UOP 543-2011 Trace Non-Aromatic Hydrocarbons in High-Purity Aromatics by GC.pdf
《UOP 543-2011 Trace Non-Aromatic Hydrocarbons in High-Purity Aromatics by GC.pdf》由会员分享,可在线阅读,更多相关《UOP 543-2011 Trace Non-Aromatic Hydrocarbons in High-Purity Aromatics by GC.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1970, 1979, 1997, 2011 UOP LLC. All ri
3、ghts reserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FA
4、X, or 610.832.9585 PHONE. Trace Non-Aromatic Hydrocarbons in High-Purity Aromatics by GC UOP Method 543-11 Scope This method is primarily for determining total non-aromatic hydrocarbons below the 1 mass-% level in individual or mixed aromatic hydrocarbon samples. For the purpose of this method, tota
5、l non-aromatics are defined as the composite of those components eluting before benzene. The applicability is limited to samples containing compounds having a maximum of 11 carbon atoms. The method may also be used to determine mass-ppm concentrations of benzene in specific sample types containing p
6、redominantly C7 and higher boiling aromatics, or benzene and toluene in specific sample types containing predominantly C8 and/or higher boiling aromatics. The lower limit of quantitation for any single component is 1 mg/kg (mass-ppm). References ASTM D4052, “Density, Relative Density, and API Gravit
7、y of Liquids by Digital Density Meter,” www.astm.org ASTM D4307, “Preparation of Liquid Blends for Use as Analytical Standards,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is injected into a gas chromatograph (GC) that is equipped wi
8、th a fused silica capillary column internally coated with 1,2,3-tris(2-cyanoethoxy)propane (TCEP), and a flame ionization detector (FID). The concentrations of individual or group impurities are determined by the external standard method of quantitation, wherein peak areas of the sample components a
9、re compared to the peak areas of a calibration blend analyzed under identical conditions and injection volumes. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability 0.1 mg 2 of 8 543-11 Chromatograph
10、ic column, 60 m of 0.25-mm ID fused silica capillary, internally coated to a film thickness of 0.4-m with TCEP, Restek Cat. No. 10999 Freezer, explosion-proof, -23 to -5C, Lab-Line Model 3552, VWR Scientific, Cat. No. 55700-158 Gas chromatograph, capable of temperature ramping, equipped with electro
11、nic pressure control, built for capillary column chromatography utilizing a split injection system, having a glass injection port insert, and equipped with a flame ionization detector that will give a minimum peak height response of 10 times the background noise for 1 mg/kg of n-undecane when operat
12、ed at the recommended conditions, Agilent Technologies, Model 7890 Data system, electronic, for obtaining peak areas. This device must integrate areas at a sufficiently fast rate so that narrow peaks typically resulting from the use of a capillary column can be accurately measured. Agilent Technolog
13、ies, ChemStation. Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, hydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen, two-stage, hig
14、h purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Sample injector, any syringe or injector capable of injecting a repeatable 1.0-L volume of sample. The use of an automatic injection device is required to achieve necessary repeatable injection volumes. See Ap
15、pendixes. Agilent Technologies, Model 7683. Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Air, zero-gas, total hydrocarbons less than 2 ppm as methane, local supply Benzene, 99.9% purity, Sigma-Aldric
16、h, Cat. No. 270709. CAUTION: Benzene is a known carcinogen. All operations involving its use must be performed in a properly ventilated area, while wearing appropriate personal protective equipment. n-Dodecane, 99+% purity, Sigma-Aldrich, Cat. No. 297879 Hydrogen, zero-gas, 99.99% minimum purity, to
17、tal hydrocarbons less than 0.5 ppm as methane, local supply Gas purifier, for hydrogen, to remove oxygen and moisture from carrier gas, VICI Mat/Sen, Cat. No. P200-1 Nitrogen, zero-gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane, local supply Pipets, disposable, Pasteur,
18、VWR, Cat. No. 14673-043 Pipet bulbs, VWR, Cat. No. 15001-362 Syringe, replacement, for recommended sample injector, 5-L, Agilent Technologies, Cat. No. 5181-1273 Toluene, 99.5% purity, Sigma-Aldrich, Cat. No. 244511 3 of 8 543-11 n-Undecane, 99+% purity, Sigma-Aldrich, Cat. No. U407 Vials, autosampl
19、er, for recommended sample injector, with caps, Agilent Technologies, Cat. No. 5182-0864 p-Xylene, 99+% purity, Sigma-Aldrich, Cat. No. 317195 Preparation of Apparatus The analyst is expected to be familiar with general laboratory practices, the technique of gas chromatography, and the equipment bei
20、ng used. 1. Install the gas purifier in the supply line between the carrier gas source and the carrier gas inlets on the gas chromatograph. Column life is significantly reduced if the gas purifier is not used. 2. Install the fused silica column in the gas chromatograph according to the column and ga
21、s chromatograph manufacturers instructions. CAUTION: Hydrogen gas leakage into the confined volume of the column oven can cause violent explosion. It is, therefore, mandatory to check for leaks each time a connection is made and periodically thereafter. 3. Establish the recommended operating conditi
22、ons as given in Table 1. Different conditions may be used provided they produce the required sensitivity and chromatographic separations equivalent to those shown in the Typical Chromatograms (Figures 1-2). Table 1 Recommended Operating Conditions Carrier gas hydrogen Head pressure 138 kPa gauge (20
23、 psig) Linear velocity 120C 40 cm/sec Equivalent flow 120C 1.6 mL/min Split flow 50 mL/min Injection port temperature 175C Column temperature program Initial temperature 120C (isothermal) Detector flame ionization Detector temperature 200C Hydrogen flow ratea 38 mL/min Air flow ratea 450 mL/min Make
24、up gas nitrogen Makeup gas flow ratea 45 mL/min Sample size 1.0 L, repeatable aConsult the manufacturers instrument manual for suggested flow rates. 4. Program the column oven to 130C and maintain this temperature until a stable baseline has been obtained at the required sensitivity. 5. Cool the col
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP5432011TRACENONAROMATICHYDROCARBONSINHIGHPURITYAROMATICSBYGCPDF

链接地址:http://www.mydoc123.com/p-1074502.html