ASTM E1570-2011 Standard Practice for Computed Tomographic (CT) Examination《计算机层析(CT)检查标准操作规程》.pdf
《ASTM E1570-2011 Standard Practice for Computed Tomographic (CT) Examination《计算机层析(CT)检查标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1570-2011 Standard Practice for Computed Tomographic (CT) Examination《计算机层析(CT)检查标准操作规程》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1570 11Standard Practice forComputed Tomographic (CT) Examination1This standard is issued under the fixed designation E1570; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in paren
2、theses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope*1.1 This practice is for computed tomography (CT), whichmay be used to n
3、ondestructively disclose physical features oranomalies within a test object by providing radiological densityand geometric measurements. This practice implicitly assumesthe use of penetrating radiation, specifically X-ray and g-ray.1.2 CT systems utilize a set of transmission measurementsmade along
4、paths through the test object from many differentdirections. Each of the transmission measurements is digitizedand stored in a computer, where they are subsequently recon-structed by one of a variety of techniques. A treatment of CTprinciples is given in Guide E1441.1.3 CT is broadly applicable to a
5、ny material or test objectthrough which a beam of penetrating radiation passes. Theprincipal advantage of CT is that it provides densitometric (thatis, radiological density and geometry) images of thin crosssections through an object without the structural superpositionin projection radiography.1.4
6、This practice describes procedures for performing CTexaminations. This practice is to address the general use of CTtechnology and thereby facilitate its use.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user o
7、f this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific safetystatements, see Section 8, NBS Handbook 114, and FederalStandards 21 CFR 1020.40 and 29 CFR 1910.96.2. Referenced Documents2.1 ASTM Standa
8、rds:2E1316 Terminology for Nondestructive ExaminationsE1441 Guide for Computed Tomography (CT) ImagingE1695 Test Method for Measurement of Computed Tomog-raphy (CT) System Performance2.2 NIST Standard:ANSI N43.3 General Radiation Safety Installations UsingNon-Medical X-Ray and Sealed Gamma Sources u
9、p to 10MeV32.3 Federal Standards:421 CFR 1020.40 Safety Requirements of Cabinet X RaySystems29 CFR 1910.96 Ionizing Radiation2.4 ASNT Documents:5SNT-TC-1A Recommended Practice for Personnel Qualifi-cation and Certification in Nondestructive TestingANSI/ASNT-CP-189 Qualification and Certification ofN
10、ondestructive Testing Personnel2.5 Military Standard:MIL-STD-410 Nondestructive Testing Personnel Qualifica-tion and Certification42.6 AIA Standard:NAS-410 Certification and Qualification of NondestructiveTesting Personnel63. Terminology3.1 DefinitionsFor definitions of terms used in this guide,refe
11、r to Terminology E1316 and Annex A1 in Guide E1441.4. Summary of Practice4.1 Requirements in this practice are intended to control thereliability and quality of the CT images.4.2 CT systems are made up of a number of subsystems; thefunction served by each subsystem is common in almost all CTscanners
12、. Section 7 describes the following subsystems:4.2.1 Source of penetrating radiation,4.2.2 Radiation detector or an array of detectors,1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.01 onRadiology (X and
13、Gamma) Method.Current edition approved July 1, 2011. Published July 2011. Originally approvedin 1993. Last previous edition approved in 2005 as E1570 - 00(2005)1. DOI:10.1520/E1570-11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm
14、.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from National Institute of Standards and Technology (NIST), 100Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http:/www.nist.gov.4Available from Standardization
15、Documents Order Desk, DODSSP, Bldg. 4,Section D, 700 Robbins Ave., Philadelphia, PA 19111-5098, http:/dodssp.daps.dla.mil.5Available fromAmerican Society for Nondestructive Testing (ASNT), P.O. Box28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http:/www.asnt.org.6Available from Aerospace Indust
16、ries Association of America, Inc. (AIA), 1000Wilson Blvd., Suite 1700,Arlington, VA22209-3928, http:/www.aia-aerospace.org.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United Sta
17、tes.4.2.3 Mechanical scanning assembly, and4.2.4 Computer system including:4.2.4.1 Image reconstruction software/hardware,4.2.4.2 Image display/analysis system,4.2.4.3 Data storage system, and4.2.4.4 Operator interface.4.3 Section 8 describes and defines the procedures forestablishing and maintainin
18、g quality control of CT services.4.4 The extent to which a CT image reproduces an object ora feature within an object is influenced by spatial resolution,statistical noise, slice plane thickness, and artifacts of theimaging system. Operating parameters should strike an overallbalance between image q
19、uality, inspection time, and cost.These parameters should be considered for CT system configu-rations, components, and procedures. The setting and optimi-zation of CT system parameters is discussed in Section 9.4.5 Methods for the measurement of CT system perfor-mance are provided in Section 10 of t
20、his practice.5. Significance and Use5.1 This practice is applicable for the systematic assessmentof the internal structure of a material or assembly using CTtechnology. This practice may be used for review by systemoperators, or to prescribe operating procedures for new orroutine test objects.5.2 Th
21、is practice provides the basis for the formation of aprogram for quality control and its continuation throughcalibration, standardization, reference samples, inspectionplans, and procedures.6. Basis of Application6.1 This practice provides the approach for performing CTexaminations. Supplemental inf
22、ormation covering specificitems where agreement between supplier7and purchaser8arenecessary is required. Generally the items are applicationspecific or performance related, or both. Examples include:system configuration, equipment qualification, performancemeasurement, and interpretation of results.
23、7. System Configuration7.1 Many different CT system configurations are possibleand it is important to understand the advantages and limitationsof each. It is important that the optimum system parameters beselected for each examination requirement, through a carefulanalysis of the benefits and limita
24、tions of the available systemcomponents and the chosen system configuration.7.2 Radiation SourcesWhile the CT systems may utilizeeither gamma-ray or X-ray generators, the latter is used formost applications. For a given focal spot size, X-ray generators(that is, X-ray tubes and linear accelerators)
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15702011STANDARDPRACTICEFORCOMPUTEDTOMOGRAPHICCTEXAMINATION 计算机 层析 CT 检查 标准 操作规程 PDF

链接地址:http://www.mydoc123.com/p-528966.html