ANSI AAMI 10079-2-1999 Medical Suction Equipment-Part 2 Manually Powered Suction Equipment.pdf
《ANSI AAMI 10079-2-1999 Medical Suction Equipment-Part 2 Manually Powered Suction Equipment.pdf》由会员分享,可在线阅读,更多相关《ANSI AAMI 10079-2-1999 Medical Suction Equipment-Part 2 Manually Powered Suction Equipment.pdf(25页珍藏版)》请在麦多课文档分享上搜索。
1、ANSI/AAMI/ISO 10079-2:1999Medical Suction EquipmentPart 2: Manually Powered Suction Equipment American National StandardThis document was approved and published when the U.S. TAG for TC 121 was held by ASTM, but it is now an AAMI standard. The original formatting has been maintained, so there are so
2、me variations from the typical AAMI style. Published by Association for the Advancement of Medical Instrumentation 4301 N. Fairfax Drive, Suite 301 Arlington, VA 22203-1633 www.aami.org 2014 by the Association for the Advancement of Medical Instrumentation All Rights Reserved This publication is sub
3、ject to copyright claims of ISO, and AAMI. No part of this publication may be reproduced or distributed in any form, including an electronic retrieval system, without the prior written permission of AAMI. All requests pertaining to this document should be submitted to AAMI. It is illegal under feder
4、al law (17 U.S.C. 101, et seq.) to make copies of all or any part of this document (whether internally or externally) without the prior written permission of the Association for the Advancement of Medical Instrumentation. Violators risk legal action, including civil and criminal penalties, and damag
5、es of $100,000 per offense. For permission regarding the use of all or any part of this document, complete the reprint request form at www.aami.org or contact AAMI, 4301 N. Fairfax Drive, Suite 301, Arlington, VA 22203-1633. Phone: 703-525-4890; Fax: 703-525-1067. Printed in the United States of Ame
6、rica ISBN 1570205388 ANS/ISO 10079-2 1999 Medical Suction Equipment Part 2: Manually Powered Suction Equipment Approved as an American National Standard by: ASTM International These materials are subject to copyright claims of ISO, ANSI, and ASTM International. Not for resale without the prior writt
7、en permission of ASTM International. No part of this publication may be reproduced in any form, including an electronic retrieval system, without the prior written permission of ASTM International. ASTM deviations to ISO 10079-2: 1999 5.2.1 General (Replace 5.2.1 with the following) Note: Suction pe
8、rformance may be markedly affected by the length diameter and degree of collapse of the suction tubing. 5.2.1.1 If supplied, suction tubing shall have an internal diameter of not less than 6 mm. 5.2.1.2 When tested in accordance with A.2, the degree of collapse of the suction tubing supplied with th
9、e equipment shall be less than 0.5 throughout its entire length. Rationale for ASTM deviation: Structured similar to ISO 10079 - 3 6 Operational requirements (Add the following) 6.8 The performance criteria in Clause 8 shall be met using a force of less than 350 N for foot-operated equipment and les
10、s than 45 N for hand-operated equipment. Rationale for ASTM deviation: These requirements are necessary so that equipment can be operated without excessive effort on the part of the operator. These requirements can be found in CGA Z168.11-94. 8.1 Vacuum (Replace with the following) When tested in ac
11、cordance with A.6, the suction equipment shall develop a vacuum of at least 40 kPa within 10 s, unless either marked as “low vacuum” or marked with the maximum vacuum that can be developed. Rationale for ASTM deviation: This allows for equipment other than high vacuum if appropriately labeled. This
12、modification is based on CGA Z168.11-94. 8.3 Free air flow (Replace with the following) When tested in accordance with A.7, the peak free air flow shall be at least 20 L/min unless marked “low flow.” Equipment intended for thoracic drainage, wound drainage, and intermittent and interrupted suction s
13、hall be exempt from the requirements of this clause. Rationale for ASTM deviation: This allows for equipment other than high flow if appropriately labeled and exempts application-specific equipment. This modification is based on CGA Z168.11-94. 10 Marking (Replace a) with the following) For equipmen
14、t not intended for field and/or transport use, the unit shall be marked appropriately based on the performance in Clause 8. Rationale for ASTM deviation: To make marking consistent with performance. 11 Information to be supplied by manufacturer (Add the following) p) whether or not the suction equip
15、ment is suitable for use in a magnetic resonance imaging (MRI) unit. Rationale for ASTM deviation: This knowledge is necessary for safe use. A.6 Test for vacuum (Replace with the following) Set up the suction equipment with the collection container in place and, using a short tube, fit a vacuum indi
16、cator to the container inlet, totally occluding the inlet. Operate the equipment at a frequency not exceeding 2 Hz. Record the reading of the vacuum indicator after 10 s. Rationale for ASTM deviation: It may not be possible to fit a vacuum indicator directly to the container inlet. AReference number
17、ISO 10079-2:1999(E)INTERNATIONALSTANDARDISO10079-2Second edition1999-08-15Medical suction equipment Part 2:Manually powered suction equipmentAppareils daspiration mdicale Partie 2: Appareils daspiration manuelleISO 10079-2:1999(E) ISO 1999All rights reserved. Unless otherwise specified, no part of t
18、his publication may be reproducedor utilized in any form or by any means, electronic or mechanical, including photocopying andmicrofilm, without permission in writing from the publisher.International Organization for StandardizationCase postale 56 CH-1211 Genve 20 SwitzerlandInternet isoiso.chPrinte
19、d in SwitzerlandiiContents1 Scope 12 Normative references 23 Terms and definitions .34 Cleaning and sterilization .35 Design requirements .36 Operational requirements .47 Physical requirements.58 Performance requirements for vacuum and flowrate 69 Resistance to environment.610 Marking .711 Informati
20、on to be supplied by manufacturer 7Annex A (normative) Test methods.8Annex B (informative) Table of typical ranges of volume for collection containers for specific uses.12Annex C (informative) Rationale statement13 ISOISO 10079-2:1999(E)iiiForewordISO (the International Organization for Standardizat
21、ion) is a worldwide federation of national standards bodies (ISOmember bodies). The work of preparing International Standards is normally carried out through ISO technicalcommittees. Each member body interested in a subject for which a technical committee has been established hasthe right to be repr
22、esented on that committee. International organizations, governmental and non-governmental, inliaison with ISO, also take part in the work. ISO collaborates closely with the International ElectrotechnicalCommission (IEC) on all matters of electrotechnical standardization.International Standards are d
23、rafted in accordance with the rules given in the ISO/IEC Directives, Part 3.Draft International Standards adopted by the technical committees are circulated to the member bodies for voting.Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote
24、.International Standard ISO 10079-2 was prepared by Technical Committee ISO/TC 121, Anaesthetic andrespiratory equipment, Subcommittee SC 8, Suction devices for hospital and emergency care use.This second edition cancels and replaces the first edition (ISO 10079-2:1992), which has been technically r
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIAAMI1007921999MEDICALSUCTIONEQUIPMENTPART2MANUALLYPOWEREDSUCTIONEQUIPMENTPDF

链接地址:http://www.mydoc123.com/p-430236.html