1、ISO/ASTM 51401:2013(E)Standard Practice forUse of a Dichromate Dosimetry System1This standard is issued under the fixed designation ISO/ASTM 51401; the number immediately following the designation indicates theyear of original adoption or, in the case of revision, the year of last revision.1. Scope1
2、.1 This practice covers the preparation, testing, and proce-dure for using the acidic aqueous silver dichromate dosimetrysystem to measure absorbed dose to water when exposed toionizing radiation. The system consists of a dosimeter andappropriate analytical instrumentation. For simplicity, the sys-t
3、em will be referred to as the dichromate system. The dichro-mate dosimeter is classified as a type I dosimeter on the basisof the effect of influence quantities. The dichromate systemmay be used as either a reference standard dosimetry system ora routine dosimetry system.1.2 This document is one of
4、a set of standards that providesrecommendations for properly implementing dosimetry inradiation processing, and describes a means of achievingcompliance with the requirements of ISO/ASTM Practice52628 for the dichromate dosimetry system. It is intended to beread in conjunction with ISO/ASTM Practice
5、 52628.1.3 This practice describes the spectrophotometric analysisprocedures for the dichromate system.1.4 This practice applies only to gamma radiation,X-radiation/bremsstrahlung, and high energy electrons.1.5 This practice applies provided the following conditionsare satisfied:1.5.1 The absorbed d
6、ose range is from 2 103to5104Gy.1.5.2 The absorbed dose rate does not exceed 600 Gy/pulse(12.5 pulses per second), or does not exceed an equivalent doserate of 7.5 kGy/s from continuous sources (1).21.5.3 For radionuclide gamma sources, the initial photonenergy shall be greater than 0.6 MeV. For bre
7、msstrahlungphotons, the initial energy of the electrons used to produce thebremsstrahlung photons shall be equal to or greater than 2MeV. For electron beams, the initial electron energy shall begreater than 8 MeV.NOTE 1The lower energy limits given are appropriate for a cylindricaldosimeter ampoule
8、of 12 mm diameter. Corrections for displacementeffects and dose gradient across the ampoule may be required for electronbeams (2). The dichromate system may be used at lower energies byemploying thinner (in the beam direction) dosimeter containers (see ICRUReport 35).1.5.4 The irradiation temperatur
9、e of the dosimeter shall beabove 0C and should be below 80C.NOTE 2The temperature coefficient of dosimeter response is knownonly in the range of 5 to 50C (see 5.2). Use outside this range requiresdetermination of the temperature coefficient.1.6 This standard does not purport to address all of thesaf
10、ety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in 9.3.2. Referenced documen
11、ts2.1 ASTM Standards:3E170 Terminology Relating to Radiation Measurements andDosimetryE178 Practice for Dealing With Outlying ObservationsE275 Practice for Describing and Measuring Performance ofUltraviolet and Visible SpectrophotometersE666 Practice for Calculating Absorbed Dose From Gammaor X Radi
12、ationE668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining AbsorbedDose in Radiation-Hardness Testing of Electronic DevicesE925 Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidthdoes not Exceed 2 nmE958 Pra
13、ctice for Estimation of the Spectral Bandwidth ofUltraviolet-Visible Spectrophotometers2.2 ISO/ASTM Standards:351261 Practice for Calibration of Routine Dosimetry Sys-tems for Radiation Processing51707 Guide for Estimating Uncertainties in Dosimetry forRadiation Processing1This practice is under the
14、 jurisdiction of ASTM Committee E61 on RadiationProcessing and is the direct responsibility of Subcommittee E61.02 on DosimetrySystems, and is also under the jurisdiction of ISO/TC 85/WG 3.Current edition approved Sept. 14, 2013. Published November 2013. Originallypublished asASTM E 1401 91.ASTM E 1
15、401 961was adopted by ISO in 1998with the intermediate designation ISO 15561:1998(E). The present InternationalStandard ISO/ASTM 51401:2013(E) replaces ISO 15561 and is a major revision ofthe last previous edition ISO/ASTM 51401:2003(E).2The boldface numbers in parentheses refer to the bibliography
16、at the end of thispractice.3For referenced ASTM and ISO/ASTM standards, visit the ASTM website,www.astm.org, or contact ASTM Customer Service at serviceastm.org. ForAnnual Book of ASTM Standards volume information, refer to the standardsDocument Summary page on the ASTM website. ISO/ASTM Internation
17、al 2017 All rights reservedThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization
18、Technical Barriers to Trade (TBT) Committee.152628 Practice for Dosimetry in Radiation Processing52701 Guide for Performance Characterization of Dosim-eters and Dosimetry Systems for Use in Radiation Pro-cessing2.3 ISO/IEC Standards:417025 General Requirements for the Competence of Testingand Calibr
19、ation Laboratories2.4 Joint Committee for Guides in Metrology (JCGM)Reports:5JCGM 100:2008, GUM 1995, with minor corrections,Evaluation of measurement data Guide to the Expres-sion of Uncertainty in Measurement2.5 International Commission on Radiation Units and Mea-surements (ICRU) Reports:6ICRU Rep
20、ort 35 Radiation Dosimetry: Electrons With InitialEnergies Between 1 and 50 MeVICRU Report 80 Dosimetry Systems for Use in RadiationProcessingICRU Report 85a Fundamental Quantities and Units forIonizing Radiation3. Terminology3.1 Definitions:3.1.1 approved laboratorylaboratory that is a recognizedna
21、tional metrology institute; or has been formally accredited toISO/IEC 17025; or has a quality system consistent with therequirements of ISO/IEC 17025.3.1.1.1 DiscussionA recognized national metrology insti-tute or other calibration laboratory accredited to ISO/IEC17025 should be used in order to ens
22、ure traceability to anational or international standard. A calibration certificateprovided by a laboratory not having formal recognition oraccreditation will not necessarily be proof of traceability to anational or international standard.3.1.2 reference standard dosimetry systemdosimetrysystem, gene
23、rally having the highest metrological qualityavailable at a given location or in a given organization, fromwhich measurements made there are derived.3.1.3 type I dosimeterdosimeter of high metrologicalquality, the response of which is affected by individual influ-ence quantities in a well-defined wa
24、y that can be expressed interms of independent correction factors.3.2 Definitions of other terms used in this practice thatpertain to radiation measurement and dosimetry may be foundin ASTM Terminology E170. Definitions in E170 are compat-ible with ICRU Report 85a; that document, therefore, may beus
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIISOASTM514012013STANDARDPRACTICEFORUSEOFADICHROMATEDOSIMETRYSYSTEMPDF

链接地址:http://www.mydoc123.com/p-437110.html