPECIFICATIONS Fedexal PPP-B-636 Military MIL-L-10547 MIL-M-7298 MIL-P-116 MIL-S-7742 MIL-T-704 STANDARDS Federal. FED-STD-595 Mi 1 it ar y MIL-STD-105 MIL-STD-129 MIL-STD-130 MIL-STD-147 MIL-STD-794 Box, Shipping, Fiberboard Liners, Case, and Sheet, Overwrap; Water- Manuals, Commexcial
7、 Off-The-Shelf Preservation, Methods of Screw Threads, Standard, Optimum Selected Treatment and Painting of Material Vaporproof or Waterproof, Flexible Series: General Specification for Colors Sampling Procedures and Tables for Marking for Shipment and Storage Identification Marking of U.S. Military
8、 Palletized Unit Loads Parts and Equipment, Procedures for Inspection by Attributes Property Packaging of (Unless othexwise indicated, copies of federal and military specifications and standards are available from the Naval Publications and Forms Center, (ATTN: NPODS), 5801 Tabor Avenue, Philadelphi
9、a, PA 19120-5099). 2.1.2 Other Government documents, drawings, and publications. The Eollowing other Government documents, drawings, and publications form a part of this document to the extent specified herein. Unless otherwise specified, the issues are those cited in the solicitation. 2 Provided by
10、 IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- MIL-H-420b7 45 9999906 021b561 7 a MIL-H-42067 DRAWING Defense Personnel Support Center (DPSC) 24346 Humidifier, Shelter, Surgery (MUST) 2.2 Non-Government publications. The following documents form a part of thi
11、s specification to the extent specified herein. Unless otherwise specified, the issues of the documents which are DOD adopted are those listed in the issue of the DODISS specified in the solicitation. Unless otherwise specified, the issues of documents not listed in the DODISS are issues of the docu
12、ments cited in the solicitation. Underwriters Laboratories, Inc. (UL) UL-73 - Electric Motor-Operated Appliance UL-507 - Electric Fans (Application for copies should be addressed to the Underwriters Laboratories, Inc. 333 Pfingsten Road, Northbrook, IL 60062; 1285 Walt Whitman Road, Melville, Long I
13、sland, NY 11746; 1655 Scott Boulevard, Santa Clara, CA 05050). (Nongovernment standards and other publications are normally available from the organizations that prepare or distribute the documents. These documents also may be available in or through libraries or other informational services.) 2.3 O
14、rder of precedence. In the event of a conflict between the text of this document and the references cited herein, (except for associated detail specifications, specification sheets or MS standards) the text of this document takes precedence. Nothing in this document, however, supersedes applicable l
15、aws and regulations unless a specific exemption has been obtained. 3. REQUIREMENTS 3.1 Specification sheets. The individual item requirements shall be as specified herein and in accordance with the applicable specification sheets. In the event of any conflict between requirements of this specificati
16、on and the specification sheets, the latter shall govern. 3.2 Fire and casualty hazards. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-. MIL-H-42067 3.2.1 The contractor shall submit proof, prior to delivery, that the items to be delivered under
17、the contract conform to the applicable requirements of UL 507. Proof shall consist of one of the following: a, UL label or UL listing of the product, provided the contractor agrees to have Underwriters Laboratories, Inc. forward a copy of the UL report directly to the contracting officer, upon the l
18、atters request. b. Laboratory report from a qualified independent laboraYory. If an independent laboratory is used, it must be adequately equipped and competent to determine compliance with UL standards. The qualification of such laboratory shall be determined by the Govemment, which reserves the ri
19、ght to require a facilitys report and to conduct a facility survey. Any laboratory report submitted shall include actual test methods, data taken, and results of tests. Detailed findings of examinations necessary to determine compliance shall be included. The test report shall also have a section th
20、at describes the test sample in such a way that the report can be used to insure identity o production items with the test sample. This section shall be supplemented with construction information such as materials used, spacings between electrical components, types of bushings, photographs, wiring d
21、iagrams, sketches, and parts list, all appropriately identified, 3.2.2 Compliance. Compliance with above paragraph does not excuse compliance with all other requirements of the contract and specifications. 3.3 MateLial, Materials shall be in accordance with DPSC drawing No. 24346 and as hereinafter
22、specified. All metals shall be corrosion-resisting or treated to resist corrosion. Materials not specified shall be of the highest quality and suitable in all respects for their intended purpose. 3.4 Design and construction. The design and construction of the humidifier shall be in accordance with D
23、PSC Drawing No. 24346 and as hereinafter specified, housing containing a water reservoir and an air circulating fan, and the other components specified on the drawing. Water level shall be maintained by a float valve in the reservoir. The water in the reservoir shall be heated by the electric heater
24、s in the reservoir:, The fan shall draw air across the evaporating water in the reservoir, and discharge the resultant humid air into the operating room of the MUST unit. provided to permit draining of the reservoir. control valve shall be provided to regulate the operation of the unit about a selec
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DLAMILH420671989HUMIDIFIEROPERATINGROOM 手术室 加湿器 PDF

链接地址:http://www.mydoc123.com/p-691369.html