UOP 868-1988 TRACE SATURATES IN HIGH PURITY AROMATICS BY GC.pdf
《UOP 868-1988 TRACE SATURATES IN HIGH PURITY AROMATICS BY GC.pdf》由会员分享,可在线阅读,更多相关《UOP 868-1988 TRACE SATURATES IN HIGH PURITY AROMATICS BY GC.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、TRACE SATURATES IN HIGH PURITY AROMATICS BY GCUOP Method 868-88SCOPEThis method is for determining trace paraffins and naphthenes by carbon number from C3through C10inolefin-free C6, C7or C8aromatics or mixtures thereof (see Note 1). The range of detection for C3to C8paraffins or naphthenes is 2 to
2、2,000 mass-ppm. The range of detection for C9and C10paraffins andnaphthenes is 100 to 2,000 mass-ppm.OUTLINE OF METHODA repeatable sample volume is injected into a gas chromatograph that is equipped with a two-columnchromatographic system and a flame ionization detector. The first column is polar, p
3、acked with OV-275 onChromosorb, and the second column is selective, packed with specially treated 13X molecular sieves.Initially, the two columns are connected in series. Immediately before the elution of benzene from the polarcolumn, the polar column is backflushed to vent while the eluted saturate
4、d hydrocarbons are analyzed onthe selective column. The external standard method of quantitation is used.APPARATUSBalance, readability 0.1-mgChromatographic columns:Column 1: 3 m of 2-mm ID (1/8-inch OD) stainless steel tubing, packed with 30% OV-275 on 60-80mesh Chromosorb PNAW, Alltech Associates,
5、 or equivalentColumn 2: 1.6 m of 2-mm ID (1/8-inch OD) stainless steel tubing, packed with specially treated 13Xmolecular sieves, Analytical Controls, Part No. 50.06.530Gas Chromatograph, built for dual packed column operation, with dual flow controllers, pressure gaugesand a single flame ionization
6、 detector, capable of column temperature programming to 400 C, andequipped with a separate isothermal heated zone large enough to contain 2 rotary valves, 2 restrictorvalves and a column. The detector must be capable of producing a minimum peak height response of10 times the background noise for 2 m
7、ass-ppm of n-hexane when operated at the recommendedconditions (see Note 2).Integrator, electronic, or equivalent for obtaining peak areasIT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TODETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIV
8、E HEALTH ANDSAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THISPROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS(MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED INTHIS PROCEDURE SHOUL
9、D BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTIONEQUIPMENT (PPE). COPYRIGHT 1986, 1988 UOP LLCALL RIGHTS RESERVEDUOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,United States. The Methods may be obtained throu
10、gh the ASTM website, www.astm.org, or by contacting Customer Service atserviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE.2 of 7868-88Oxygen remover, Oxy-Trap, Alltech Associates, Cat. No. 4001, or equivalent. (An oxygen remover in thecarrier gas line ensures optimum column life.)Recorder, 1-m
11、V full scale, 1-sec or less full-scale responseRegulator, air, two-stage, high purity, Matheson, Model 3104-590, or equivalentRegulator, hydrogen, two-stage, high purity, Matheson, Model 3104-350, or equivalentRegulator, nitrogen, two-stage, high purity, Matheson, Model 3104-580, or equivalentRestri
12、ctor, fine metering valve, Nupro, Cat. No. SS-1-SG, or equivalent, 2 required (see Note 2).Sample injector, any syringe or injection system capable of injecting a 1.0-L volume of sample, such asa Hamilton 701-NWG syringe, Alltech Associates, or equivalentStopwatch, electronic, digital, Alltech Assoc
13、iates, Cat. No. 4061, or equivalentValve, 6-port rotary, Valco Instruments Co., Model C6UWP, or equivalent, 2 required (see Note 2).REAGENTS AND MATERIALSAll reagents shall conform to the specifications established by the Committee on Analytical Reagents ofthe American Chemical Society, when such sp
14、ecifications are available, unless otherwise specified.References to water mean deionized or distilled water, except where noted.Air, zero gas, total hydrocarbons less than 2.0 ppm as methaneBenzene, 99.9% minimum purity, Wiley Organics, Cat. No. 0160.84, or equivalentButylcyclohexane, 99.9% minimum
15、 purity, Wiley Organics, Cat. No. 0419.71-1, or equivalent (C10naphthene)Cyclohexane, 99.9% minimum purity, Wiley Organics, Cat. No. 0679.02, or equivalent (C6naphthene)n-Decane, 99.9% minimum purity, Wiley Organics, Cat. No. 0812.57, or equivalent (C10paraffin)Ethylcyclohexane, 99.9% minimum purity
16、, Wiley Organics, Cat. No. 2442.00, or equivalent (C8naphthene)n-Heptane, 99.9% minimum purity, Wiley Organics, Cat. No. 3111.00, or equivalent (C7paraffin)n-Hexane, 99.9% minimum purity, Wiley Organics, Cat. No. 3442.00, or equivalent (C6paraffin)Hydrogen, zero gas, total hydrocarbons less than 0.5
17、 ppm as methaneMethylcyclohexane, 99.9% minimum purity, Wiley Organics, Cat. No. 4082.00, or equivalent (C7naphthene)Nitrogen, zero gas, total hydrocarbons less than 0.5 ppm as methanen-Nonane, 99.5% minimum purity, Wiley Organics, Cat. No. 6782.00 or equivalent (C9paraffin)n-Octane, 99.9% minimum p
18、urity, Wiley Organics, Cat. No. 7056.50, or equivalent (C8paraffin)Propylcyclohexane, 99.8% minimum purity, Wiley Organics, Cat. No. 8260.05-1, or equivalent (C9naphthene)Toluene (methylbenzene), 99.9% minimum purity, Wiley Organics, Cat. No. 3829.90, or equivalent3 of 7868-88PROCEDUREInstall the ox
19、ygen remover in the supply line between the hydrogen source regulator and the carrier gasinlet on the gas chromatograph.The recommended operating conditions for this analysis are given in Table 1. Other conditions may beused provided they produce the required sensitivity and chromatographic separati
20、ons equivalent to thoseshown in the Typical Chromatogram (Fig. 2).Table 1Operating ConditionsCarrier gas hydrogenCarrier gas flow rate 15 mL/minDetector flame ionizationDetector temperature 250 CInjection port temperature 200 CHydrogen flow rate* 15 mL/minAir flow rate* 400 mL/minMakeup gas nitrogen
21、Makeup gas flow rate* 30 mL/minColumn 1 temperature 120 CColumn 2 temperature programInitial temperature 100 CProgramming rate 6 C/minFinal temperature 400 CFinal time 10 minChart speed 0.5 cm/minSample size 1.0 L repeatable*Consult the manufacturers instrument manual for suggested flow rates.Prepar
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP8681988TRACESATURATESINHIGHPURITYAROMATICSBYGCPDF

链接地址:http://www.mydoc123.com/p-1074558.html