《BS 684-1 11-1976 Methods of analysis of fats and fatty oils Physical methods Determination of penetration value《脂肪和油脂分析方法 第1部分 物理法 第11节 渗透值测定》.pdf》由会员分享,可在线阅读,更多相关《BS 684-1 11-1976 Methods of analysis of fats and fatty oils Physical methods Determination of penetration value《脂肪和油脂分析方法 第1部分 物理法 第11节 渗透值测定》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、BRITISH STANDARD CONFIRMED APRIL1993 BS684-1.11: 1976 Methods of analysis of Fats and fatty oils Part1: Physical methods Section 1.11 Determination of penetration value IMPORTANT NOTE. It is essential that this Section be read in conjunction with the information in the “General introduction” to BS68
2、4, which is published separately. UDC665.1.014:543:539.533.083:620.178.15BS684-1.11:1976 This British Standard, having been prepared under the directionof the Oils and Fats Standards Committee, was published under the authorityofthe Executive Boardon 29October1976 BSI12-1999 The following BSI refere
3、nces relate to the work on this standard: Committee reference OFC/24 Draft for comment74/50396DC ISBN 0 580 09165 1 A British Standard does not purport to include all the necessary provisions of a contract. Users of British Standards are responsible for their correct application. Compliance with a B
4、ritish Standard does not of itself confer immunity from legal obligations. Summary of pages This document comprises a front cover, an inside front cover, pagesi andii, pages1 to3 and a back cover. This standard has been updated (see copyright date) and may have had amendments incorporated. This will
5、 be indicated in the amendment table on the inside front cover. Amendments issued since publication Amd. No. Date of issue CommentsBS684-1.11:1976 BSI 12-1999 i Contents Page 0 Scope 1 1 Method1. For fats such as margarine, butter and shortenings 1 1.1 Field of application 1 1.2 Reference 1 1.3 Defi
6、nition 1 1.4 Principle 1 1.5 Apparatus 1 1.6 Sampling and preparation of sample 1 1.7 Conditioning 1 1.8 Procedure 1 1.9 Expression of results 2 1.10 Test report 2 2 Method 2. For hard confectionery butters 2 2.1 Field of application 2 2.2 References 2 2.3 Definition 2 2.4 Principle 2 2.5 Apparatus
7、2 2.6 Sampling and preparation of sample 2 2.7 Conditioning 2 2.8 Procedure 3 2.9 Expression of results 3 2.10 Test report 3 Figure 1 Penetrometer cone 3 Figure 2 Penetration needle 3ii blankBS684-1.11:1976 BSI 12-1999 1 0 Scope This Section describes two methods for the determination of penetration
8、 value. The procedure described in method1 is based on the appropriate parts of ISO2137 and the corresponding appropriate parts of IP50/59 “Standard method of test for cone penetration of lubricating grease” and is suitable for fats such as margarine, butter and shortenings. The modified procedure d
9、escribed in method2, using a needle identical to that described in IP49/67 “Penetration of bitumen”, is suitable for hard confectionary butters, cocoa butter, chocolate,etc. 1 Method 1. For fats such as margarine, butter and shortenings 1.1 Field of application The method described is suitable for t
10、he determination of the penetration value of fats such as margarine, butter and shortenings. 1.2 Reference The following standards publication is referred to in this Section: ISO2137, Petroleum products Lubricating grease Determination of cone penetration. 1.3 Definition For the purposes of this met
11、hod the following definition applies. penetration value the distance, in tenths of a millimetre, that a standard cone penetrates vertically into the sample under specified conditions. It is a measure of the consistency, or hardness, of the sample 1.4 Principle A prepared sample is conditioned at a s
12、elected temperature and the depth of penetration of a conical plunger into the sample is measured under specified conditions. 1.5 Apparatus The following items of apparatus are required (seealso BS684: General introduction). 1.5.1 Penetrometer, an apparatus of any suitable pattern that permits a con
13、e(1.5.2) to move vertically into the sample without appreciable friction, and that indicates the depth of penetration of the cone to the nearest0.1mm. The base of the instrument shall be fitted with levelling screws. 1.5.2 Penetration cone, with a conical body of brass or corrosion-resistant steel,
14、and a detachable hardened steel or stainless steel tip. The outside dimensions shall conform to those shown in Figure 1 (the interior construction and dimensions, shown without tolerances, are intended merely as suggestions and may be modified as required). The outside surface of the cone and tip sh
15、all be given a very smooth finish. When fitted in the penetrometer, the cone is attached to a metal cone holder that moves with the cone. The total moving mass of the cone and cone holder shall be150g. 1.5.3 Constant temperature cabinet (fan circulation), thermostatically controlled to an accuracy o
16、f 0.2 C. NOTEThe temperatures normally used are5,10,20 and25 C. 1.5.4 Stopwatch 1.6 Sampling and preparation of sample See BS684: General introduction. 1.6.1 Small packages. Carry out the measurement directly on the sample in the container. 1.6.2 Large packages. Remove1kg to2kg portions of sample in
17、 the form of rectangular blocks or cylinders, not less than50mm thick, taking not less than4 portions for measurement. Wrap in greaseproof paper or place in a carton. It is essential that the portions of sample used are not deformed or distorted. Place soft samples in rigid containers such as sample
18、 rings. If the testing surface of a sample is irregular, cut it carefully to produce a smooth uniform surface. NOTEIt is important that the top and bottom surfaces of the block samples are flat and parallel to ensure that the blocks cannot wobble on the base plate and that the cone axes are perpendi
19、cular to the top surfaces. 1.7 Conditioning Keep the samples or portions of sample, prepared as described in1.6, in the constant temperature cabinet at the required temperature for not less than16h. NOTEFor certain fats exhibiting strong polymorphic behaviour, it may be necessary to extend the condi
20、tioning time to at least3 days to obtain a stable structure. 1.8 Procedure Place the penetrometer on a firm flat surface. Insert the cone and level the penetrometer by means of the levelling screws in the base. Set the pointer on the calibrated scale to the zero position. Ensure that the temperature
21、 of the penetrometer, and especially that of the cone and table, are approximately equal to that of the sample.BS684-1.11:1976 2 BSI 12-1999 NOTEIf the temperature of the cone and plate differs substantially from the temperature of the sample, unreliable results may be obtained. A warmer cone, for e
22、xample, will melt some of the sample and lead to inaccuracies. Remove the sample from the constant temperature cabinet and immediately place it on the penetrometer table. Make the measurements as soon as possible in order to minimize any change in the temperature of the sample. Lower the penetromete
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- BS6841111976METHODSOFANALYSISOFFATSANDFATTYOILSPHYSICALMETHODSDETERMINATIONOFPENETRATIONVALUE 脂肪 油脂 分析

链接地址:http://www.mydoc123.com/p-546283.html