1、Page 1 of 1 2015 ASHRAE. All Rights reserved. Interpretation IC 170-2013-7 of ANSI/ASHRAE/ASHE Standard 170-2013 Ventilation of Health Care Facilities Date Approved: June 30, 2015 Request from: Travis English, Kaiser Permanente, 1800 Harrision Street, Oakland, CA 94612. Reference: This request for i
2、nterpretation refers to the requirements in ANSI/ASHRAE/ASHE Standard 170-2013, Table 7.1, regarding design temperature range. Background: Table 7.1, Column 8, lists design temperature ranges. For the three spaces in this interpretation request (under Central Medical and Surgical Supply), Soiled or
3、decontamination room (entry 38), Clean workroom (entry 39), and Sterile storage (entry 40), the temperature range is 72F-78F (22C-26C). These temperatures are not specifically cited to a particular body of evidence. Per IC 170-2013-2 of ANSI/ASHRAE/ASHE Standard 170-2013, these temperature ranges do
4、 not provide Thermal Comfort for Human Occupancy (i.e. compliance to ASHRAE Standard 55-2013). Another ANSI standard, ANSI/AAMI ST79, authored by the Association for the Advancement of Medical Instrumentation (AAMI) calls for a temperature range of 60-65 F for “decontamination area” (3.3.6.5). The A
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASHRAE170INT72015VENTILATIONOFHEALTHCAREFACILITIESPDF

链接地址:http://www.mydoc123.com/p-450214.html