REG NASA-LLIS-0622-1999 Lessons Learned Use of Obsolescent Test Facilities Pose Risk to Hardware (1997).pdf
《REG NASA-LLIS-0622-1999 Lessons Learned Use of Obsolescent Test Facilities Pose Risk to Hardware (1997).pdf》由会员分享,可在线阅读,更多相关《REG NASA-LLIS-0622-1999 Lessons Learned Use of Obsolescent Test Facilities Pose Risk to Hardware (1997).pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Lessons Learned Entry: 0622Lesson Info:a71 Lesson Number: 0622a71 Lesson Date: 1999-05-21a71 Submitting Organization: JPLa71 Submitted by: A. Acord/W. Moore/D. OberhettingerSubject: Use of Obsolescent Test Facilities Pose Risk to Hardware (1997) Abstract: Cassini: At the beginning of Assembly, Test,
2、 and Launch Operations (ATLO), vibration testing of the massive Cassini spacecraft was performed using facility test equipment that was barely up to the task. Powering an aged power amplifier for a shaker resulted in multiple failures of the test equipment, including board fires. Proactively perform
3、 periodic test facility maintenance and upgrade, and validate the current performance of the environmental test facility prior to use.Description of Driving Event: At the beginning of Cassini Assembly, Test, and Launch Operations (ATLO), the readiness and condition of the JPL Environmental Test Labo
4、ratory equipment was poorly understood.Vibration testing of the massive Cassini spacecraft was performed using facility test equipment that was barely up to the task.a71 Proof mass vibration tests had to be aborted due to instrumentation problems. Since not all elements of the standard instrumentati
5、on suite were needed, various test instruments were uncoupled until the configuration provided a workable subset.a71 a71 The amplifier used to drive the shaker was approaching the end of its useful life during the Cassini proof mass vibration tests. The power amplifier includes 3000 power transistor
6、s on 10 circuit boards. Several days were spent getting the amplifier operational. Powering the amplifier resulted in multiple failures of the test equipment, including board fires. The resultant uncontrolled shutdown of the shaker would have posed a significant risk had the test been conducted on f
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- REGNASALLIS06221999LESSONSLEARNEDUSEOFOBSOLESCENTTESTFACILITIESPOSERISKTOHARDWARE1997PDF

链接地址:http://www.mydoc123.com/p-1018269.html