ANSI AAMI 10079-1-1999 Medical Suction Equipment-Part I Electrically Powered Suction Equipment-Safety Requirements.pdf
《ANSI AAMI 10079-1-1999 Medical Suction Equipment-Part I Electrically Powered Suction Equipment-Safety Requirements.pdf》由会员分享,可在线阅读,更多相关《ANSI AAMI 10079-1-1999 Medical Suction Equipment-Part I Electrically Powered Suction Equipment-Safety Requirements.pdf(42页珍藏版)》请在麦多课文档分享上搜索。
1、ANSI/AAMI/ISO 10079-1:1999Medical Suction EquipmentPart I: Electrically Powered Suction EquipmentSafety Requirements American National StandardThis document was approved and published when the U.S. TAG for TC 121 was held by ASTM, but it is now an AAMI standard. The original formatting has been main
2、tained, so there are some variations from the typical AAMI style. Published by Association for the Advancement of Medical Instrumentation 4301 N. Fairfax Drive, Suite 301 Arlington, VA 22203-1633 www.aami.org 2014 by the Association for the Advancement of Medical Instrumentation All Rights Reserved
3、This publication is subject to copyright claims of ISO, and AAMI. No part of this publication may be reproduced or distributed in any form, including an electronic retrieval system, without the prior written permission of AAMI. All requests pertaining to this document should be submitted to AAMI. It
4、 is illegal under federal law (17 U.S.C. 101, et seq.) to make copies of all or any part of this document (whether internally or externally) without the prior written permission of the Association for the Advancement of Medical Instrumentation. Violators risk legal action, including civil and crimin
5、al penalties, and damages of $100,000 per offense. For permission regarding the use of all or any part of this document, complete the reprint request form at www.aami.org or contact AAMI, 4301 N. Fairfax Drive, Suite 301, Arlington, VA 22203-1633. Phone: 703-525-4890; Fax: 703-525-1067. Printed in t
6、he United States of America ISBN 157020537X ANS/ISO 10079-1 1999 Medical Suction Equipment Part 1: Electrically Powered Suction Equipment Safety Requirements Approved as an American National Standard by: ASTM International These materials are subject to copyright claims of ISO, ANSI, and ASTM Intern
7、ational. Not for resale without the prior written permission of ASTM International. No part of this publication may be reproduced in any form, including an electronic retrieval system, without the prior written permission of ASTM International. ASTM Deviations to ISO 10079-1: 1999 2 Normative refere
8、nces (Add the following) CGA Z168.11-94, Vacuum Devices Used for Suction and Drainage Rationale for ASTM deviation: CGA Z168.11-94 has requirements for intermittent and interrupted regulators that are not part of ISO 10079-1. 3 Terms and Definitions (Add the following) interrupted suction suction in
9、 which the negative pressure is automatically and periodically relieved to a lower negative pressure (for example, from -50 kPa to -5 kPa). single-fault condition condition in which a single means of protection against a hazard in equipment is defective or a single external abnormal condition is pre
10、sent Rationale for ASTM deviation: These definitions are needed because the terms are used in the documents. 6 Identification, marking and documents (Add the following to 6.8.2) 16) The manufacturer shall disclose whether or not the suction equipment is suitable for use in a magnetic resonance imagi
11、ng (MRI) unit. Rationale for ASTM deviation: This knowledge is necessary for safe use. 16.3 Components and general assembly (Make changes to the following subclauses) 56.5 (Add the following) 3) An intermittent regulator shall cycle through its off-time and on-time in accordance with the manufacture
12、rs specifications. Compliance shall be checked with the following test: Connect the regulator to a vacuum source as recommended by the manufacturer. Set the regulator to the intermittent mode and occlude the intermittent tubing. Set the vacuum to a level in the middle third of the gauge range during
13、 the on cycle. Record the on time and off time during five full cycles. 4) An intermittent regulator shall return to atmospheric pressure during the off-time for at least 3 s before commencing the next on-cycle. Compliance shall be checked with the following test: Connect the regulator to a vacuum s
14、ource as recommended by the manufacturer. Connect the patient side of the regulator to the vacuum end of a 2-L collection container. Occlude the patient side of the collection container with a pressure indicator having a range similar to that of the intermittent regulator. Set the intermittent regul
15、ator to the intermittent mode, occlude the intermediate tubing between the intermittent regulator and the collection container, and adjust the vacuum on the regulator to a level in the middle third of the gauge range. Release the occlusion on the intermittent tubing, ensure that the intermittent reg
16、ulator is cycling normally, and record the amount of time that the vacuum indicator indicates 0 kPa during the off cycle. 5) An interrupted regulator shall cycle between the set vacuum and a less negative vacuum in accordance with the manufacturers specifications. Compliance shall be checked with th
17、e following tests: a) Fixed vacuum relief Connect the regulator to a vacuum source as recommended by the manufacturer. Adjust the regulator to a vacuum 2.7 kPa more negative than the vacuum relief pressure. Allow the regulator to cycle through five full cycles and record the maximum and minimum pres
18、sures shown on the gauge and the times that these pressures are maintained. b) Adjustable vacuum relief Connect the regulator to a vacuum source as recommended by the manufacturer. Adjust the regulator to a vacuum 2.7 kPa more negative than the minimum vacuum relief pressure. Allow the regulator to
19、cycle through five full cycles and record the maximum and minimum pressures shown on the gauge and the times these pressures are maintained. Adjust the regulator to a vacuum of 2.7 kPa more negative than the maximum vacuum relief pressure. Allow the regulator to cycle through five full cycles and re
20、cord the maximum and minimum pressures on the gauge and the times these pressures are maintained. 6) An interrupted regulator shall return to a lower vacuum level before returning to the vacuum relief valve set point. Compliance shall be checked with the following tests: a) Fixed vacuum relief valve
21、s Connect the regulator to a vacuum source as recommended by the manufacturer. Connect the patient side of the interrupted regulator to the vacuum side of a 2-L collection container. The patient side of the collection container shall have a vacuum indicator having a range similar to that of the regu
22、lator. Adjust the regulator static set point to 2.7 0.2 kPa (20 mmHg) above the vacuum relief valve set point. Allow the vacuum relief valve to cycle and record the minimum vacuum level on the 2-L collection container. b) Adjustable vacuum relief valves Connect the regulator to a vacuum source as re
23、commended by the manufacturer. Connect the patient side of the interrupted regulator to the vacuum side of a 2-L collection container. The patient side of the collection container shall have a vacuum indicator having a range similar to that of the regulator. Adjust the regulator static set point to
24、2.7 0.2 kPa (20 mmHg) above the minimum vacuum relief valve set point. Allow the vacuum relief valve to cycle and record the minimum vacuum level on the 2-L collection container. Repeat for the maximum vacuum relief set point. Rationale for ASTM deviations: Above requirements can be found in CGA Z16
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIAAMI1007911999MEDICALSUCTIONEQUIPMENTPARTIELECTRICALLYPOWEREDSUCTIONEQUIPMENTSAFETYREQUIREMENTSPDF

链接地址:http://www.mydoc123.com/p-430235.html