UOP 294-2014 Apparent Bulk Density of Catalyst Particles.pdf
《UOP 294-2014 Apparent Bulk Density of Catalyst Particles.pdf》由会员分享,可在线阅读,更多相关《UOP 294-2014 Apparent Bulk Density of Catalyst Particles.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1966, 1988, 1993, 2014 UOP LLC. All rights reserved. Nonco
3、nfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 P
4、HONE. Apparent Bulk Density of Catalyst Particles UOP Method 294-14 Scope This method is for determining the apparent bulk density (ABD) of loosely packed spherical catalyst particles with an average particle diameter of 1.6- to 3.2-mm ID. It is also applicable to extruded catalyst particles having
5、a maximum diameter of 3.2 mm, a maximum average length of 9.5 mm, and a maximum individual length of 16 mm. Apparent bulk density is the mass per unit volume of the particulate material. Reference UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is pou
6、red into a weighed and volume-calibrated cylinder. The catalyst is leveled to the top of the cylinder and weighed. ABD is calculated by dividing the mass of the catalyst by the volume of the cylinder. Apparatus References to catalog numbers and suppliers are included as a convenience to the method u
7、ser. Other suppliers may be used. Balance, readable to 0.01-g Beaker, 250-mL, VWR, Cat. No. 89000-204 Cover, cylinder, flat plastic, glass, or other transparent and non-porous material, approximately 50-mm (2-inches) square, local supply Cylinder, graduated, glass, 250-mL, 38-mm ID, Fisher Scientifi
8、c, Cat. No. 08-555E, cut down to the 100-mL mark and fire polished, “weighing cylinder.” A cylindrical crucible, 100-mL, 39-mm diameter, 91-mm height, CoorsTek, Cat. No. 65539, is also suitable Funnel, powder, polypropylene, 75 to 80-mm top diameter, Fisher Scientific, Cat. No. 10-500-1 Ring, suppor
9、t, iron, with screw clamp, slotted, VWR, Cat. No. 60130-004 Spatula, 152-mm straight-edge blade, VWR, Cat. No. 82027-462 Support, rectangular base, steel rod, VWR, Cat. No. 60110-222 2 of 4 294-14 Thermometer, -l to 51 C range, 0.1 C subdivisions, partial immersion, spirit filled, VWR, Cat. No. 6122
10、2-624 Timer/stopwatch, VWR, Cat. No. 98000-186 Tray, polypropylene, medium-deep form to catch scattered particles, large enough to accommodate ring support, Fisher Scientific, Cat. No. 13-359-20A Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the
11、method user. Other suppliers may be used. References to water mean deionized or distilled. Tissues, Kimwipes, Fisher Scientific, Cat. No. 06-666A Water, deionized or distilled Procedure The analyst is expected to be familiar with general laboratory procedures and the equipment being used. Dispose of
12、 used supplies and samples in an environmentally safe manner according to applicable regulations. Volume Calibration of the Weighing Cylinder The weighing cylinder must be calibrated for volume initially and on a periodic basis thereafter to ensure accurate analyses. For a certified calibration, the
13、 weighing cylinder is sent out to a certified calibration laboratory, e.g., The Calibration Solution, for calibration. Alternatively, if a certified calibration is not necessary, the calibration may be done in house by filling the cylinder with water and weighing it as described below. For in house
14、calibration only: 1. Weigh the dry cylinder and cylinder cover to the nearest 0.01 g. 2. Fill the cylinder to the brim with water. Dislodge any bubbles adhering to the cylinder walls by gently tapping the sides, being careful not to spill the water. Add additional water, if necessary, without introd
15、ucing bubbles. 3. Touch one edge of the cover to the rim of the cylinder. Lower the cover onto the cylinder while carefully excluding all air bubbles from below the cover. Dry the outside of the cylinder carefully. 4. Weigh the filled cylinder and cover to the nearest 0.01 g. 5. Remove the cylinder
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP2942014APPARENTBULKDENSITYOFCATALYSTPARTICLESPDF

链接地址:http://www.mydoc123.com/p-1074481.html