UOP 954-2011 Loss on Ignition (LOI) for Fresh Regenerated Used and Spent Catalysts Catalyst Supports and Adsorbents.pdf
《UOP 954-2011 Loss on Ignition (LOI) for Fresh Regenerated Used and Spent Catalysts Catalyst Supports and Adsorbents.pdf》由会员分享,可在线阅读,更多相关《UOP 954-2011 Loss on Ignition (LOI) for Fresh Regenerated Used and Spent Catalysts Catalyst Supports and Adsorbents.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2003, 2011 UOP LLC. All rights reserve
3、d. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.83
4、2.9585 PHONE. Loss on Ignition (LOI) for Fresh, Regenerated, Used, and Spent Catalysts, Catalyst Supports, and Adsorbents UOP Method 954-11 Scope This method is for determining the loss on ignition (LOI) in air, using a specified ignition temperature and time, for fresh, regenerated, used, and spent
5、 catalysts, catalyst supports, and molecular sieves of various shapes and sizes. This method is quantitative down to 0.1 mass-%. The method is not water specific; any components that volatilize at the selected ignition temperature, for example, carbon, chlorine, and fluorine, are also included in th
6、e LOI value. Some components, for example sulfates, may volatilize fully or partially depending on ignition time and temperature. Mass gains are possible with catalysts that contain reduced metals. The criteria for selecting the temperature and time for the LOI are based on the composition of the sa
7、mple and the expected volatility of the impurities. Examples are listed in Procedure, Table 1. Length of time at temperature is dependent on the application of the LOI value. For example, it is important to drive off defined volatiles, such as halogens, for the determination of platinum concentratio
8、n. References UOP Method 918, “Collection use a porcelain cover, Cat. No. Z247235 (optional, see Note 1) Crucibles, porcelain, low, wide-form, 30-, 50-, and 100-mL capacities, Fisher Scientific, Cat. Nos. 07-955E, -955F, and 955G, respectively, with covers (optional) 07-970G, -970H, and 970J, respec
9、tively. For powder samples, a 35-mL round bottom crucible may be used to reduce the depth of the material, Fisher Scientific, Cat. No, 08-103D; the -970G cover will fit this crucible. Desiccator, 250-mm ID, and porcelain plate, Fisher Scientific, Cat. Nos. 08-631B and 08-636F, respectively Furnace,
10、muffle, capable of operation at 1000 15C, Fisher Scientific, Cat. No. 10-550-126 Hot plate, variable temperature to 250C, Fisher Scientific, Cat. No. 11-497-2A Thermometer, surface, with temperature probe, Omega Engineering, Cat. Nos. HH81 and 88108K, respectively Tongs, crucible, 530- and 230-mm le
11、ngth, Fisher Scientific, Cat. Nos. 15-207 and 15-186, respectively Reagents and Materials Reference to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Desiccant, molecular sieve, activated, moisture indicating, type 4A, 8-12 mesh, Cole-Par
12、mer, Cat. No. M1373, or, Drierite, indicating, 8-mesh, Fisher Scientific, Cat. No. 07-578-3A Grease, high-vacuum, Fisher Scientific, Cat. No. 14-635-5D Introduction There are a number of reasons for determination of the LOI. In some cases, the volatile component, typically water, is of interest. In
13、other cases, the intent is to place samples at a baseline state so that some other analysis may be made, and comparisons to another material are thus made on the same basis. The importance of the LOI value will impact the number of replicate samples analyzed. For general purposes, such as base metal
14、 determination, a single LOI determination is sufficient. For determination of noble metal on catalyst to meet specifications, duplicate LOI analyses are generally requested. Check the specific information for the material being analyzed to determine the LOI requirements before proceeding. In all ca
15、ses, the LOI must be performed with the sample in the appropriate physical state. For analysis of material to be shipped, the sample used to obtain the LOI must be in the same state as the sample when shipped: whole pill (sphere), whole extrudate, or powder; not ground. For noble metal analysis, sin
16、ce the noble metals are determined on a ground sample, the LOI must also be determined on a ground sample. Check the requirements or specifications of the material being analyzed, or any established method, standard operating procedure, or work instruction. For analyses where the LOI is used to calc
17、ulate the analysis results to a volatile free (VF) basis, always weigh the samples for LOI and elemental analysis at the same time. Where duplicate LOIs are stipulated, report the average of the two LOI values. If the two values differ by more than that expected from Precision, rerun both the LOIs a
18、nd the associated elemental analysis, since the sample may have gained or lost moisture during the interim. 954-11 3 of 8 For certain samples, or for experimental purposes, it may be desirable to measure an LOI to a constant weight. This procedure is described in Appendix B. Procedure The analyst is
19、 expected to be familiar with general laboratory practices and the equipment being used. CAUTION: Spent or used catalysts may ignite violently if introduced directly from the hot plate into the muffle furnace at 900C, due to carbon or hydrocarbon combustion. For this application, the sample is warme
20、d on a hot plate for an hour, or longer if necessary until the material is no longer visually wet or fuming. It is then introduced into the muffle furnace set at 500C or less for a specified time. The sample is then placed into a second muffle furnace set at the desired final temperature, often 900C
21、. Alternatively, the temperature of the first muffle furnace can be ramped to 900C, or other final temperature, over a period of time. The additional hardware required and the procedure are described in Appendix A. Materials that contain a high moisture level may be analyzed in the same manner. Samp
22、les should be introduced into the furnace in small batches to avoid overloading the furnace particularly for samples with specific characteristics, e.g, spent or used catalysts, etc. Typical LOI temperatures and times for various sample types are listed in Table 1. These are only examples. Specific
23、conditions are to be established for individual applications. If desired, more than one sample can be placed, each in its own crucible, into the muffle furnace at one time. Each crucible must then be marked for identification. Table 1 Time at Specified Temperature Sample Type Typical LOI Temperature
24、, C Typical Time in Muffle Furnace, hours Catalysts with Tungsten or Molybdenum 500 1 Catalysts with Potassium 700 11 Catalysts 900 11 Molecular sieves 1,000 1 Spent or used catalyst 500 to 900 See caution statement above or Appendix A.2 Calcined supports 700 or 900 1 1For platinum analysis, use 6 h
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9542011LOSSONIGNITIONLOIFORFRESHREGENERATEDUSEDANDSPENTCATALYSTSCATALYSTSUPPORTSANDADSORBENTSPDF

链接地址:http://www.mydoc123.com/p-1074584.html