ANSI AAMI 10079-3-1999 Medical Suction Equipment-Part 3 Suction Equipment Powered from a Vacuum or Pressure Source.pdf
《ANSI AAMI 10079-3-1999 Medical Suction Equipment-Part 3 Suction Equipment Powered from a Vacuum or Pressure Source.pdf》由会员分享,可在线阅读,更多相关《ANSI AAMI 10079-3-1999 Medical Suction Equipment-Part 3 Suction Equipment Powered from a Vacuum or Pressure Source.pdf(33页珍藏版)》请在麦多课文档分享上搜索。
1、ANSI/AAMI/ISO 10079-3:1999Medical Suction EquipmentPart 3: Suction Equipment Powered from a Vacuum or Pressure Source American National StandardThis document was approved and published when the U.S. TAG for TC 121 was held by ASTM, but it is now an AAMI standard. The original formatting has been mai
2、ntained, so there are some variations from the typical AAMI style. Published by Association for the Advancement of Medical Instrumentation 4301 N. Fairfax Drive, Suite 301 Arlington, VA 22203-1633 www.aami.org 2014 by the Association for the Advancement of Medical Instrumentation All Rights Reserved
3、 This publication is subject to copyright claims of ISO, and AAMI. No part of this publication may be reproduced or distributed in any form, including an electronic retrieval system, without the prior written permission of AAMI. All requests pertaining to this document should be submitted to AAMI. I
4、t is illegal under federal law (17 U.S.C. 101, et seq.) to make copies of all or any part of this document (whether internally or externally) without the prior written permission of the Association for the Advancement of Medical Instrumentation. Violators risk legal action, including civil and crimi
5、nal penalties, and damages of $100,000 per offense. For permission regarding the use of all or any part of this document, complete the reprint request form at www.aami.org or contact AAMI, 4301 N. Fairfax Drive, Suite 301, Arlington, VA 22203-1633. Phone: 703-525-4890; Fax: 703-525-1067. Printed in
6、the United States of America ISBN 1570205434 ANS/ISO 10079-3 1999 Medical Suction Equipment Part 3: Suction Equipment Powered from a Vacuum or Pressure Source Approved as an American National Standard by: ASTM International These materials are subject to copyright claims of ISO, ANSI, and ASTM Inter
7、national. Not for resale without the prior written permission of ASTM International. No part of this publication may be reproduced in any form, including an electronic retrieval system, without the prior written permission of ASTM International. ASTM deviations to ISO 10079-3: 1999 2 Normative refer
8、ences (Add the following) CGA Z168.11-94, Vacuum Devices Used for Suction and Drainage Rationale for ASTM deviation: CGA Z168.11-94 has requirements for intermittent and interrupted regulators that are not part of ISO 10079-3 5.2 Suction tubing (Add the following) Suction tubing supplied with the su
9、ction equipment shall have a minimum length of 1.3 m. Rationale for ASTM deviation: This minimum length is specified in ISO 10079-1and 10079-2 8 Performance requirements for vacuum and flow rate (Replace the following clauses and delete Subclauses 8.5 and 8.6) 8.2 Vacuum When tested in accordance wi
10、th A.12, suction equipment shall develop a vacuum of at least 40 kPa, unless either marked “low vacuum” or marked with the maximum vacuum that can be developed. 8.3 Free Air Flow When tested in accordance with A.14, the peak free air flow shall be at least 20 L/min unless marked “low flow” in which
11、case the peak air flow shall be less than 20 L/min. Equipment intended for thoracic drainage, wound drainage, and intermittent and interrupted suction shall be exempt from the requirements of this clause. Rationale for ASTM deviation: This simplifies performance categories and allows all combination
12、s of high and flow vacuum and flow but exempts application-specific apparatus such as thoracic suction equipment. 10 Regulators (Add the following) 10.3 Intermittent Regulators 10.3.1 Intermittent Regulator Function When tested in accordance with A.19, an intermittent regulator shall cycle through i
13、ts off-time according to the manufacturers published specifications. 10.3.2 Intermittent Regulator Venting When tested in accordance with A.20, an intermittent regulator shall return to atmospheric pressure (0-kPa gauge) for at least 3 s before commencing the next on-cycle. 10.4 Interrupted Regulato
14、rs 10.4.1 Interrupted Regulator Function When tested in accordance with A.21, an interrupted regulator shall cycle between the vacuum relief set point and that of a lower level according to the manufacturers published specification. 10.4.2 Interrupted Regulator Venting When tested in accordance with
15、 A.22, an interrupted regulator shall return to a lower vacuum level before returning to the vacuum relief set point. Rationale for ASTM deviations: Above requirements can be found in CGA Z16811-94 but not ISO 10079-3 12.2 Equipment on carrying case (Change title to the following) Equipment or carry
16、ing case Rationale for ASTM deviation: Typing error in ISO document. (Replace a) with the following) a) the performance category as indicated in Clause 8 or the vacuum and flow rate ranges for patient use, with the marking visible in the normal operating position; Rationale for ASTM deviation: To ma
17、ke the marking consistent with performance. 13 Information to be supplied by manufacturer (Add the following) k) whether or not the suction equipment is suitable for use in a magnetic resonance imaging (MRI) unit. Rationale for ASTM deviation: This knowledge is necessary for safe use. Annex A Test m
18、ethods A.14 Test for free air flowrate of low vacuum equipment (Change title and test as follows) A.14 Test for free air flow With the collection container(s) empty, switch on the suction equipment with the regulator adjusted to give the maximum vacuum. Open the inlet and attach a low resistance flo
19、wmeter to it. Note the mean free air flow when stable conditions are reached. Rationale for ASTM deviation: To make the test procedure consistent with performance modifications. Vacuum level is tested in A.12. Therefore A.14 only needs to test for free air flow. A.17 Test for vacuum regulator with v
20、ariable setting (Modify as follows) Commence with the vacuum source at 79 kPa below atmospheric pressure and reduce the vacuum to 53 kPa below atmospheric pressure adjust the vacuum source from 53 kPa to 80 kPa below atmospheric pressure Rationale for ASTM deviation: Source pressure of 53 kPa is nee
21、ded in order to set the regulator to 53 kPa as required in test procedure. CGA Z168.11-94 also uses a supply pressure of 53 kPa for this test. Annex A (Add the following) A.19 Test for intermittent function of intermittent regulators (See Figure A.10) Connect the regulator to a vacuum source as reco
22、mmended by the manufacturer. Set the regulator to the intermittent mode and occlude the tubing at the patient port. Set the occluded vacuum level to a point in the middle third of the gauge range during the on cycle. With the unit still occluded, record the on time and off time during five full cycl
23、es. A.20 Test for venting of intermittent regulators (See figure A.11) Connect the regulator to a vacuum source as recommended by the manufacturer. Connect the patient side of the regulator to the vacuum side of a 2-L collection container. Occlude the patient side of the regulator. Set the regulator
24、 to the intermittent mode, occlude the tubing between the regulator and the collection container, and adjust the regulator to a point in the middle third of the gauge range during the on cycle. Release the occlusion on the tubing, ensuring that the regulator is cycling normally, and record the amoun
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIAAMI1007931999MEDICALSUCTIONEQUIPMENTPART3SUCTIONEQUIPMENTPOWEREDFROMAVACUUMORPRESSURESOURCEPDF

链接地址:http://www.mydoc123.com/p-430237.html