VA 26 55 71-2010 MEDICAL AND SURGICAL LIGHTING FIXTURES.pdf
《VA 26 55 71-2010 MEDICAL AND SURGICAL LIGHTING FIXTURES.pdf》由会员分享,可在线阅读,更多相关《VA 26 55 71-2010 MEDICAL AND SURGICAL LIGHTING FIXTURES.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、09-10 26 55 71 - 1 SECTION 26 55 71 MEDICAL AND SURGICAL LIGHTING FIXTURES SPEC WRITER NOTES: 1. Delete between / - / if not applicable to project. Also delete any other item or paragraph not applicable in the section and renumber the paragraphs. 2. The A/E should coordinate with the Architect, Stru
2、ctural Engineer, and Mechanical Engineer to ensure that the ceiling, structure, and ceiling cavity are suitable and free of obstructions for surgical light installation. 3. The A/E shall verify with VHA the fixture type(s) desired for the operating room(s) affected by the project. 4. The A/E should
3、determine the need for spare conduits for future addition of arm-mounted video cameras or monitors. PART 1 - GENERAL 1.1 DESCRIPTION This section specifies the furnishing, installation, and connection of the surgical lighting fixtures. 1.2 RELATED WORK /A. Section 13 05 41, SEISMIC RESTRAINT REQUIRE
4、MENTS FOR NON-STRUCTURAL COMPONENTS: Seismic supports and lateral bracing for light fixtures./ B. Section 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS: General electrical requirements and items that are common to more than one section of Division 26. C. Section 26 05 26, GROUNDING AND BONDING
5、 FOR ELECTRICAL SYSTEMS: Requirements for personnel safety and to provide a low impedance path for possible ground fault currents. 1.3 QUALITY ASSURANCE Refer to Paragraph, QUALIFICATIONS, in Section 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS. 1.4 SUBMITTALS A. Submit in accordance with Sec
6、tion 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS. B. Shop Drawings: 1. Clearly present sufficient information to determine compliance with drawings and specifications. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-09-10 26 55 71 - 2 2. Incl
7、ude electrical ratings, dimensions, mounting details, materials required clearances, terminations, wiring and connection diagrams, lubrication, ballasts, lenses, louvers, lamps, and controls. 3. Include photometric data for surgical lighting fixture Types A, B, C, and D from an independent testing l
8、aboratory. The photometric report shall include data to show that the surgical light fixtures are in full compliance with requirements for illumination level, shadow reduction, beam heat, and color temperature. Where testing procedures or parameters are specified, the report shall indicate the surgi
9、cal light fixtures were tested to those criteria. C. Manuals: 1. When submitting the shop drawings, submit companion copies of complete maintenance and operating manuals, including technical data sheets, wiring diagrams, and information for ordering replacement parts. a. Wiring diagrams shall have t
10、heir terminals identified to facilitate installation, operation, and maintenance. b. Wiring diagrams shall indicate internal wiring for each item of equipment and the interconnections between the items of equipment. c. Provide a clear and concise description of operation including all details requir
11、ed to properly operate the equipment and system. d. Approvals will be based on complete submissions of manuals together with shop drawings. 2. Two weeks prior to final inspection, submit four copies of a final updated maintenance and operating manual to the /Resident Engineer/ /COTR/. a. The manual
12、shall be updated to include any information necessitated by shop drawing approval. b. Complete “As Installed“ wiring and schematic diagrams shall be included, showing all pieces of equipment and their interconnecting wiring. c. Show all terminal identification. d. Include information for testing, re
13、pair, troubleshooting, lubrication, assembly, disassembly, and recommended maintenance procedures and intervals. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-09-10 26 55 71 - 3 e. Provide a replacement parts list with current prices. Include a lis
14、t of recommended spare parts, lamps, tools, and instruments for testing and maintenance. D. Certifications: Two weeks prior to final inspection, submit four copies of the following to the /Resident Engineer/ /COTR/: 1. Certification by the manufacturer that the materials are in accordance with the d
15、rawings and specifications. 2. Certification by the contractor that the equipment has been properly adjusted, installed, and tested. 1.5 APPLICABLE PUBLICATIONS A. Publications listed below (including amendments, addenda, revisions, supplements, and errata) form a part of this specification to the e
16、xtent referenced. Publications are referenced in the text by designation only. B. Illuminating Engineering Society of North America (IESNA): RP-29-06 . Lighting for Hospitals and Health Care Facilities HB-9-00 Lighting Handbook Reference and Application C. National Fire Protection Association (NFPA)
17、: 70-08 National Electrical Code (NEC) 99-05 Health Care Facilities D. Underwriters Laboratories, Inc. (UL): 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety 1598-08 Luminaires PART 2 - PRODUCTS 2.1 SURGICAL LIGHTING FIXTURES, GENERAL A. Fixtures shall be in accordance w
18、ith UL 1598, NEC, NFPA 99, and IESNA RP-29, as shown on the drawings and as specified. B. Fixtures shall be complete, grounded, fungi-proof, adequately enclosed for asepsis, and designed for use in human operating rooms by a manufacturer that regularly produces such fixtures. C. Fixtures shall be su
19、pplied complete with suspension systems, lightheads, transformers, and controls. Components shall be products of a single manufacturer. D. Suspension components shall not flex during normal use. Articulation of the suspension to any position in its range shall maintain the lighthead at that point wi
20、thout drift. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-09-10 26 55 71 - 4 E. All exposed surfaces shall be free of burrs and sharp edges. Finishes on all exposed surfaces shall be specifically designed to resist scuffing and deleterious effects
21、 of the use of hospital cleaning materials. F. Except for finished aluminum, stainless steel, chrome, nickel and brass surfaces, all metal surfaces shall be thoroughly cleaned and painted at the factory with a corrosion-resistant primer and not fewer than two coats of lacquer or baked enamel finish.
22、 G. Maximum leakage current of each lighthead and its respective control shall not exceed 100 microamperes as measured in accordance with UL 60601-1. SPEC WRITER NOTE: Select type(s) appropriate for the project. 2.2 SURGICAL LIGHTING FIXTURE TYPES A. Single Lighthead and Pivot Arm, Single Point Susp
23、ension (Type A): Shall be a surgical light system of the single point suspension type with a single lighthead unit, mounted from a pivotal arm assembly. Lighthead shall rotate within a clearance circle of 142.67 in 3624 mmto 257.63 in 6544 mm, depending on light head site horizontal arm selection. C
24、enter of lighthead adjusted vertically from 46.85 in 1190 mm to 88.58 in 2250 mm above the floor. B. Dual Lightheads and Pivot Arms, Single Point Suspension (Type B): Shall be a major light system incorporating two identical lighthead units, each mounted on an independent arm assembly. The arm assem
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- VA2655712010MEDICALANDSURGICALLIGHTINGFIXTURESPDF

链接地址:http://www.mydoc123.com/p-1074977.html