ASTM D7278-2011 5343 Standard Guide for Prediction of Analyzer Sample System Lag Times《预测分析仪采样系统滞后时间的标准指南》.pdf
《ASTM D7278-2011 5343 Standard Guide for Prediction of Analyzer Sample System Lag Times《预测分析仪采样系统滞后时间的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7278-2011 5343 Standard Guide for Prediction of Analyzer Sample System Lag Times《预测分析仪采样系统滞后时间的标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7278 11Standard Guide forPrediction of Analyzer Sample System Lag Times1This standard is issued under the fixed designation D7278; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in
2、 parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONLag time, as used in this guide, is the time required to transport a representative sample from theprocess tap to the analyzer. Sample system de
3、signs have infinite configurations so this guide gives theuser guidance, based on basic design considerations, when calculating the lag time of online sampledelivery systems. Lag time of the analyzer sample system is a required system characteristic whenperforming system validation in Practice D3764
4、 or D6122 and in general the proper operation of anyonline analytical system. The guide lists the components of the system that need to be considered whendetermining lag time plus a means to judge the type of flow and need for multiple flushes beforeanalysis on any sample.1. Scope1.1 This guide cove
5、rs the application of routine calculationsto estimate sample system lag time, in seconds, for gas, liquid,and mixed phase systems.1.2 This guide considers the sources of lag time from theprocess sample tap, tap conditioning, sample transport, pre-analysis conditioning and analysis.1.3 Lag times are
6、estimated based on a prediction of flowcharacteristics, turbulent, non turbulent, or laminar, and thecorresponding purge requirements.1.4 Mixed phase systems prevent reliable representativesampling so system lag times should not be used to predictsample representation of the stream.1.5 The values st
7、ated in inch-pound units are to be regardedas standard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate
8、safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D3764 Practice for Validation of the Performance of Pro-cess Stream Analyzer SystemsD6122 Practice for Validation of the Performance of Mul-tivariate Online,
9、 At-Line, and Laboratory Infrared Spec-trophotometer Based Analyzer Systems3. Terminology3.1 Definitions:3.1.1 continuous analyzer unit cycle timethe time intervalrequired to replace the volume of the analyzer measurementcell.3.1.2 intermittent analyzer unit cycle timethe time inter-val between succ
10、essive updates of the analyzer output.3.1.3 purge volumethe combined volume of the fullanalyzer sampling and conditioning systems.3.1.4 sample system lag timethe time required to transporta representative sample from the process tap to the analyzer.3.1.5 system response timethe sum of the analyzer u
11、nitresponse time and the analyzer sample system lag time.3.2 Abbreviations:3.2.1 I.D.Internal Diameter3.2.2 ReReynolds Number4. Summary4.1 The lag time of an analyzer sample system is estimatedby first determining the flow characteristics. The flow isassigned as turbulent or non-turbulent to assign
12、the number ofpurges required to change out the sample. Based on thehardware employed in the sample system an estimation of thelag time can be calculated.1This guide is under the jurisdiction of ASTM Committee D02 on PetroleumProducts and Lubricants and is the direct responsibility of Subcommittee D0
13、2.25 onPerformance Assessment and Validation of Process Stream Analyzer Systems.Current edition approved Oct. 15, 2011. Published December 2011. Originallyapproved in 2006. Last previous edition approved in 2006 as D727806. DOI:10.1520/D7278-11.2For referenced ASTM standards, visit the ASTM website,
14、 www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5
15、. Significance and Use5.1 The analyzer sample system lag time estimated by thisguide can be used in conjunction with the analyzer output toaid in optimizing control of blender facilities or process units.5.2 The lag time can be used in the tuning of controlprograms to set the proper optimization fre
16、quency.5.3 The application of this guide is not for the design of asample system but to help understand the design and toestimate the performance of existing sample systems. Addi-tional detailed information can be found in the referencesprovided in the section entitled Additional Reading Material.6.
17、 Basic Design Considerations6.1 Acceptable Lag TimeA one to two minute samplesystem lag time should be maintained to give acceptableperformance. Flow is a key component in the determination ofsample system lag time, and in most systems the desiredsystem lag time is impossible to achieve with maximum
18、allowable sample flow rate to the analyzer. A fast loop orbypass can be ways to improve lag time by increasing samplevelocity. A slipstream is taken from the bypass to feed theanalyzer at its optimum flowrate. Excess sample in the slip-stream is vented to atmosphere, to flare or to the process strea
19、mdependent upon application and regulatory requirements.6.2 Physical State of Sample:6.2.1 Liquid SamplesPressure drop properties often gov-ern the design of a liquid system. This is due for the most parton the close relationship between pressure drop and systemflowrate and the fixed pressure differ
20、ential available from theprocess for sample transport. The sizing of the sample compo-nents is a tradeoff between pressure drop and sample flowrate.High sample flowrates in small sized component systems causehigh-pressure drops and low sample transport times. The sameflowrate in a larger tubing syst
21、em will yield significant im-provements in pressure drop through the system, but will alsosignificantly increase the time for sample transport.6.2.2 Vapor SamplesVapor phase sampling is governedless by pressure drop and more by pressure compressionproperties of gases relative to liquids. In compress
22、ible gasesthe higher the pressure in a given volume, the more sample ispresent in that volume. For this reason, and different fromliquids, the selection and location of pressure regulatingdevices in the vapor sample system has a significant impact onthe overall system design. The optimal location fo
23、r a high-pressure regulator in a vapor sample is immediately down-stream of the sample tap or high-pressure location therebylimiting the volume of the system under high pressure. Sincethe density of a compressible fluid is a function of the pressure,compressible fluid flow rate calculations are some
24、times doneover segmental lengths where average properties adequatelyrepresent the fluid conditions of the line segment.6.2.3 Liquid to Vapor SamplesA change of phase due tosample vaporization can also impact the sample lag time. Thevolume change from the liquid phase to the vapor phase issubstantial
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD727820115343STANDARDGUIDEFORPREDICTIONOFANALYZERSAMPLESYSTEMLAGTIMES 预测 分析 采样系统 滞后 时间 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-525027.html