EN ISO 6817-1995 en Measurement of Conductive Liquid Flow in Closed Conduits - Method Using Electromagnetic Flowmeters《封闭管道中导电液体流量的测量采用 电磁流率计的方法 ISO 6817-1992》.pdf
《EN ISO 6817-1995 en Measurement of Conductive Liquid Flow in Closed Conduits - Method Using Electromagnetic Flowmeters《封闭管道中导电液体流量的测量采用 电磁流率计的方法 ISO 6817-1992》.pdf》由会员分享,可在线阅读,更多相关《EN ISO 6817-1995 en Measurement of Conductive Liquid Flow in Closed Conduits - Method Using Electromagnetic Flowmeters《封闭管道中导电液体流量的测量采用 电磁流率计的方法 ISO 6817-1992》.pdf(39页珍藏版)》请在麦多课文档分享上搜索。
1、STD-BSI BS-EN IS b817-ENGL 1777 1b2LibbS 0578807 ATi = BRITISH STANDARD BS EN IS0 6817 : 1997 Measurement of conductive liquid flow in closed conduits - Method using electromagnetic flowmeters BS 5792 : Part 1 : 1993 ?WVhU*, iwmting Amendment No. 1 The Euro- Standar EN IS0 6817 : 1995 has the status
2、 of a British standard ICs 17.120.10 NO COPYING WEOuT BSI PERMISSION EXCEPT AS PEBMiTED BY COPYBIGHT LAW STD-BSI BS EN IS0 bBL7-ENGL L7 LbZLi,bb 588A 73T BS EN IS0 6817: 1996 Issue 2, January 1997 mis British Standard. having been prepared under the direction of the industrial-process Measurement an
3、d Control Standards Policy Committee, WS publislied under the authority of tlic Standards hard and coines into effect on 15 March 1993 BSI 1997 The following in closed conduits (Identical) BS 5844 : 1980 Methods afmew.rm af.fluid.fnu: IS0 5168 : 1978 - estimatwn ofuncertaint b) eliminate, as far as
4、possible, spurious e.m.fs. These include common mode and quadrature signals; . c) provide means of compensating for supply volt- age and frequency variations where necessary; d) provide means of compensating or minimizing magnetic field strength variatioris in the primary device. This is important s
5、ince it directly affects repeatability of the voltage at the measurement electrodes. Compensation is achieved by the following means: a) a gain-compensated amplifier in which the gain is proportional to the supply frequency and in- versely proportional to the supply voltage; b) a system in which the
6、 output is proportional to the ratio of the flow signal and a reference signal derived from the field current. At a given flow- rate both signals may vary with supply voltage and frequency, but their ratio will remain con- stant: c) a system in which the field current is stabilized. For alternating
7、current (a.c.) energized systems with unregulated coil current, the secondary device measures the ratio of V/B (see clause 5). Voltages other than the flow signal (V) may be picked up by electrode leads. These voltages may be generated by the varying flux intersecting a loop composed of the electrod
8、e leads, the electrodes, and the liquid connecting the electrodes (transformer effect). Such a voltage will be approximately 90” out of phase with the flow signal. That portion which is 90” out of phase is called “quadrature”. The remainder is called the “in-phase” component. The “in-phase” componen
9、t is zeroed at no-flow during initial instal- lation, unless the flowmeters have a device which provides this function automatically. If the coil current is regulated, the magnetic field is considered to be constant and it is only necessary to measure the electrode signal. If the coil current is not
10、 regulated, then, in order to compensate for variations in the magnetic field, the secondary de- vice may use a reference signal obtained from the primary element. This reference signal may be de- rived from the supply voltage, the supply current, the flux density in the metal or the flux density in
11、 the air gap. In a pulsed d.c. system, under ideal or reference conditions, the peak-to-peak value of the electrode signals, (V, + V,), is proportional to the flow velocity in the pipeline and V, is also equal to V, see figure 4a), where Vp = positive voltage and V, = negative voltage. In a practica
12、l situation, if the zero or “no-flow” signal is offset in the positive direction by an amount Ve, then the positive signal is (V, + Ve) and the negative signal is (i,- Ve) Cfigure4b)l. Hence the overall value of the electrode signal is (V, + V,) and the offset zero is eliminated. The same applies if
13、 the offset is in the negative direction. The system thus eliminates zero errors automati- cally at all times and zero adjustment is not usually required, either at start-up/commissioning or at any time during subsequent operation. General guidance on the function and installation of secondary devic
14、es is presented in 7.2. 6.4 System output The system output can be one or more of the fol- lowing: a) analog direct current in accordance with IEC 381-1; b) analog direct voltage in accordance with IEC 381-2; c) a frequency output in the form of scaled or un- scaled pulses; d) digital. 6.5 Effect of
15、 the liquid conductivity If the electrical conductivity of the liquid is uniform in the measuring section of the meter, the electric field distribution is independent of the liquid con- ductivity and therefore the meter output is generally independent of the liquid conductivity. Minimum op- erationa
16、l conductivity requirements should be ob- tained from the manufacturers. The internal impedance of the primary device obvi- ously depends upon the liquid conductivity, and very large changes in this impedance may produce er- rors in the output signal. If the conductivity is not uniform throughout th
17、e meter, errors may also oc- cur. A heterogeneous fluid composed of small parti- cles uniformly distributed in a medium can be considered as a homogeneous liquid. Deposition of electrically conducting layers on the inside surface of the liner may also lead to errors. 6 STD-BSI BS EN IS0 b8L7-ENGL 19
18、97 Lb24bb7 0578820 237 m f . . . IS0 68171992(E) . . i . sp . . . . sc 6.6 Reynolds number effect In industrial electromagnetic flowmeters, the effect of Reynolds number is usually so small that for practical purposes it can be ignored. 1 . - I 6.7 Velocity profile effect Distortions in velocity pro
19、files may be caused by pipe fittings (bends, valves, reducers, etc.) placed upstream or downstream from the flowmeter; the resulting flow patterns may have an influence on the performance of the meter. In general, the user should comply with the manu- facturers recommendations for installation in or
20、der to minimize these effects. I Flow pattern effects are described in 7.1.2.1. 7 Installation design and practice 7.1 Primary devices 7.1.1 Sire Usually the bore of the primary device tube will be the same as that of the adjacent pipework. If, in this case, the mean axial velocity corresponding to
21、the maximum flow-rate is less than that recommended by the manufacturer, a primary device with a smaller bore should be used. A primary device with a bore smaller than that of the adjacent pipework may also be used for other reasons, e.g. to reduce cost or in the interests of rationalization. Inform
22、ation on the allowable tolerances for matching the pipe and meter tube bores is given in IS0 9104. Zero al Ideal condltlons Random noise No-flow Zero signal b) Practlcal condltlons Figure 4 - Principle of pulsed d.c. (bipolar) system 7 STD*BSI BS EN IS0 b817-ENGL 1977 3b24bb7 0578823 373 = IS0 681 7
23、:1992(E) 7.1.2 Layout not be within the manufacturers stated accuracy. If necessary, warning devices should be installed to There is no theoretical restriction on the attitude at preserve measurement integrity. Partially-filled pri- which a primary device may be mounted, provided mary device meters
24、are used, for example in sew- the pipe remains full at all times. Locations close to age applications, but these merit special electrical equipment which may interfere with the consideration outside the scope of this International flow measurement signal, or locations where cur- Standard. rents may
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ENISO68171995ENMEASUREMENTOFCONDUCTIVELIQUIDFLOWINCLOSEDCONDUITSMETHODUSINGELECTROMAGNETICFLOWMETERS

链接地址:http://www.mydoc123.com/p-723063.html