BS 7887-1997 Ultrasonics - Real-time pulse-echo systems - Guide for test procedures to determine performance specifications《超声波学 实时脉冲回波系统 测定性能规范的试验程序指南》.pdf
《BS 7887-1997 Ultrasonics - Real-time pulse-echo systems - Guide for test procedures to determine performance specifications《超声波学 实时脉冲回波系统 测定性能规范的试验程序指南》.pdf》由会员分享,可在线阅读,更多相关《BS 7887-1997 Ultrasonics - Real-time pulse-echo systems - Guide for test procedures to determine performance specifications《超声波学 实时脉冲回波系统 测定性能规范的试验程序指南》.pdf(24页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD BS 7887:1997 IEC 1390:1996 Ultrasonics Real-time pulse-echo systems Guide for test procedures to determine performance specifications ICS 11.040.50BS7887:1997 This British Standard, having been prepared under the directionof the Electrotechnical Sector Board, was published underthe
2、authority of the Standards Board and comes intoeffect on 15March1997 BSI 11-1998 The following BSI references relate to the work on this standard: Committee reference EPL/87 Draft for comment 91/20276 DC ISBN 0 580 26627 3 Committees responsible for this British Standard The preparation of this Brit
3、ish Standard was entrusted to Technical Committee EPL/87, Ultrasonics, upon which the following bodies were represented: Association of British Health-Care Industries British Dental Association British Institute of Radiology British Medical Ultrasound Society British Society for Rheumatology Departm
4、ent of Health Institution of Electrical Engineers Institution of Physics and Engineering in Medicine and Biology National Physical Laboratory Institute of Laryngology and Otology Amendments issued since publication Amd. No. Date CommentsBS 7887:1997 BSI 11-1998 i Contents Page Committees responsible
5、 Inside front cover National foreword ii Introduction 1 1 Scope 1 2 Normative references 1 3 Definitions 2 4 Environmental conditions 4 5 Equipment required 4 6 Test methods 4 6.1 Instruments 4 6.2 Test settings 5 6.3 Test parameters 6 Annex A (normative) Test objects and tissue-mimicking material 1
6、0 Annex B (informative) Bibliography 17 Figure A.1 Soft tissue-mimicking test object 11 Figure A.2 Axial resolution test object 12 Figure A.3 Resolution test object 13 Figure A.4 Contrast test object 14 Figure A.5 Non-scattering spheres test object 15 Figure A.6 Slice thickness measurement and calcu
7、lation 16 List of references Inside back coverBS7887:1997 ii BSI 11-1998 National foreword This British Standard has been prepared by Technical Committee EPL/87 and is identical with Technical Report IEC 1390:1996 Ultrasonics Real-time pulse-echo systems Test procedures to determine performance spec
8、ifications, published by the International Electrotechnical Committee (IEC). A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application. Compliance with a British Standard does not of itself conf
9、er immunity from legal obligations. Cross-references International standard Corresponding British Standard IEC 50 (801):1994 BS 4727 Glossary of electrotechnical power, telecommunications, electronics, lighting and colour terms Part 3: Terms particular to telecommunications and electronics Part 3:Gr
10、oup 08:1995 Acoustics and electroacoustics (Identical) IEC 866:1987 BS 7041:1989 Specification for characteristics and calibration of hydrophones for operation in the frequency range 0.5MHz to 15MHz (Identical) IEC 1102:1991 BS EN 61102:1994 Specification for measurement and characterisation of ultr
11、asonic fields using hydrophones in the frequency range of 0.5MHz to 15MHz (Identical) IEC 1157:1992 BS EN 61157:1995 Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment (Identical) Summary of pages This document comprises a front cover, an inside front
12、cover, pages i and ii, pages1 to 18, an inside back cover and a back cover. This standard has been updated (see copyright date) and may have had amendments incorporated. This will be indicated in the amendment table on theinside front cover. BS 7887:1997 BSI 11-1998 1 Introduction An ultrasonic puls
13、e-echo scanner produces images of tissue in a scan plane by sweeping a narrow pulsed beam of ultrasound through the section of interest and detecting the echoes generated at tissue boundaries. A variety of transducer types is employed to operate in a transmit/receive mode for the ultrasonic signals.
14、 Ultrasonic scanners are widely used in medical practice to produce images of many soft-tissue organs throughout the human body. This technical report describes test procedures which should be widely acceptable and valid for a wide range of types of equipment. Manufacturers should use the report to
15、prepare their specifications, the users should use the report to check specifications. The measurements can also be carried out without interfering with the normal working conditions of the machine. The structures of the test objects have not been specified in detail rather, suitable types of overal
16、l and internal structures are described, together with typical test objects, in annex A. The specific structure of a test object should be reported, together with the results obtained using it. Similar commercial versions of these test objects are also available. The performance parameters specified
17、, and the corresponding methods of measurement have been chosen to provide a basis for comparison with the manufacturers specifications and between similar types of apparatus of different makes, intended for the same kind of diagnostic application. The manufacturers specifications should allow compa
18、rison with the results obtained from the tests in this report. Specific values of parameters and the tolerances on them have not been recommended, since these are constantly changing. Furthermore, it is intended that the sets of results and values obtained from the use of the recommended methods wil
19、l provide useful criteria for predicting the performance of equipment in appropriate diagnostic application. The procedures recommended in this report are in accordance with IEC 601-1 9 1) . Where a diagnostic system accommodates more than one option in respect of a particular system component, for
20、example the transducer, it is intended that each option be regarded as a separate system. However, it is considered that the performance of a machine is adequately specified if measurements are undertaken for the most significant combinations of machine control settings and accessories. Further eval
21、uation of equipment is obviously possible, but this should be considered as a special case rather than a routine requirement. Data relating to measuring methods, principles and equipment that are common to two or more sections of this report are given in annex A. The measurement of acoustic output p
22、ower levels and the assessment of electrical safety are dealt with in other IEC standards, and are therefore specifically excluded from this report. 1 Scope This technical report describes methods of measuring the performance of real-time medical ultrasonic imaging equipment in the frequency range 0
23、,5MHz to 15MHz. This report is relevant for real-time ultrasonic scanners based on the pulse-echo principle, for the types listed below: mechanical sector scanner; electronic phased array sector scanner; electronic linear array scanner; electronic curved array sector scanner; water-bath scanner base
24、d on any of the above four scanning mechanisms. 2 Normative references The following normative documents contain provisions which, through reference in this text, constitute provisions of this technical report. At the time of publication, the editions indicated were valid. All normative documents ar
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS78871997ULTRASONICSREALTIMEPULSEECHOSYSTEMSGUIDEFORTESTPROCEDURESTODETERMINEPERFORMANCESPECIFICATIONS

链接地址:http://www.mydoc123.com/p-547139.html