ASTM F1218-1989(2014) Standard Specification for Bronchoscopes &40 Rigid&41 《支气管镜40 刚性41的标准规范》.pdf
《ASTM F1218-1989(2014) Standard Specification for Bronchoscopes &40 Rigid&41 《支气管镜40 刚性41的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1218-1989(2014) Standard Specification for Bronchoscopes &40 Rigid&41 《支气管镜40 刚性41的标准规范》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1218 89 (Reapproved 2014)Standard Specification forBronchoscopes (Rigid)1This standard is issued under the fixed designation F1218; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number i
2、n parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers definitions and requirementsfor rigid bronchoscopes and certain bronchoscopic accessoriesused in the practice of medi
3、cine.2. Referenced Documents2.1 ASTM Standards:2F1054 Specification for Conical Fittings (Withdrawn 2005)33. Terminology3.1 Definitions:3.1.1 bronchoscopes (rigid)a medical instrument havingviewing means, with or without optics, introduced into thelarynx or tracheobronchial airway, or both, through
4、a natural orsurgically created body opening for examination, diagnosis, ortherapy, and intended to be unyielding to natural or surgicallycreated body cavities.3.1.2 optical endoscopea medical instrument with optics,having viewing means, which may be introduced into a bodycavity through a bronchoscop
5、e, or through a naturally orsurgically created body opening for examination, diagnosis, ortherapy. An optical endoscope may be of rigid or flexibledesign.3.1.3 endoscopic accessorya medical instrument insertedthrough a bronchoscope for diagnosis or therapy.3.1.3.1 rigid accessoryan accessory whose i
6、nsertion por-tion is intended to be unyielding to natural or surgically createdbody cavities or instrument lumens.3.1.3.2 flexible accessoryan accessory whose insertionportion is intended to conform to natural or surgically createdbody cavities or instrument lumens.3.1.3.3 DiscussionThe intent is to
7、 include forceps, snares,electrodes, and other such instruments which can be passedthrough a bronchoscope or with a bronchoscope throughanother accessory. The intent is also to exclude certaininstruments, such as electrosurgical units, light sources, othersuch instruments, and ventilation systems ex
8、ternal to thebronchoscope.3.1.4 distalthe location of that portion of a bronchoscopeor endoscopic accessory which is farther from the user thansome reference point.3.1.4.1 DiscussionThe terms given in 3.1.4 and 3.1.5,commonly used in endoscopy, are defined in their most generalform to avoid the need
9、 for such definitions as “distal tip,”“distal end,”“ area proximal to .,” “X cm distal to the . .”3.1.5 proximalthe location of that portion of a broncho-scope or endoscopic accessory which is closer to the user thansome reference point.3.1.5.1 See 3.1.4.3.1.6 insertion portionthat portion of a bron
10、choscope orendoscopic accessory which is intended to be inserted into anatural or surgically created body opening; or which is in-tended to be inserted into the lumen of a bronchoscope orendoscopic accessory.3.1.6.1 DiscussionAlthough the term defined seems self-explanatory, different expressions fo
11、r the same portion of theinstrument are used by different manufacturers.3.1.7 maximum insertion portion widththe maximum ex-ternal width of a bronchoscope or endoscopic accessorythroughout the length of the insertion portion.3.1.7.1 DiscussionBy defining external sizes as maximaand internal sizes as
12、 minima, sufficient instrument informationfor selection of an instrument will be provided to users.3.1.8 minimum lumen widththe minimum internal widthof a bronchoscope or endoscopic accessory through which abronchoscope or endoscopic accessory is intended to pass.3.1.8.1 See 3.1.7.3.1.9 working leng
13、ththe maximum length of the insertionportion.3.1.10 overall lengththe distance between the proximaland distal ends of a rigid bronchoscope or bronchoscopicaccessory, expressed in metric units.3.1.11 field of viewthe size of the object field viewedthrough an optical endoscope and stated by the vertex
14、 angle (in1This specification is under the jurisdiction of ASTM Committee F29 onAnesthetic and Respiratory Equipment and is the direct responsibility of Subcom-mittee F29.12 on Airways, Bronchoscopes and Laryngoscopes.Current edition approved Jan. 1, 2014. Published March 2014. Originallyapproved in
15、 1989. Last previous edition approved in 2007 as F1218 89 (2007).DOI: 10.1520/F1218-89R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Sum
16、mary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1degrees) of the cone whose vertex is at the distal windowsurface of the i
17、nstrument.3.1.11.1 DiscussionThe endoscope is not intended to be incontact with the object. (See Fig. 1.)3.1.11.2 DiscussionThe definitions given in 3.1.11 and3.1.12, as illustrated, describe characteristics of optical endo-scopes paramount for effective use, thereby allowing the userto select instr
18、uments of specific values for particular medicalpurposes.3.1.12 direction of viewthe location of the center of theobject field relative to the normal axis of the optical endoscope,and stated as the angle (in degrees) between the normal axis ofthe optical endoscope and the center of the field of view
19、. (SeeFig. 2.)3.1.12.1 See 3.1.11.3.1.13 controllable portionthat part of the insertion por-tion of an optical endoscope whose motion is intended to beremotely controlled by the user.3.1.13.1 Many optical endoscopes incorporate provisionsfor remote control of the motion of the distal tip by the user
20、.The mechanisms for these controls vary widely, and areobviously of interest to the user. In order to avoid constraint offuture designs, control characteristics are defined in the mostgeneral terms.3.1.14 French (Charriere) size (Fr)a measure of the sizeof certain endoscopic accessories of circular
21、and noncircularcross-section. The measure has several definitions:Fr 5 3 3 u/3.1416 pi! for noncircular cross 2 sectionwhere u is the perimeter of the cross-section, mm.Fr 5 3.1416pi! 3 d for circular cross 2 sectionwhere d is the diameter of the cross-section, mm.Fr 5 3.0 3 d for circular cross 2 s
22、ectionwhere d is the diameter of the cross-section, mm.NOTE 1Fr = 3.0 d for circular cross-section, commonly used bymany manufacturers of endoscopic accessories.3.1.14.1 DiscussionThis measure of size enjoys world-wide favor in many branches of medicine. In usage, themeasurement is imprecise, as sho
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF121819892014STANDARDSPECIFICATIONFORBRONCHOSCOPES40RIGID41 支气管 40 刚性 41 标准规范 PDF

链接地址:http://www.mydoc123.com/p-534441.html